ADVIA Centaur CP Immunoassay System, Siemens Material… recallZ-1615-2016 · Siemens Healthcare Diagnostics, Inc. · initiated Mar 30, 2016

Class III Terminated All recalls posted in May 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm identified an issue in sample processing when using EZee-Nest Insert Cups due to two inaccuracies that exist in the ADVIA Centaur CP Operators Guide.

FDA root cause category: Use error

What the company is doing

Siemens sent an Urgent Medical Device Correction letter dated March 2016, to all affected customers. The letter was sent to the US customers via Fed Ex beginning on March 30, 2016., and were e-mailed dated March 20, 2016, to regional countries for implementation outside the US. Field service personnel were sent a support bulletin describing the issue and instructing them how to manage customer questions. " Complete and return the Field Correction Effectiveness Check form attached to this letter within 30 days. " Please review this letter with your Medical Director. Please retain this letter with your laboratory records, and forward this letter to those who may have received this product. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative. For further questions, please call (914) 524-2868.

Recall details

Recall number
Z-1615-2016
Recall class
Class III
Status
Terminated
Initiated by the company
Mar 30, 2016
Posted by the FDA
May 6, 2016
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
US: 442 units; Foreign: 3237 units
Product code
JJE
FDA submissions
K041133
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Afghanistan, Albania, Algeria, Angola, Argentina, Australia, Austria, Bangladesh, Bosnia Herzeg., Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Caledonia, Norway, Oman, Pakistan, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uganda, United Kingdom, Uruguay, Vietnam, Yemen.

Product description

ADVIA Centaur CP Immunoassay System, Siemens Material Number (SMN) 10309710, IVD. --- CLASSIFICATION NAME: Immunoassay Analyzers. This system is intended for professional use in a laboratory environment only. Tests performed using this system are intended for in vitro diagnostic use.

Code information (models, lots, serial numbers)

All serial numbers

Other devices under the same FDA submission

4 devices cleared under K041133 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.