BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x recallZ-0567-2016 · Becton Dickinson & Company · initiated Dec 3, 2015
Class II Terminated All recalls posted in December 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
FDA root cause category: Process change control
What the company is doing
The firm, BD, issued an "URGENT PRODUCT RECALL" notifications/Recall Response Forms to their customers dated 12/3/2015 via email and/or UPS 2nd Day Mail. The notification identified the reason for the recall; how to identify affected product; and actions to be taken. The customers were instructed customers to immediately review their inventory; complete the enclosed Recall Response Form and fax (1-866-873-0312 it to BD or email it to [email protected]; and return all affected products with the completed Recall Response Form following instructions on the enclosed packing instruction, even if you do not have any affected lot in your inventory. If customers have any questions or require assistance with the return of the recalled product and/or availability of replacement product, they were instructed to contact 1-888-628-0732 between 8AM and 5 PM ET Monday through Friday.
Recall details
- Recall number
- Z-0567-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 3, 2015
- Posted by the FDA
- Dec 31, 2015
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 29,815,600 units in total
- Product code
- JKA
- FDA submissions
- K981013
- Distribution
- Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.
Product description
BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 2.0 mL (Pink) BD Hemogard Additive: K2EDTA (spray dried), 3.6mg Sterile IVD BD, Made in USA The Vacutainer Brand PLUS (Plastic) Tube with EDTA Vacutainer Brand Serum Tube are evacuated blood collection tubes which provide a means of collecting, transporting, separating and processing blood in a plastic tube.
Code information (models, lots, serial numbers)
Catalog Reference #367842; Lot numbers 5089731 and 5175923
Other devices under the same FDA submission
1 device cleared under K981013 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BD Vacutainer® Serum Blood Collection Tubes 367815Becton, Dickinson And Company · Tubes, vials, systems, serum separators, blood collection462 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 72868).
- Z-0568-2016 · TerminatedBD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 2.0 mL (Pink) BD Hemogard Additive: K2EDTA (spray dried), 3.6mg Sterile IVD BD, Made in USA BD Vacutainer tubes are used …
- Z-0569-2016 · TerminatedBD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 3.0 mL (Lavender) BD Hemogard Additive: K2EDTA (spray dried), 5.4mg Sterile IVD BD, Made in USA BD Vacutainer tubes are u…
- Z-0570-2016 · TerminatedBD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 3.0 mL (Tan) BD Hemogard Additive: K2EDTA (spray dried), 5.4mg Sterile IVD BD, Made in USA BD Vacutainer tubes are used t…
- Z-0571-2016 · TerminatedBD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 4.0 mL (Lavender) BD Hemogard Additive: K2EDTA (spray dried), 7.2mg Sterile IVD BD, Made in USA BD Vacutainer tubes are u…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page