Skip to content

FMS neXtra 3.5mm Full Radius Cutter Shaver Blades recallZ-0656-2016 · DePuy Mitek, Inc., a Johnson & Johnson Co. · initiated Nov 20, 2015

Class II Terminated All recalls posted in January 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DePuy Mitek, Inc., a Johnson & Johnson Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product is incorrectly labeled. The blade configuration in the affected lots does not match the description on the label.

FDA root cause category: Process control

What the company is doing

On November 20, 2015, customers were notified of the recall via letter. They were asked to take the following actions: Customer immediate actions: 1. Review inventory located within your facility and quarantine any products listed in this product recall notification. Please follow the instructions on the Business Reply Form to return any affected inventory located within your facility; 2. Review, complete, sign and return the attached business reply form to Elizabeth Messana ([email protected]) in accordance with the directions on the business reply form within 5 business days of receipt of this notification; 3. Forward this notice to anyone in your facility that needs to be informed; 4. If any potentially affected product has been forwarded to another facility, contact Elizabeth Messana at 508-828-3150; 5. Keep a copy of this notice.

Recall details

Recall number
Z-0656-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 20, 2015
Posted by the FDA
Jan 15, 2016
Recalling firm
DePuy Mitek, Inc., a Johnson & Johnson Co.
Quantity
50 units
Product code
HRX
FDA submissions
K041824
Distribution
Worldwide Distribution. US states of TX, OH, MI, and ME; and the countries of FR, DE, SK, IT, NL, CH, and CZ.

Product description

FMS neXtra 3.5mm Full Radius Cutter Shaver Blades; Product Code: 283305 Intended to provide controlled cutting, shaving, and abrading of tissue during orthopedic procedures.

Code information (models, lots, serial numbers)

Lot Number: M1505055

Other devices under the same FDA submission

4 devices cleared under K041824 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 72697).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page