Evita Infinity V500 Ventilator recallZ-0827-2016 · Draeger Medical, Inc. · initiated Feb 4, 2016

Class I Terminated All recalls posted in February 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

FDA root cause category: Software design

What the company is doing

Draeger Medical, Inc. will send an Urgent Medical Device Recall Letter to consignees in a method that confirms delivery. For consignees that have received new devices with firmware version 1.50 (127 devices) or through device updates (644 devices; reference Recall Z-0436-2016), those devices will be downgraded to firmware version 1.49 and the batteries replaced. Batteries will continue to be replaced for all consignees every six months until a final solution is available.

Recall details

Recall number
Z-0827-2016
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Feb 4, 2016
Posted by the FDA
Feb 25, 2016
Recalling firm
Draeger Medical, Inc.
Quantity
2553
Product code
CBK
FDA submissions
K093632, K093633
Distribution
US Nationwide Distribution including Puerto Rico.

Product description

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Code information (models, lots, serial numbers)

Catalog #s 8416400 and 8417400

Other devices under the same FDA submission

2 devices cleared under K093632, K093633 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates