Dimension Vista System (MG) recallZ-0339-2016 · Siemens Healthcare Diagnostics, Inc. · initiated Sep 30, 2015
Class II Terminated All recalls posted in November 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Erroneous low results on a small number of reagent wells. There is the potential for under-recovery on certain wells for quality control (QC) and patient sample results. The under-recovery ranged from -0.3 mg/dL [0.12 mmol/L] to -1.6 mg/dL [0.66 mmol/L]; equivalent to a maximum negative bias of -88%.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An Urgent Medical Device Recall dated September 2015 was provided to all Dimension Vista MG customers who received the affected lot to notify them of the issue, the potential risk to health, and actions to be taken by customers. 1. An Urgent Medical Device Recall was provided to all customers who received the affected lot to notify them of the following: Siemens has received customer complaints for Dimension Vista MG lot 15063BA, regarding sporadic depressed serum/plasma and QC results. Discontinue use and discard their remaining inventory of the affected lot. Review the letter with their Medical Director. Siemens will replace any unused inventory of the affected lot.
Recall details
- Recall number
- Z-0339-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 30, 2015
- Posted by the FDA
- Nov 25, 2015
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 3601
- Product code
- JGJ
- FDA submissions
- K061655
- Distribution
- Distributed US (nationwide) including DC and the states of IL, CA, GA, LA, TX, TN, FL, SC, IA, TN, DE, NJ, WA, MS, KY, NY, WI, OH, OH, MO, UT, KS, VA, AR, WV, MA, PA, AL, CO, IN, MT, NM, ME, AZ, NM, MO, and NC, and the country of Canada.
Product description
Dimension Vista System (MG) ; Flex reagent cartridge Catalog number K3057, SMN 10445158. For the quantitative measurement of magnesium in human serum, plasma and urine on the Dimension Vista System.
Code information (models, lots, serial numbers)
Lot Number 15063BA
Other devices under the same FDA submission
6 devices cleared under K061655 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Dimension Vista® Flex® reagent cartridge VANC 10445128Siemens Healthcare Diagnostics Inc. · RADIOIMMUNOASSAY, VANCOMYCIN13 reports
- Dimension Vista® Flex® reagent cartridge ACTM 10445110Siemens Healthcare Diagnostics Inc. · COLORIMETRY, ACETAMINOPHEN2 reports
- Dimension Vista® Flex® reagent cartridge CHOL 10445143Siemens Healthcare Diagnostics Inc. · ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL1 report
- Dimension Vista® Flex® reagent cartridge SAL 10445108Siemens Healthcare Diagnostics Inc. · COLORIMETRY, SALICYLATE1 report
- Dimension Vista® Flex® reagent cartridge GLU 10445162Siemens Healthcare Diagnostics Inc. · HEXOKINASE, GLUCOSE1 report
- Dimension Vista® Flex® reagent cartridge PHOS 10445144Siemens Healthcare Diagnostics Inc. · Phosphomolybdate (colorimetric), inorganic phosphorus1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page