Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR… recallZ-0280-2016 · Cyberonics, Inc · initiated Oct 7, 2015

Class II Terminated All recalls posted in November 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cyberonics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.

FDA root cause category: Manufacturing material removal

What the company is doing

Cyberonics sent written recall notifications beginning October 14, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. The notice also directs customers to contact Cyberonics to schedule additional on-site troubleshooting. The on-site troubleshooting proposed by Cyberonics will utilize a feature included within the Model 250 Version 9.0 Programming Software as an alternate method of verifying heart rate and determining that the appropriate heart beat sensitivity setting is programmed. The Version 9.0 software displays an instantaneous heart rate when the user selects the 'Refresh' feature. This heart rate is measured directly by the generator and displayed on the programmer. For further questions, please call (866) 882-8804.

Recall details

Recall number
Z-0280-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 7, 2015
Posted by the FDA
Nov 17, 2015
Recalling firm
Cyberonics, Inc
Quantity
27 units
Product code
LYJ
FDA submissions
P970003
Distribution
US Distribution to the states of : OR, NY, GA, PA, CA, NC, KS, FL, KY, MI, TX, IL, OH, IN, UT, CO and WV.

Product description

Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106

Code information (models, lots, serial numbers)

Serial Numbers: 37719, 37899, 38123, 38414, 38428, 39385, 39531, 39607, 39801, 39821, 40037, 40048, 40249, 40251, 40271, 40354, 40428, 40488, 40526, 40603, 40751, 41000, 41109, 41152, 41805, 42022, 42780

Other devices under the same FDA submission

23 devices cleared under P970003 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.