Why it was recalled

N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.

FDA root cause category: Under Investigation by firm

What the company is doing

Customer notification letters were sent on 01/11/16. Customers were provided the following instructions/actions: When performing tests that use the Tinder reaction, the blood draw should be done before treatment is given to the patient. Reagent inserts for the concerned HORIBA Medical Products have been updated and now include the following additional warning: Patients treated with N-Acetylcysteine (NAC) for Paracetamol overdose may generate a false low result. The letter request customers share the information with laboratory staff and retain the notification as part of their laboratory documentation. The letter states it is mandatory for customers to complete and return the enclosed response form within 10 days in order for the firm to verify the understanding of the notice. If any questions contact the local HORIBA Medical representative.

Recall details

Recall number
Z-0740-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 11, 2016
Posted by the FDA
Feb 2, 2016
Recalling firm
Horiba Instruments Inc
Quantity
216
Product code
CFR
FDA submissions
K052007
Distribution
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.

Product description

ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.