Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI… recallZ-0375-2016 · Synthes (USA) Products LLC · initiated Nov 2, 2015
Class II Terminated All recalls posted in December 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the handle. Or 2. Cause the bottom of the cannulated connecting screw to tighten against the top of the locking mechanism preventing advancement of the locking mechanism.
FDA root cause category: Process control
What the company is doing
The firm, DePuy Synthes, sent an "URGENT NOTICE:MEDICAL DEVICE FIELD SAFETY NOTIFICATION" letter dated 11/2/2015 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Review the instructions listed in this notification; forward this notice to anyone in your customer's facility that needs to be informed; complete the attached Verification Section and send a copy of the completed Verification Section by: Fax: (866) 629-6186 or Scan/email: [email protected], if the Verification Form is answered on behalf of more than one facility and/or individual, please clearly indicate the name and address of the facility and/or individual, and maintain a copy of this notice. If you have any questions, please call (610)719-5450.
Recall details
- Recall number
- Z-0375-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 2, 2015
- Posted by the FDA
- Dec 4, 2015
- Recalling firm
- Synthes (USA) Products LLC
- Quantity
- 5
- Product code
- HSB
- FDA submissions
- K011857
- Distribution
- US Distribution to states of: CA, FL, MI and WI.
Product description
Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System) The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures. Intramedullary fixation rod, orthopedic.
Code information (models, lots, serial numbers)
Part Number Lot Number 04.037.162S 7950511
Other devices under the same FDA submission
44 devices cleared under K011857 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 456.305Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES9 reports
- NA 456.303Synthes GmbH · Screw, fixation, bone7 reports
- NA 456.315SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES7 reports
- NA 456.318SSynthes GmbH · Screw, fixation, bone6 reports
- NA 456.306Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES5 reports
- NA 456.302Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.304SSynthes GmbH · Screw, fixation, bone4 reports
- NA 456.306SSynthes GmbH · Screw, fixation, bone4 reports
- NA 456.322SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.420SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.305SSynthes GmbH · Screw, fixation, bone3 reports
- NA 456.415SSynthes GmbH · Screw, fixation, bone3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page