Medical device recalls: December 2015
The FDA posted 108 medical device recall records in December 2015, from 65 recall events by 57 companies. 9 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips Medical Systems (Cleveland) Inc · 3 events, 8 records
- Becton Dickinson & Company · 2 events, 6 records
- Zimmer, Inc. · 2 events, 4 records
- Merz North America, Inc. · 2 events, 3 records
- CareFusion 303, Inc. · 2 events, 2 records
- Draeger Medical, Inc. · 2 events, 2 records, 2 Class I
- Siemens Medical Solutions USA, Inc · 2 events, 2 records
- LeMaitre Vascular, Inc. · 1 event, 13 records
FDA root cause
- Software design · 16
- Process change control · 13
- Packaging process control · 13
- Nonconforming Material/Component · 13
- Process control · 11
- Under Investigation by firm · 10
- Other · 7
- Employee error · 7
Class I recalls (9)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0436-2016 · Draeger Medical, Inc. · posted Dec 23, 2015 · Terminated
The firm became aware of cases in which the battery run times of the optional PS 500 power supply unit with the Infinity Workstation Critical Care (Evita Infinity V 500) were unexpectedly short due to the design of the charging algorithm in the current…
- Z-0439-2016 · Biomerieux · posted Dec 23, 2015 · Terminated
Due to the potential for reporting erroneous results for Piperacillin/Tazobactam; specifically false susceptible results.
- Z-0440-2016 · Biomerieux · posted Dec 23, 2015 · Terminated
Due to the potential for reporting erroneous results for Piperacillin/Tazobactam; specifically false susceptible results.
- Z-0441-2016 · Biomerieux · posted Dec 23, 2015 · Terminated
Due to the potential for reporting erroneous results for Piperacillin/Tazobactam; specifically false susceptible results.
- Z-0442-2016 · Biomerieux · posted Dec 23, 2015 · Terminated
Due to the potential for reporting erroneous results for Piperacillin/Tazobactam; specifically false susceptible results.
- Z-0392-2016 · Boston Scientific Corporation · posted Dec 18, 2015 · Terminated
Boston Scientific is voluntarily recalling its Chariot" Guiding Sheath. To date, Boston Scientific has received fourteen complaints for shaft separation, four for distal shaft separation. The most severe outcome of this failure is embolism of device fragments.
- Z-0393-2016 · Insulet Corporation · posted Dec 15, 2015 · Terminated
Pod's needle mechanism fails to deploy or there is a delay in the deployment of the needle mechanism.
- Z-0394-2016 · Insulet Corporation · posted Dec 15, 2015 · Terminated
Pod's needle mechanism fails to deploy or there is a delay in the deployment of the needle mechanism.
- Z-0360-2016 · Draeger Medical, Inc. · posted Dec 12, 2015 · Terminated
There is a potential problem with the device main power switch of the Perseus A500 Anesthesia Workstation. The device power switch may spontaneously fail during use, causing one of the following to occur: 1. The workstation has a fresh gas delivery failure…
Class II recalls (98)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0564-2016 · Biomet 3i, LLC · posted Dec 31, 2015 · Terminated
Images contained within radiographic transparency ART 1036 are undersized.
- RenovoCath RC 120 ↗Class IIZ-0565-2016 · RenovoRx, Inc. · posted Dec 31, 2015 · Terminated
Incorrect version of Instructions for Use were shipped to two US customers.
- Z-0567-2016 · Becton Dickinson & Company · posted Dec 31, 2015 · Terminated
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
- Z-0568-2016 · Becton Dickinson & Company · posted Dec 31, 2015 · Terminated
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
- Z-0569-2016 · Becton Dickinson & Company · posted Dec 31, 2015 · Terminated
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
- Z-0570-2016 · Becton Dickinson & Company · posted Dec 31, 2015 · Terminated
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
- Z-0571-2016 · Becton Dickinson & Company · posted Dec 31, 2015 · Terminated
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
- Granada Biphasic Broth ↗Class IIZ-0556-2016 · Biomerieux Inc · posted Dec 29, 2015 · Terminated
FDA has determined that the product requires 510(k) clearance to continue marketing/distribution in the United States.
