Emergency Transport Ventilator recallZ-0691-2016 · Draeger Medical, Inc. · initiated Dec 22, 2015
Class I Terminated All recalls posted in February 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
FDA root cause category: Device Design
What the company is doing
The firm, Draeger Medical, Inc., sent an "Urgent Medical Device Recall" letter dated 12/21/2015 and a copy of the Instructions for Use (IFU) Supplement to consignees in a method that confirms delivery. The letter described the product, problem and actions to be taken. The consignees were instructed to include a copy of the supplement with their Oxylog IFU; inform all affected users in their facility; and contact Draeger Customer Support at 1-800-543-5047 (press 4 at the prompt) to obtain additional copies of the IFU Supplement Free of Charge. If there are any questions regarding this letter, please contact Michael Kelhart at 1-800-543-5047 (press 1 at the prompt, then 2, then 32349). For questions regarding the operation and/or servicing of your Drger Oxylog ventilator please contact DrgerService Technical Support at 1-800-543-5047 (press 4 at the prompt). On 12/09/2016, Draeger mailed an Urgent Medical Device Recall notification letter to affected customers to inform them of a revised new software for the Oxlog 3000/3000plus (version 1.06 for Oxylog 3000 plus and version 1.23 for Oxylog 3000) that reduces the impact of the error condition. Customers will be contacted by a Drager Service Representative to schedule a time to perform the software update free of charge. The new software changes the device behavior in the event of the error (a controller failure). With the new software, in the event of a controller failure, ventilation will not stop. The device will continue to ventilate with the last setting. An alarm will annunciate and be displayed, alerting the user of the controller failure.
Recall details
- Recall number
- Z-0691-2016
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Dec 22, 2015
- Posted by the FDA
- Feb 3, 2016
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 0
- Product code
- CBK
- FDA submissions
- K062267, K082600, K103625
- Distribution
- Nationwide Distribution -- including states of AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Product description
Emergency Transport Ventilator Oxylog 2000 Plus Ventilator Catalog Number: 5705080. Volume controlled emergency and transport ventilator with pressure support.
Code information (models, lots, serial numbers)
All Oxylog 2000 plus Ventilators with the identified catalog numbers distributed between April 2007 - December 2015.
Other devices under the same FDA submission
1 device cleared under K062267, K082600, K103625 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Oxylog 3000plus 5704833Drägerwerk AG & Co. KGaA · Ventilator, Continuous, Facility Use4 deaths58 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 72914).
- Z-0692-2016 · TerminatedEmergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000 plus Ventilator Catalog Numbers: 2M86300/2M86965 and 5704813/5704831. Volume controlled and pressure-controll…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page