Medical device recalls: May 2016
The FDA posted 106 medical device recall records in May 2016, from 46 recall events by 39 companies. 18 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Biomet, Inc. · 2 events, 15 records
- Covidien LLC · 2 events, 15 records
- Biomerieux Inc · 2 events, 10 records, 9 Class I
- Alcon Research, Ltd. · 2 events, 4 records
- Siemens Healthcare Diagnostics, Inc. · 2 events, 2 records
- Siemens Medical Solutions USA, Inc · 2 events, 2 records
- Skytron, Div. The KMW Group, Inc · 2 events, 2 records
- Zimmer Manufacturing B.V. · 1 event, 12 records
FDA root cause
- Equipment maintenance · 14
- Packaging process control · 13
- Packaging change control · 12
- Device Design · 11
- Process control · 10
- Labeling Change Control · 9
- Software design · 7
- Other · 5
Class I recalls (18)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1636-2016 · Hummingbird Med · posted May 20, 2016 · Terminated
Hummingbird Med Devices, Inc. is recalling ABG-HM-1 Hummi Micro Draw Blood Transfer because the connection between the "Hummi" Y connector and the yellow cannula hub may not be sufficiently secure enough and could separate while in use.
- Z-1614-2016 · Monteris Medical Corp · posted May 16, 2016 · Terminated
Monteris received a report that the sapphire lens on a 2.2SF probe fractured during a LITT procedure, with resulting introduction of C02 into the patient's brain.
- Z-1577-2016 · Medtronic · posted May 12, 2016 · Terminated
Medtronic Respiratory & Monitoring Solutions Announces a Voluntary Recall of a Battery Pack Used with the Capnostream"20 and Capnostream"20p Patient Monitor Due to the Battery Pack Might Experience Thermal Runaway with the Potential of Fire Risk to the…
- Z-1639-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1640-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1641-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1642-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1643-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1644-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1645-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1646-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1647-2016 · Biomerieux Inc · posted May 11, 2016 · Terminated
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
- Z-1531-2016 · B Braun Medical Inc · posted May 3, 2016 · Terminated
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- Z-1532-2016 · B Braun Medical Inc · posted May 3, 2016 · Terminated
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- Z-1533-2016 · B Braun Medical Inc · posted May 3, 2016 · Terminated
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- Z-1534-2016 · B Braun Medical Inc · posted May 3, 2016 · Terminated
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- Z-1535-2016 · B Braun Medical Inc · posted May 3, 2016 · Terminated
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
- Z-1536-2016 · B Braun Medical Inc · posted May 3, 2016 · Terminated
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
Class II recalls (87)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1838-2016 · Baxter Healthcare Corp · posted May 31, 2016 · Terminated
Potential for incomplete dissolution of the polyethylene glycol (PEG) component during the reconstitution of the product, which may affect the consistency of the hydrogel formation during use.
- Z-1839-2016 · Baxter Healthcare Corp · posted May 31, 2016 · Terminated
Potential for incomplete dissolution of the polyethylene glycol (PEG) component during the reconstitution of the product, which may affect the consistency of the hydrogel formation during use.
- Z-1840-2016 · Baxter Healthcare Corp · posted May 31, 2016 · Terminated
Potential for incomplete dissolution of the polyethylene glycol (PEG) component during the reconstitution of the product, which may affect the consistency of the hydrogel formation during use.
- Z-1841-2016 · Integra York PA, Inc · posted May 31, 2016 · Terminated
One lot of self-lock chisel blade handles have a wider slot that holds the chisel blade that may result in the inability to firmly secure the blade in the handle.
- Z-1842-2016 · ConMed Corporation · posted May 31, 2016 · Terminated
Metal shavings released from burs during use are due to contact between the bur and the hood of the PreBent Bur assembly. During the manufacturing process an assembly error occurred which caused inadvertent mixing of two sizes of bearings.
- Z-1844-2016 · Skytron, Div. The KMW Group, Inc · posted May 31, 2016 · Terminated
The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached…
- Z-1846-2016 · Sterilmed, Inc. · posted May 31, 2016 · Terminated
Sterlilmed is recalling all unexpired Scissor Tips because of a potential manufacturing defect that can possibly result in a patient injury.
- Z-1847-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1848-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1849-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1850-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1851-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1852-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1853-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Curity X-Ray Detectable Laparotomy Sponges, Soft Pouch Pack, 18" x 18" (45.7 cm x 45.7 cm) ↗Class IIZ-1854-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Curity X-Ray Detectable Laparotomy Sponges, Soft Pouch Pack, 12" x 12" (30.5 cm x 30.5 cm) ↗Class IIZ-1855-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1856-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Curity" X-Ray Detectable Laparotomy Sponges, Soft Pouch Pack, 8" x 36" (20.3 cm x 91.4 cm) ↗Class IIZ-1857-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1858-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1859-2016 · Covidien LLC · posted May 31, 2016 · Terminated
Compromised sterility due to breach of barrier
- Z-1866-2016 · Integra LifeSciences Corp. · posted May 31, 2016 · Terminated
Reports relating to unintended patient or operator burns or operator electric shock for some of Integra's MicroFrance electrosurgical instruments.
