Reagent Management System (RMS) recallZ-0461-2016 · Siemens Healthcare Diagnostics Inc. · initiated Nov 10, 2015
Class II Terminated All recalls posted in December 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens has identified RMS refrigeration compressors that are missing the safety cover on the electrical termination block. The exposed electrical hazard on the electrical termination block is located directly behind the RMS Waste Container.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Siemens sent an Urgent Medical Device Correction (UMDC # DI-16-01.A.US) and Urgent Field Safety Notice (UFSN # DI-16-01.A.OUS) letters dated November 2015 to all affected customers. The letters identified the affected product, problem and actions to be taken. For questions contact your local Siemens technical support representative.
Recall details
- Recall number
- Z-0461-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 10, 2015
- Posted by the FDA
- Dec 22, 2015
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Quantity
- 151
- Product code
- JJE
- FDA submissions
- K112999
- Distribution
- US Nationwide Distribution: FL, NY, SC, TX, LA, AZ, WI, IA, IL, OH, AL, AK, NH, OK, MD, CA, NM, MS, KY, NJ, MN, GA, CO, MA, SD, TN, MO, PA, MT, SC, NE, KS, VA, AL, NV, WY, OR, PR, MO, OK, GU, HI, AR, ME, VT, CT, NC and the countries of Canada and Mexico.
Product description
Reagent Management System (RMS) Dimension RxL Max HM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS doubles the available storage of refrigerated Flex reagent cartridges on the Dimension system from 44 to 88. The RMS loads, hydrates, and automatically removes Flex reagent cartridges, allowing the Dimension system to process samples without interruption.
Code information (models, lots, serial numbers)
Dimension RXL MAX HM STM Catalog Numbers 752000.941 (Domestic)/752000.951 (Intl). RMS for RxL Catalog Numbers 753000.901 (Domestic)/753000.902 (Intl). All serial numbers.
Other devices under the same FDA submission
1 device cleared under K112999 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Dimension® QuikLYTE® Multisensor Na+/K+/Cl- 10445058Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION-SPECIFIC, CHLORIDE34 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 72738).
- Z-0458-2016 · TerminatedReagent Management System (RMS) Dimension EXL with LM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The RMS…
- Z-0459-2016 · TerminatedReagent Management System (RMS) Dimension EXL with LM STM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The…
- Z-0460-2016 · TerminatedReagent Management System (RMS) Dimension RxL Max w/o HM Product Usage: The Reagent Management System (RMS) is a module for use with the Dimension clinical chemistry system. The R…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page