- Z-0201-2019 · GE Medical Systems, LLC · posted Dec 28, 2015 · Terminated
GE Healthcare has recently become aware of a potential safety issue with the 3.0T 6 Channel Flex Coil used with the 3T MR750W Surgical Suite Scanners. Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient…
- Z-0553-2016 · GE Medical Systems, LLC · posted Dec 28, 2015 · Terminated
Coil overheating can occur when the device is used in Mode 2 setup. This could lead to a serious patient thermal injury. There have been no injuries reported as a result of this issue.
- Z-0481-2016 · Siemens Medical Solutions USA, Inc · posted Dec 24, 2015 · Terminated
Due to improper soldering in a limited number of large display bypass module components of a specific production lot, a loss of video signal could potentially occur. This has not been observed in the field and only sporadic cases have been observed in…
- Z-0482-2016 · Cardica, Inc. · posted Dec 24, 2015 · Terminated
Cardica, Inc. has received reports that use of the Blue Cartridges with the MicroCutter XCHANGE 30 Stapler was associated with incomplete firing of staples, potentially resulting in an incomplete transection or anastomosis of tissue during the surgical…
- Z-0483-2016 · Cardica, Inc. · posted Dec 24, 2015 · Terminated
Cardica, Inc. has received reports that use of the Blue Cartridges with the MicroCutter XCHANGE 30 Stapler was associated with incomplete firing of staples, potentially resulting in an incomplete transection or anastomosis of tissue during the surgical…
- Cardinal Health Tiny Toes ↗Class IIZ-0543-2016 · Philips Electronics North America Corporation · posted Dec 24, 2015 · Terminated
Heel warmers may cause first and second degree burns.
- Z-0544-2016 · Philips Medical Systems (Cleveland) Inc · posted Dec 24, 2015 · Terminated
Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not…
- Z-0545-2016 · Philips Medical Systems (Cleveland) Inc · posted Dec 24, 2015 · Terminated
Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not…
- Z-0546-2016 · Philips Medical Systems (Cleveland) Inc · posted Dec 24, 2015 · Terminated
Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not…
- Z-0547-2016 · Philips Medical Systems (Cleveland) Inc · posted Dec 24, 2015 · Terminated
Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not…
- Z-0548-2016 · Philips Medical Systems (Cleveland) Inc · posted Dec 24, 2015 · Terminated
Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not…
- Z-0549-2016 · Philips Medical Systems (Cleveland) Inc · posted Dec 24, 2015 · Terminated
Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not…
- Z-0551-2016 · Terumo Medical Corp · posted Dec 24, 2015 · Terminated
Potential improper bonding between the tube and luer for certain lots of Surflo Winged Infusion Set with Filter and Needle Protection (Surshield).
- Z-0552-2016 · Airways Development LLC · posted Dec 24, 2015 · Terminated
Airways Development LLC has received a complaint about a canister leaking from WaterPAP Lot 8605A. During a review with their supplier it could not be determined if this was an isolated incident or if the whole lot is impacted.
- FB Broth, 10ml ↗Class IIZ-0470-2016 · Hardy Diagnostics · posted Dec 23, 2015 · Terminated
The firm is recalling FB Broth Cat. no. K31, Lot no. 15231 because of potential Burkholderia fungorum contamination.
- Vancomycin Screen Agar ↗Class IIZ-0473-2016 · Becton Dickinson & Co. · posted Dec 23, 2015 · Terminated
The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth of vancomycin susceptible…
- Z-0474-2016 · Becton Dickinson & Co. · posted Dec 23, 2015 · Terminated
The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth of vancomycin susceptible…
- Z-0475-2016 · Arthrosurface, Inc. · posted Dec 23, 2015 · Terminated
An oversight in the implementation of the design change record resulted in outdated components being shipped to customers. The component did not pass a half-cycle sterilization.
- Z-0479-2016 · DePuy Mitek, Inc., a Johnson & Johnson Co. · posted Dec 23, 2015 · Terminated
Mitek identified that on VAPR Tripolar 90 Degree Suction Electrode the ablation and coagulation buttons are colored incorrectly. The wiring for each button function was confirmed as correct and the non-conformance impacted button color only
- Z-0480-2016 · Synaptive Medical, Inc. · posted Dec 23, 2015 · Terminated
Out of tolerance for radio frequency emissions. At the 150-1000MHz frequency, the testing indicated the BrightMatter Navigation system was up to 20dB uV/meter higher than the applicable IEC 60601-1-2:2007 (Ed3.0) standard specification.
