Medical device recalls: February 2016
The FDA posted 136 medical device recall records in February 2016, from 66 recall events by 54 companies. 9 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips Electronics North America Corporation · 4 events, 7 records
- Cook Inc. · 3 events, 6 records, 5 Class I
- Bard Peripheral Vascular Inc · 2 events, 3 records
- Draeger Medical, Inc. · 2 events, 3 records, 3 Class I
- Siemens Healthcare Diagnostics, Inc. · 2 events, 3 records
- C.R. Bard, Inc. · 2 events, 2 records
- Leica Microsystems, Inc. · 2 events, 2 records
- Siemens Medical Solutions USA, Inc · 2 events, 2 records
FDA root cause
- Process control · 36
- Under Investigation by firm · 27
- Labeling Change Control · 14
- Device Design · 14
- Software design · 9
- Nonconforming Material/Component · 8
- Error in labeling · 6
- Component design/selection · 4
Class I recalls (9)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0827-2016 · Draeger Medical, Inc. · posted Feb 25, 2016 · Terminated
The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015…
- Z-0798-2016 · Cook Inc. · posted Feb 24, 2016 · Terminated
Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result. Recall expanded 10/07/2015 to…
- Z-0799-2016 · Cook Inc. · posted Feb 24, 2016 · Terminated
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding…
- Z-0800-2016 · Cook Inc. · posted Feb 24, 2016 · Terminated
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding…
- Z-0801-2016 · Cook Inc. · posted Feb 24, 2016 · Terminated
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding…
- Z-0802-2016 · Cook Inc. · posted Feb 24, 2016 · Terminated
A manufacturing process may lead to catheter tip fracture and or separation. Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding…
- Flow-i Anesthesia SystemClass IZ-0804-2016 · Maquet Cardiovascular Us Sales, Llc · posted Feb 22, 2016 · Terminated
Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system…
- Z-0691-2016 · Draeger Medical, Inc. · posted Feb 3, 2016 · Terminated
Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
- Z-0692-2016 · Draeger Medical, Inc. · posted Feb 3, 2016 · Terminated
Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
Class II recalls (124)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1026-2016 · Ivoclar Vivadent, Inc. · posted Feb 29, 2016 · Terminated
Complaints were received claiming the material failed to polymerize under the curing light. After investigation, it was determined that a wrong photo-initiator was used in the production batch U30755 of Helioseal Clear. This failure can lead to…
- Z-1028-2016 · Cook Inc. · posted Feb 29, 2016 · Terminated
Product potentially exposed to body fluids during processing. Potential adverse events that may occur as a result of the foreign matter include blood borne pathogen transmission. However, the risk of transmission of blood borne pathogens is thought to be…
- Z-1024-2016 · Philips Electronics North America Corporation · posted Feb 26, 2016 · Terminated
Wrong positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to coupling to the QBC. The combined use of the Sense coil, high SAR level scanning and placement of the cables contrary to…
- Z-0867-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0868-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0869-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0870-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0871-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0872-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Stryker, T2 K-Wire, 3x285 mm ↗Class IIZ-0873-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0874-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0875-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0876-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0877-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0878-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0879-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0880-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0881-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0882-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0883-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0884-2016 · Stryker Howmedica Osteonics Corp. · posted Feb 25, 2016 · Terminated
During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased.
- Z-0885-2016 · Biomet, Inc. · posted Feb 25, 2016 · Terminated
A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked.
- Z-0886-2016 · Teleflex Medical · posted Feb 25, 2016 · Terminated
The clip in the applier may be missing a boss. A boss is the part of the clip that is used to hold the clip within the applier jaws prior to application.
- Z-0887-2016 · Fresenius Medical Care Renal Therapies Group, LLC · posted Feb 25, 2016 · Terminated
On the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of constituents represents the actual calcium concentration of the…
- Z-1017-2016 · The Anspach Effort, Inc. · posted Feb 25, 2016 · Open, Classified
May operate solely in reverse mode, not operate in reverse mode when intended, not function or an unintentionally start.
- Z-0753-2016 · Regulatory Insight, Inc · posted Feb 24, 2016 · Terminated
It was discovered that the Remote Touch Panel (RTP) of the GE Precision MPi X-ray system may not always boot up as intended and needs to be updated to properly accomplish its intended purpose.
