Skip to content

Terumo Surflo Winged Infusion Set with Filter and Needle… recallZ-0551-2016 · Terumo Medical Corp · initiated Nov 23, 2015

Class II Terminated All recalls posted in December 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Terumo Medical Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential improper bonding between the tube and luer for certain lots of Surflo Winged Infusion Set with Filter and Needle Protection (Surshield).

FDA root cause category: Process control

What the company is doing

Terumo N.V. Europe notified their US Importer (Termumo Medical Corporation, Somerset, NJ) of the voluntary recall on 11/23/2015 via phone and email. Terumo Medical Corporation in turn notified their sole consignee on 11/24/2015 by email that included the manufacturers Recall Notification. The notification informed the customer of the reason for the recall; how to identify affected product; to immediately identify and quarantine any available stock; and complete and return the attached Reply Form.

Recall details

Recall number
Z-0551-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 23, 2015
Posted by the FDA
Dec 24, 2015
Recalling firm
Terumo Medical Corp
Quantity
80,500 units
Product code
MEG
FDA submissions
K133894
Distribution
California

Product description

Terumo Surflo Winged Infusion Set with Filter and Needle Protection (Surshield), Product Code: SV-S25FL35.

Code information (models, lots, serial numbers)

Product Code: SV-S25FL35 Lot Numbers 1508023, 1509011 and 1510004.

Other devices under the same FDA submission

1 device cleared under K133894 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page