- Reagent Management System (RMS)Class IIZ-0458-2016 · Siemens Healthcare Diagnostics Inc. · posted Dec 22, 2015 · Terminated
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
- Reagent Management System (RMS)Class IIZ-0459-2016 · Siemens Healthcare Diagnostics Inc. · posted Dec 22, 2015 · Terminated
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
- Reagent Management System (RMS)Class IIZ-0460-2016 · Siemens Healthcare Diagnostics Inc. · posted Dec 22, 2015 · Terminated
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
- Reagent Management System (RMS)Class IIZ-0461-2016 · Siemens Healthcare Diagnostics Inc. · posted Dec 22, 2015 · Terminated
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
- Z-0467-2016 · Siemens Medical Solutions USA, Inc · posted Dec 22, 2015 · Terminated
Possible installation error of the quench lines of superconducting magnets, used to vent helium gas in the rare event of a quench. Any restriction, obstruction, or improper installation of the quench line may cause the helium gas to be blown directly into…
- Z-0468-2016 · Siemens Healthcare Diagnostics, Inc. · posted Dec 22, 2015 · Terminated
Factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Systems. This could lead, under certain circumstances, to an erroneously…
- Z-0469-2016 · Instratek, Incorporated · posted Dec 22, 2015 · Terminated
The sterile blade will not fit into the handle provided with the ETTD system.
- Z-0452-2016 · Zimmer, Inc. · posted Dec 21, 2015 · Terminated
The firm identified a malfunction which made it possible for a handpiece to start by itself when the power source is connected.
- Z-0453-2016 · Zimmer, Inc. · posted Dec 21, 2015 · Terminated
The firm identified a malfunction which made it possible for a handpiece to start by itself when the power source is connected.
- Z-0454-2016 · DePuy Spine, Inc. · posted Dec 21, 2015 · Terminated
Manufacturing error within the Torque Handle may prohibit some handles from being able to fully assemble with the tightener shaft. The design of the inspection gage does not account for the additional shift length
- Z-0455-2016 · CareFusion 303, Inc. · posted Dec 21, 2015 · Terminated
CareFusion is recalling the Alaris EtCO2 module model 8300 because it may have been tested during service using a higher than specified concentration of CO2. The effect to the EtCO2 module tested with a higher concentration of CO2 could lead to an…
- Z-0456-2016 · Becton Dickinson & Company · posted Dec 21, 2015 · Terminated
Some tubes were manufactured with the stopper not fully inserted into the tube. This may result in the stopper popping-off in a small number of instances which could lead to blood exposure and in some cases to blood splatter or spraying.
- Z-0450-2016 · Biomet, Inc. · posted Dec 18, 2015 · Terminated
These screws have a single lead (instead of a triple lead) thread. The screw may sit proud in the plate, If the screw does not lock into the plate; the screw may back out. If the screw sits proud, soft tissue irritation may occur. If the screw does not…
- Z-0437-2016 · Medtronic MiniMed Inc. · posted Dec 17, 2015 · Terminated
The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the…
- Z-0438-2016 · Medtronic MiniMed Inc. · posted Dec 17, 2015 · Terminated
The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the…
- Z-0444-2016 · Myco Medical Supplies Inc · posted Dec 17, 2015 · Terminated
Particulates: Complaints of rubber coring from the closure of injection vials.
- Z-0445-2016 · CareFusion 303, Inc. · posted Dec 17, 2015 · Terminated
CareFusion is recalling the Alaris Pump Module SmartSite Infusion Set because the disconnection and leakage may occur during infusion.
- Z-0412-2016 · Bestmed, LLC · posted Dec 16, 2015 · Terminated
An incorrect calibration of the thermometer that causes the device to display inaccurate and frequently lower temperatures than the actual body temperature of the user.
- Flexible Screw Driver SJ706R ↗Class IIZ-0422-2016 · Aesculap, Inc. · posted Dec 16, 2015 · Terminated
The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.
- Flexible Bone Awl ↗Class IIZ-0423-2016 · Aesculap, Inc. · posted Dec 16, 2015 · Terminated
The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.
- Flexible Drill ↗Class IIZ-0424-2016 · Aesculap, Inc. · posted Dec 16, 2015 · Terminated
The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.