- Z-0864-2016 · EKOS Corporation · posted Feb 24, 2016 · Terminated
The EkoSonic kit, including the individually labeled MicroSonic Device (MSD) and Intelligent Drug Delivery Catheter (IDDC) are labeled with a 135 cm working length, and programmed with a 135 cm family code, however the MSD actual working length is 106 cm.
- Z-0856-2016 · Philips Electronics North America Corporation · posted Feb 23, 2016 · Terminated
Reconstructed ECG leads viewed or printed at the Information Center iX may misrepresent the ECG waveform in specific leads.
- Z-0857-2016 · Philips Electronics North America Corporation · posted Feb 23, 2016 · Terminated
Reconstructed ECG leads viewed or printed at the Information Center iX may misrepresent the ECG waveform in specific leads.
- Z-0858-2016 · Philips Electronics North America Corporation · posted Feb 23, 2016 · Terminated
Reconstructed ECG leads viewed or printed at the Information Center iX may misrepresent the ECG waveform in specific leads.
- Z-0860-2016 · Hill-Rom, Inc. · posted Feb 23, 2016 · Terminated
The recliners were shipped without the foot rest ottoman linkage side shields, which could result in consumer injury.
- Z-0861-2016 · Centurion Medical Products Corporation · posted Feb 23, 2016 · Terminated
Convenience Kits contain an In-Vision-Plus Needleless IV Connector, code RYM5001 that is being recalled by the manufacturer, RyMed Technologies, LLC. A component molding abnormality may cause female luer threads to crack, leading to potential for the…
- Z-0862-2016 · Siemens Medical Solutions USA, Inc · posted Feb 23, 2016 · Terminated
Siemens is releasing an updated software version to address several software issues including RGB images will show "?" since calculation of HU is not possible; save as option enabled; changes in access for loading studies; breast region is now properly…
- LPS-FLEX GSF OPTClass IIZ-0830-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- NexGen CR-FLEX Option FEMClass IIZ-0831-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- NexGen LPS-FLEX Option FemoralClass IIZ-0832-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- NexGen Knee CR Porous FEM ↗Class IIZ-0833-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- NexGen Knee LPS Porous FEMClass IIZ-0834-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- NexGen Knee CR Option FemClass IIZ-0835-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- ZIMMER UNI HIGH FLEX PRECOAT FEMClass IIZ-0836-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- Gender PFJ FEMORAL COMP,Class IIZ-0837-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- CR Flex GSF Precoat FemClass IIZ-0838-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- NKII CSTI Porous FemoralClass IIZ-0839-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- CR Flex GSF Porous FemoralClass IIZ-0840-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- NexGen LPS Flex Option FemoralClass IIZ-0841-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- PRCT MIS CR MOB PLT ↗Class IIZ-0842-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- CPT 12/14 COCRClass IIZ-0843-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- CPT 12/14 COCRClass IIZ-0844-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- Bipolar ShellClass IIZ-0845-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- VERSYS 6 INCH BEADED FC 1 ↗Class IIZ-0846-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- VERSYS 7.5 INCH BEADED FC ↗Class IIZ-0847-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- VERSYS 6 INCH BEADED FC 1Class IIZ-0848-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- VERSYS 6 INCH BEADED FC 1Class IIZ-0849-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- Metasul Head 40, 12/14, SZ M/0 ↗Class IIZ-0850-2016 · Zimmer Biomet, Inc. · posted Feb 22, 2016 · Terminated
LDPE bag containing the implant adheres to the highly polished implant surface.
- Z-0823-2016 · Leica Microsystems, Inc. · posted Feb 19, 2016 · Terminated
The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining when using a detection protocol where one of its components - Peroxide Block / Peroxidase Block is applied after the primary marker incubation as…
- Z-0824-2016 · Siemens Healthcare Diagnostics, Inc. · posted Feb 19, 2016 · Terminated
False short clotting times for PT on BCS and BCS XP with Dade Innovin can occur. This has also an effect for false low PT INR values. These false short clotting times were observed only with turbid, hemolytic and/or icteric samples from intensive care…
- Z-0825-2016 · Siemens Healthcare Diagnostics, Inc. · posted Feb 19, 2016 · Terminated
False short clotting times for PT on BCS and BCS XP with Dade Innovin can occur. This has also an effect for false low PT INR values. These false short clotting times were observed only with turbid, hemolytic and/or icteric samples from intensive care…
- software for Syngo Dynamics ↗Class IIZ-0826-2016 · Siemens Medical Solutions USA, Inc · posted Feb 19, 2016 · Terminated
Siemens is releasing a software update that addresses an issue of mixing data from multiple patients. In rare situations, echo trend graphs may mix data from multiple patients.
