Percutaneous Insertion Tray recallZ-1063-2016 · Arrow International, Inc., Division of Teleflex Medical Inc. · initiated Feb 10, 2016
Class I Terminated All recalls posted in March 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, Inc., Division of Teleflex Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The sheath body may become separated from the sheath hub. If the separation occurs, there is a potential for bleeding from the device.
FDA root cause category: Process control
What the company is doing
The firm, Teleflex Arrow International, sent an "Urgent Medical Device Recall Notification" letter dated 2/11/2016 via FedEx to its Customers. The letter described the product, problem and actions to be taken. The customers were informed of the reason for recall and were instructed to take the following actions: 1. If you have any products referenced in the attached enclosure, immediately discontinue use and quarantine those products appropriately. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507 Attn: Customer Service or email it to [email protected]. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of product to Arrow International. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email it to [email protected]. If you have any other questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990.
Recall details
- Recall number
- Z-1063-2016
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Feb 10, 2016
- Posted by the FDA
- Mar 11, 2016
- Recalling firm
- Arrow International, Inc., Division of Teleflex Medical Inc.
- Quantity
- 13,405 US and 33,735 OUS in total
- Product code
- DSP
- FDA submissions
- K981660
- Distribution
- Worldwide Distribution: US (Nationwide) and countries of: Argentina, Australia, Belgium, Brunei Darussalam, Brazil, Bahamas, Canada, Chile, China, Colombia, Ecuador, Guatemala, Hong Kong, Indonesia, India, Japan, Korea (Republic of [South] Korea), Myanmar, Mexico, Malaysia, Nepal, New Zealand, Panama, Peru, Philippines, Pakistan, Singapore, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Viet Nam, and South Africa (Zuid Afrika).
Product description
Percutaneous Insertion Tray; Product Code: IAK-S7IT The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diastole and deflation, during systole increases blood supply to the heart muscle and decreases work of the left ventricle.
Code information (models, lots, serial numbers)
Batch: 18F14A0037 18F14B0040 18F14E0027 18F14F0027 18F14F0028
Other devices under the same FDA submission
9 devices cleared under K981660 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ARROW IPN915325Teleflex Incorporated · System, balloon, intra-aortic and control35 reports
- ARROW IPN038265Teleflex Incorporated · System, balloon, intra-aortic and control16 reports
- ARROW IPN039783Teleflex Incorporated · System, balloon, intra-aortic and control16 reports
- ARROW IPN038077Teleflex Incorporated · System, balloon, intra-aortic and control13 reports
- ARROW IPN039784Teleflex Incorporated · System, balloon, intra-aortic and control12 reports
- ARROW IPN915324Teleflex Incorporated · System, balloon, intra-aortic and control9 reports
- ARROW IPN039780Teleflex Incorporated · System, balloon, intra-aortic and control4 reports
- ARROW IPN915328Teleflex Incorporated · System, balloon, intra-aortic and control2 reports
- ARROW IPN915327Teleflex Incorporated · System, balloon, intra-aortic and control1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 73266).
- Z-1056-2016 · TerminatedUltra 8 IAB: 8Fr 30cc; Product Code: IAB-05830-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diasto…
- Z-1058-2016 · TerminatedUltra 8 IAB: 8Fr 40cc; Product Code: IAB-05840-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diasto…
- Z-1059-2016 · TerminatedUltraFlex IAB: 7.5Fr 30cc; Product Code: IAB-06830-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during di…
- Z-1060-2016 · TerminatedUltraFlex IAB: 7.5Fr 40cc; Product Code: IAB-06840-U The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during di…
- Z-1061-2016 · TerminatedRediGuard IAB: 7Fr 30cc; Product Code: IAB-S730C The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during diasto…
- Z-1062-2016 · TerminatedPercutaneous Insertion Tray; Product Code: IAK-06845 The Arrow IAB is utilized for intra aortic balloon counterpulsation therapy in the aorta, whereby balloon inflation, during di…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.