- Z-0432-2016 · Beckman Coulter Inc. · posted Dec 16, 2015 · Terminated
Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but…
- Z-0433-2016 · Beckman Coulter Inc. · posted Dec 16, 2015 · Terminated
Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but…
- Z-0434-2016 · Beckman Coulter Inc. · posted Dec 16, 2015 · Terminated
Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen identification (ID) but…
- Z-0448-2016 · Aesculap, Inc. · posted Dec 16, 2015 · Terminated
The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.
- Z-0419-2016 · Teleflex Medical · posted Dec 15, 2015 · Terminated
Product is being recalled due to complaints that the inspiratory and expiratory tubes are transposed.
- Z-0420-2016 · Verathon, Inc. · posted Dec 15, 2015 · Terminated
Certain models and serial numbers of GlideScope GVL and AVL laryngoscope blades may separate, due to inadequate bonding of the silicone adhesive affixing the halves of the blade. The potential for blade separation may not be readily visible during routine…
- Z-0421-2016 · Verathon, Inc. · posted Dec 15, 2015 · Terminated
Certain models and serial numbers of GlideScope GVL and AVL laryngoscope blades may separate, due to inadequate bonding of the silicone adhesive affixing the halves of the blade. The potential for blade separation may not be readily visible during routine…
- Z-0416-2016 · Cook Inc. · posted Dec 14, 2015 · Terminated
Cook Medical initiated a voluntary recall of specific lot numbers of the VentriClear II Ventricular Drainage Catheter Set due to the device being sterilized twice, which may result in the minocycline and rifampin catheter coating to be less than what is…
- Z-0417-2016 · Stryker Instruments Div. of Stryker Corporation · posted Dec 14, 2015 · Terminated
The Guidewire Introducer Needle (0910-192-000) was shipped after the documented expiration date or with minimal time until expiration. There is a potential risk of a soft tissue infection to the patient if device is used after expiration date.
- Z-0399-2016 · Visicu, Inc. · posted Dec 10, 2015 · Terminated
A software defect may cause incorrect medication order change. If the user decides to edit the order prior to electronic signature AND selects the return to Previous Screen function, the previously selected medication is replaced with whatever medication…
- Z-0400-2016 · Philips Medical Systems (Cleveland) Inc · posted Dec 10, 2015 · Terminated
Brilliance Big Bore- An anomaly was identified with software version 2.0 tumor localization application when printing work sheets. this causes information regarding relative locations of marked regions of interest on the patient work sheets to be…
- Z-0402-2016 · Boston Scientific Corp · posted Dec 10, 2015 · Terminated
Boston Scientific Corp. initiated a voluntary recall of specific lots of the EndoVive One Step Button, including the Decompression Tube and Low Profile Replacement Button devices because they may contain decompression tube adapters with a protrusion that…
- Z-0403-2016 · Olympus Corporation of the Americas · posted Dec 10, 2015 · Terminated
A damaged temperature sensor at the ENDOEYE tip caused the distal end to become abnormally hot. Excessive heating of the ENDOEYE distal end could result in patient or user injury.
- Z-0409-2016 · Philips Medical Systems (Cleveland) Inc · posted Dec 10, 2015 · Terminated
Multiple software defects which could impact the performance or reliability of the system. These impacts include, but are not limited to: error messages which prevent operation, incorrect scan region, couch movement failure, inaccurate CTDI reporting,…
- Z-0410-2016 · Zimmer, Inc. · posted Dec 10, 2015 · Terminated
Complaints that the ratchet teeth of the tube are not latching onto the rod.
- Z-0411-2016 · Zimmer, Inc. · posted Dec 10, 2015 · Terminated
Complaints that the ratchet teeth of the tube are not latching onto the rod.
- Z-0397-2016 · In2Bones, SAS · posted Dec 9, 2015 · Terminated
One batch of osteosynthesis screws were manufactured with the incorrect Torx imprint on the screw head.
- Z-0398-2016 · Edwards Lifesciences, LLC · posted Dec 9, 2015 · Terminated
Edwards Lifesciences initiated a field action for the Thin-Flex Venous Cannula (Model TF292902A) because the connector size is incorrectly described in the Edwards product guide as 3/8 inch. The actual and correct connector size is 1/2 inch.