- Z-0819-2016 · Medtronic · posted Feb 18, 2016 · Terminated
Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
- Z-0820-2016 · Medtronic · posted Feb 18, 2016 · Terminated
Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
- Z-0821-2016 · Medtronic · posted Feb 18, 2016 · Terminated
Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
- Best Practice Kits containing certain production lots of Endo GIA Ultra Universal Stapler HandlesClass IIZ-0822-2016 · Medtronic · posted Feb 18, 2016 · Terminated
Staplers fail to fire or partially fire and reports of the instrument articulating lever disengaging during use
- Z-0817-2016 · Nihon Kohden America Inc · posted Feb 17, 2016 · Terminated
The MEE-1000A Neuromaster had defective printed circuit boards (PCBs). Gas arrestors on these boards which were intended to provide protection from electrostatic discharge failed to prohibit unintentional current flow.
- Z-0810-2016 · Philips Electronics North America Corporation · posted Feb 16, 2016 · Terminated
The actuator assembly became detached and the monitor carriage with a flexvision monitor dropped to its lowest position.
- Z-0813-2016 · Sorin Group USA, Inc. · posted Feb 16, 2016 · Terminated
Sorin Group initiated field correction for the Dual Stage Venous Return Catheter because the incorrect Instructions For Use (IFU) was packaged in the case.
- Z-0814-2016 · CareFusion 303, Inc. · posted Feb 16, 2016 · Terminated
Customers reported issues when scanning medications with the scanners. When a user scans a medication using the affected barcode scanner, information on a different medication could be communicated to the Pyxis product.
- Z-0807-2016 · Bio-Rad Laboratories, Inc. · posted Feb 13, 2016 · Terminated
The error "Washer Reagent Clean Fluid" level low alarmed, which prompted the customer to open the system drawer. When the system drawer was opened the customer noted the plate transport was still running and as a consequence the plate was pushed off the…
- Z-0809-2016 · Exactech, Inc. · posted Feb 13, 2016 · Terminated
Incorrect external and internal packaging labels. Exactech has identified nineteen (19) Acapella One Interbody Fusion Device Lordotic (05-090-01-1610) were packaged as Acapella One Interbody Fusion Device Parallel (05-090-02-1610)
- Z-0805-2016 · Varian Medical Systems, Inc. · posted Feb 12, 2016 · Terminated
When using PBC 11.0.31 to calculate the dose for a conventional arc field with more than 100 segments for Eclipse versions 11.0, 13.0, 13.5 or 13.6, the displayed dose does not correspond to the calculated Monitor Units (MU). Potential for unintended…
- Z-0806-2016 · Viewray Incorporated · posted Feb 12, 2016 · Terminated
The system loaded a completion fraction in the incorrect order after a treatment interruption.
- Z-0808-2016 · Leica Microsystems, Inc. · posted Feb 12, 2016 · Terminated
The Firm informed the importer/distributor, Leica Microsystems, a manufacturing error occurred during the filling of some containers of the Bond Polymer Refine Red Detection DS9390 kits, which may result in inadequate staining.
- Z-0688-2016 · MOOG Medical Devices Group · posted Feb 11, 2016 · Terminated
Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015 because the master pumps used to calibrate them were out of tolerance and may cause an over delivery of…
- Z-0797-2016 · Orthofix, Inc · posted Feb 11, 2016 · Terminated
There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization. In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical…
- Z-0795-2016 · Pharmacaribe llc · posted Feb 10, 2016 · Terminated
Vials were labeled as USP 7% Hypertonic saline instead of PulmoSal 7% (pH+)
- Z-0796-2016 · Salter Labs · posted Feb 10, 2016 · Terminated
Reports that the T-piece does not fit on the nebulizer.
- Z-0752-2016 · Toshiba American Medical Systems Inc · posted Feb 9, 2016 · Terminated
When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defined the data as an "abnormal value"…
- CyberKnife Robotic ↗Class IIZ-0771-2016 · Accuray Incorporated · posted Feb 9, 2016 · Terminated
The gun box mounting bracket may fail to support the weight of the gun box when in vertical (inverted) position. If this failure occurs the gun box may become loose and could come into contact with a patient.
