AXIOM Artis zee/zeego, system, x-ray, angiographic recallZ-1283-2016 · Siemens Medical Solutions USA, Inc · initiated Feb 18, 2016
Class II Terminated All recalls posted in March 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a leak in the cooling system, fluids could potentially seep into the equipment cabinet. Loss of coolant can result in a cooling system failure and subsequent system failure. This leak can occur sporadically in affected systems.
FDA root cause category: Unknown/Undetermined by firm
What the company is doing
Siemens sent an Important Customer Safety Notice to customers on February 18, 2016 informing them that their service organization will contact them to arrange a date to perform the corrective action (replacement of affected plastic couplings with metal couplings to prevent fluid leaks). Customers were advised to contact Siemens service organization for an earlier appointment at 1-800-888-7436.
Recall details
- Recall number
- Z-1283-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 18, 2016
- Posted by the FDA
- Mar 31, 2016
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 986 systems in total
- Product code
- OWB
- FDA submissions
- K123529
- Distribution
- US Nationwide Distribution in the states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY including Puerto Rico.
Product description
AXIOM Artis zee/zeego, system, x-ray, angiographic Product Usage: AXIOM Artis zee and Q/Q.zen is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis zee/zeego and Q/Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis zee/zeego and Q/Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Code information (models, lots, serial numbers)
Serial Numbers: Artis Q biplane Serial # 121061, 121089, 121094, 121006, 121013, 121040, 121062, 121082, 121011, 121039, 121134, 121053, 121054, 121014, 121112, 121051, 121004, 121133, 121018, 121036, 121137, 121012, 121049, 121078, 121019, 121010, 121139, 121081, 121096, 121015, 121115, 121016, 121138, 121020, 121140, 121106, 121120, 121122, 121121, 121109, 121144, 121047, 121088, 121027, 121136, 121141, 121069, 121009, 121057, 121056, 121034, 121008, 121058, 121085, 121071, 121023, 121079, 121041, 121037, 121046, 121072
Other devices under the same FDA submission
7 devices cleared under K123529 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Artis zeego 10848283Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system5 recalls1 death7 reports
- Artis Q ceiling 10848281Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls1 death49 reports
- Artis Q biplane 10848282Siemens Healthcare GmbH · System, x-ray, fluoroscopic, image-intensified4 recalls1 death46 reports
- Artis Q floor 10848280Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls18 reports
- Artis Q.zen ceiling 10848354Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls10 reports
- Artis Q.zen floor 10848353Siemens Healthcare GmbH · System, x-ray, angiographic4 recalls5 reports
- Artis Q.zen biplane 10848355Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system4 recalls5 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page