- Z-0396-2016 · Implant Direct Sybron Manufacturing, LLC · posted Dec 8, 2015 · Terminated
Implant Direct Sybron Manufacturing, LLC is recalling 26 lots of DGDE DirectGuide Drill Extender because of a possible dropping of the drill.
- Z-0391-2016 · BioHorizons Implant Systems Inc · posted Dec 7, 2015 · Terminated
An incorrect label reading 10.5mm and not the specified 15mm was placed on the implant inner vial packaging.
- Z-0371-2016 · Valeant Pharmaceuticals North America LLC · posted Dec 4, 2015 · Terminated
SPAG-2 6000 Lot J3522070 was released to commerce without being tested for particle size per quality control procedure SPAG-2515.
- Z-0372-2016 · Smith & Nephew, Inc. · posted Dec 4, 2015 · Terminated
The devices were packaged and laser marked anatomically incorrect. The Left plates were laser marked and packaged as Right plates and the Right plates were laser marked and packaged as Left plates.
- Z-0373-2016 · Smith & Nephew, Inc. · posted Dec 4, 2015 · Terminated
The devices were packaged and laser marked anatomically incorrect. The Left plates were laser marked and packaged as Right plates and the Right plates were laser marked and packaged as Left plates.
- Z-0374-2016 · Philips Medical Systems, Inc. · posted Dec 4, 2015 · Terminated
Sp02 and/or Non Invasive Blood Pressure (NBP) alarms may become disabled without visual notification
- Z-0375-2016 · Synthes (USA) Products LLC · posted Dec 4, 2015 · Terminated
TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the handle. Or 2. Cause the…
- Z-0378-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0379-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0380-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0381-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0382-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0383-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0384-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0385-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0386-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0387-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0388-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0389-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0390-2016 · LeMaitre Vascular, Inc. · posted Dec 4, 2015 · Terminated
Manufacturing issue that could cause a pinhole in the formed end of the packaging, which may lead to compromised sterility.
- Z-0366-2016 · Merz North America, Inc. · posted Dec 3, 2015 · Terminated
The expiration date encoded in the UDI barcode on the label on the product carton and the peelable patient identification label is incorrect.
- Z-0367-2016 · Merz North America, Inc. · posted Dec 3, 2015 · Terminated
The expiration date encoded in the UDI barcode on the label on the product carton and the peelable patient identification label is incorrect.
- Z-0368-2016 · Merz North America, Inc. · posted Dec 3, 2015 · Terminated
The expiration date encoded in the UDI barcode on the label on the product carton and the peelable patient identification label is incorrect.
- Z-0361-2016 · Quest Medical, Inc. · posted Dec 2, 2015 · Terminated
The products have been found to intermittently exhibit a seal failure during use.
- Z-0362-2016 · Stryker Howmedica Osteonics Corp. · posted Dec 2, 2015 · Terminated
Stryker Orthopaedics has received two customer complaints regarding Duracon Tibial Wedge implants puncturing the packaging's Tyvek lid(s).
- Z-0363-2016 · DeRoyal Industries Inc · posted Dec 2, 2015 · Terminated
These custom sterile surgical kits contain the component, ECOLAB ORS Fluid Warming and Slush Drape, which was recalled due to the possibility that the sterile packaging may contain small channels in the seal of the pouch.
- Z-0355-2016 · Procedure Products, Inc. · posted Dec 1, 2015 · Terminated
The firm had received a complaint alleging that a medical professional had difficulty inserting a guide wire into a needle
- Z-0356-2016 · Baylis Medical Corp * · posted Dec 1, 2015 · Terminated
Microcatheter may have circumferential defects (cracks) along its shaft.
- Z-0357-2016 · Carl Zeiss Meditec AG · posted Dec 1, 2015 · Terminated
IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
- Z-0358-2016 · Carl Zeiss Meditec AG · posted Dec 1, 2015 · Terminated
IOL Master software versions 7.5 and 7.7 calculation printouts and exported reports can contain the wrong IOL power data.
- Z-0359-2016 · ConMed Corporation · posted Dec 1, 2015 · Terminated
Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop…
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0457-2016 · BC Group International Inc · posted Dec 22, 2015 · Terminated
Device is not functioning as intended: Two wires running to the ECG distribution circuit board are reversed resulting in inverted waveforms.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.