- IMPLANT,TM, MTX ↗Class IIZ-0772-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX ↗Class IIZ-0773-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX ↗Class IIZ-0774-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX ↗Class IIZ-0775-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX ↗Class IIZ-0776-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX ↗Class IIZ-0777-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX FULL ↗Class IIZ-0778-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX FULL ↗Class IIZ-0779-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX FULL ↗Class IIZ-0780-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX FULL ↗Class IIZ-0781-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX FULL ↗Class IIZ-0782-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX FULL ↗Class IIZ-0783-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- IMPLANT,TM, MTX FULL ↗Class IIZ-0784-2016 · Zimmer Dental Inc · posted Feb 9, 2016 · Terminated
Information specific to the 3.7 mmD and 4.1 mmD Trabecular Metal Zimmer Dental Implants was omitted from the latest revision of the applicable IFU in the Indications, Contraindications, and Precautions sections.
- Z-0785-2016 · Bard Peripheral Vascular Inc · posted Feb 9, 2016 · Terminated
Size mislabeling; the graft is labeled as a 6 mm diameter graft when the actual packaged product is an 8 mm graft.
- Z-0786-2016 · Bard Peripheral Vascular Inc · posted Feb 9, 2016 · Terminated
Size mislabeling; the graft is a 6 mm diameter graft when the actual packaged product is an 8 mm graft.
- Z-0787-2016 · DePuy Orthopaedics, Inc. · posted Feb 9, 2016 · Terminated
Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin. Surgical delay may result if the punch dislodges from the inserter and needs to be removed from…
- Z-0788-2016 · Biomet France SARL · posted Feb 9, 2016 · Terminated
Cytotoxicity tests were not passed at 36 months. Use of Endobon-Xenograft Granules past 18 months of its specified shelf life may result in patient illness/injury.
- Z-0789-2016 · Richard Wolf Medical Instruments Corp. · posted Feb 9, 2016 · Terminated
Weak weld in the reusable rotary blade/abrader (burr) may result in inner part breaks at the distal or proximal end making the product dysfunctional.
- Z-0790-2016 · Richard Wolf Medical Instruments Corp. · posted Feb 9, 2016 · Terminated
Weak weld in the reusable rotary blade/abrader (burr) may result in inner part breaks at the distal or proximal end making the product dysfunctional.
- Z-0791-2016 · Richard Wolf Medical Instruments Corp. · posted Feb 9, 2016 · Terminated
Weak weld in the reusable rotary blade/abrader (burr) may result in inner part breaks at the distal or proximal end making the product dysfunctional.
- Mizuho OSI Trios, Table Base ↗Class IIZ-0792-2016 · Mizuho OSI · posted Feb 9, 2016 · Terminated
Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly. Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult. If the clutch cannot disengage, the user would be…
- Z-0767-2016 · Mevion Medical Systems, Inc. · posted Feb 8, 2016 · Terminated
The treatment table positioning belt can slip gear positions, resulting in deviation of patient setup and treatment positions from the time of slippage until the discovery of slippage and the halting of treatment.
- Z-0768-2016 · GE Medical Systems, LLC · posted Feb 8, 2016 · Terminated
Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
- Z-0769-2016 · GE Medical Systems, LLC · posted Feb 8, 2016 · Terminated
Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
- Z-0770-2016 · GE Medical Systems, LLC · posted Feb 8, 2016 · Terminated
Potential safety latch mechanism failure. The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.
- Z-0758-2016 · Spectranetics Corp. · posted Feb 5, 2016 · Terminated
The product is labeled with an expiration date that is past its shelf life.
- Z-0760-2016 · C.R. Bard, Inc. · posted Feb 5, 2016 · Terminated
Improper targeted temperature therapy was delivered to patients because the Temperature Simulator Key was connected to the patient temperature cable instead of the patient temperature probe.
- Z-0764-2016 · Philips Electronics North America Corporation · posted Feb 5, 2016 · Terminated
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced…
- Z-0765-2016 · Philips Electronics North America Corporation · posted Feb 5, 2016 · Terminated
Normally, movement of the C-arc is initially stopped by the motor (holding torque) then by the electromagnetic brake, leading to a full stop of the C-arc. Because of improper adjustment and tolerances, when the C-arc is initially stopped in an unbalanced…
- ADVIA Centaur PSA Assay ↗Class IIZ-0754-2016 · Siemens Healthcare Diagnostics, Inc · posted Feb 4, 2016 · Terminated
PSA assay is not meeting the High Dose Hook Effect expectation in the Instructions for Use (IFU).
- Z-0755-2016 · Siemens Healthcare Diagnostics, Inc. · posted Feb 4, 2016 · Terminated
Multiple Software issues. Auto Start-UP Fail, Calibration Interval Resets when a Reagent Blank is run, Control Definition screen assumes range defined is 2 SD, Laboratory Information System (LIS) communication/Laboratory Automation (LAS) issue, Printer…
- API ZYM Bx2 (Ref 70493) ↗Class IIZ-0756-2016 · BioMerieux SA · posted Feb 4, 2016 · Terminated
Discrepant identification results related to API identification strips. These discrepant identification results are due to false negative results on the tests of the strips linked to the ZYM B reagent.
- Z-0757-2016 · Olympus Corporation of the Americas · posted Feb 4, 2016 · Terminated
Olympus America Inc is conducting a voluntary removal/corrective action of all TJF-Q180V duodenoscopes in order to replace the forceps elevator mechanism Olympus is replacing the forceps elevator mechanism with a new forceps elevator design consistent with…
- Z-0696-2016 · Global Healthcare Inc · posted Feb 3, 2016 · Terminated
Firm was notified of a potential product defect related to kinking (flattening) of anesthesia extension tubes which may represent a potential health hazard.
- Astral 100, 150 ↗Class IIZ-0746-2016 · Resmed Corporation · posted Feb 3, 2016 · Terminated
Resmed Corporation is recalling Astral 100, 150 External Battery because of a start-up issue.
- Z-0747-2016 · The Anspach Effort, Inc. · posted Feb 3, 2016 · Terminated
The affected Small Electric Drive (SED) may operate solely in reverse mode, not operate in reverse mode when intended, not function, or unintentionally start.
- AQURE System ↗Class IIZ-0748-2016 · Radiometer America Inc · posted Feb 3, 2016 · Terminated
The AQURE System has a design error regarding sample type in which sample type may be specified for some results and left blank for other results. The fact that the sample type field is left blank in some cases could lead to misinterpretation of results.
- Z-0738-2016 · Horiba Instruments Inc · posted Feb 2, 2016 · Terminated
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents…
- Z-0739-2016 · Horiba Instruments Inc · posted Feb 2, 2016 · Terminated
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents…
- Z-0740-2016 · Horiba Instruments Inc · posted Feb 2, 2016 · Terminated
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents…
- Z-0741-2016 · Horiba Instruments Inc · posted Feb 2, 2016 · Terminated
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents…
- Z-0742-2016 · Horiba Instruments Inc · posted Feb 2, 2016 · Terminated
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents…
- Z-0743-2016 · Horiba Instruments Inc · posted Feb 2, 2016 · Terminated
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents…
- Z-0744-2016 · Acist Medical Systems · posted Feb 2, 2016 · Terminated
Reports related to ingress of air into the manifold.
- Z-0745-2016 · Philips Medical Systems · posted Feb 2, 2016 · Terminated
In DS/US proton planning, there is a correction factor used internally to the dose engine that is calculating improperly in some cases. This can result in an incorrect dose calculation.
- Z-0736-2016 · Instrumentation Laboratory Co. · posted Feb 1, 2016 · Terminated
This Lot is not meeting labeled on-board instrument stability claims of 5 days.
- Z-0737-2016 · Bard Peripheral Vascular Inc · posted Feb 1, 2016 · Terminated
IFU missing contraindications: Patients with uncontrolled sepsis and Patients with known hypersensitivity to nickel-titanium alloys.
Class III recalls (3)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1023-2016 · Medtronic Perfusion Systems · posted Feb 26, 2016 · Terminated
Medtronic is notifying customers who have or may receive specific lots of Medtronic Perfusion Tubing Packs that include the potentially affected cuvettes manufactured by Terumo Cardiovascular Systems (Terumo CVS). Terumo Cardiovascular Systems (Terumo…
- Z-0815-2016 · Lab Vision Corporation · posted Feb 17, 2016 · Terminated
Certain lots of CEA/CD66e Ab2 were produced with a dilution factor of 1:100 instead of the recommended 1:250.
- Z-0812-2016 · C.R. Bard, Inc. · posted Feb 16, 2016 · Terminated
Report by a customer of an open package seal.