- Z-1867-2016 · Integra LifeSciences Corp. · posted May 31, 2016 · Terminated
Reports relating to unintended patient or operator burns or operator electric shock for some of Integra's MicroFrance electrosurgical instruments.
- Z-1871-2016 · Skytron, Div. The KMW Group, Inc · posted May 31, 2016 · Terminated
Potential for the weld to fail at the pivot point of the monitor bracket. This weld failure could cause the display interface and bracket to fall from its mounted location, resulting in injury.
- Z-1834-2016 · MicroAire Surgical Instruments, LLC · posted May 27, 2016 · Terminated
MicroAire discovered that 20 suspect Impactor instruments ( MicroAire Pneumatic Impactor) were shipped to end users and were missing the lock and unlock symbol markings on them.
- Z-1836-2016 · Intuitive Surgical, Inc. · posted May 27, 2016 · Terminated
Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the cannula lumen when insufflation is connected during a surgical procedure.
- Z-1837-2016 · Intuitive Surgical, Inc. · posted May 27, 2016 · Terminated
Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the cannula lumen when insufflation is connected during a surgical procedure.
- Z-1815-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1816-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1817-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1818-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1819-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1820-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1821-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1822-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1823-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1824-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1825-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1826-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1827-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1828-2016 · Biomet, Inc. · posted May 26, 2016 · Terminated
Insufficient seal strength on sterile barrier packaging. The seal could be jeopardized causing the sterile barrier to be compromised and the product could become nonsterile. If the product becomes non-sterile and is used or implanted, it may lead to an…
- Z-1829-2016 · RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.) · posted May 26, 2016 · Terminated
Potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury.
- Z-1712-2016 · Raysearch Laboratories AB · posted May 25, 2016 · Terminated
A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views. The views can become unsynchronized with the stored data if simultaneously right clicking, pressing…
- Z-1814-2016 · Biomet, Inc. · posted May 25, 2016 · Terminated
The Trocar and Plunger Assemblies are missing from the instrument, which could result in a delay in surgery greater than 30 minutes.
- Z-1705-2016 · Elekta, Inc. · posted May 23, 2016 · Terminated
If by mistake the initial pre-treatment imaging was performed BEFORE the HexaPOD was moved to the DRIVE (*START) position, iGUIDE offers the possibility to branch into a specific catch-up workflow, which was implemented to use the initial scan data.
- Z-1706-2016 · GE Medical Systems, LLC · posted May 23, 2016 · Terminated
Potential non-recoverable loss of image acquisition. The affected Discovery systems may experience multiple X-Ray abort errors before or during a real-time fluoroscopic Interventional procedure.
- Z-1707-2016 · GE Medical Systems, LLC · posted May 23, 2016 · Terminated
Potential non-recoverable loss of image acquisition. The affected Discovery systems may experience multiple X-Ray abort errors before or during a real-time fluoroscopic Interventional procedure.
- Z-1709-2016 · Synthes (USA) Products LLC · posted May 23, 2016 · Terminated
The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.
- Z-1710-2016 · Synthes (USA) Products LLC · posted May 23, 2016 · Terminated
The sterile packaging may be compromised. A protective cap is to be secured on the tip of the drill bit. If the protective cap is not secured, the drill bit may puncture the sterile barrier.
- Z-1701-2016 · Angiodynamics Inc. (Navilyst Medical Inc.) · posted May 20, 2016 · Terminated
Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in…
- Z-1702-2016 · Angiodynamics Inc. (Navilyst Medical Inc.) · posted May 20, 2016 · Terminated
Maquet Cardiopulmonary, the manufacturer of the RotaFlow Centrifugal Pump, determined that the products may possess levels of endotoxin that exceed regulatory limits. AngioDynamics confirmed that the affected RotaFlow Centrifugal Pumps were included in…
- Z-1703-2016 · B. Braun Medical, Inc. · posted May 20, 2016 · Terminated
Potential for the tip of the Huber needle on the WHIN Infusion Set to protrude from the needle guard.
- Z-1690-2016 · Becton Dickinson & Company · posted May 18, 2016 · Terminated
The device may have a defect in the catheter. In some instances this defect could result in catheter separation or breakage.
- Z-1692-2016 · Alcon Research, Ltd. · posted May 18, 2016 · Terminated
The product has potential to leak beyond their design specification.
- Z-1693-2016 · Alcon Research, Ltd. · posted May 18, 2016 · Terminated
The product has potential to leak beyond their design specification.
- Z-1694-2016 · Alcon Research, Ltd. · posted May 18, 2016 · Terminated
The product has potential to leak beyond their design specification.
- Z-1672-2016 · Medtronic Navigation, Inc. · posted May 16, 2016 · Terminated
Potential for the Visualase Cooled Laser Applicator System (VCLAS) catheters to melt during procedures.
- Zimmer Plates and Screws (ZPS)Class IIZ-1673-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Zimmer Plates and Screws (ZPS)Class IIZ-1674-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Zimmer universal locking system (ULS) 3.5 mm cortical screw selftapping 18 mm length 47493501801 ↗Class IIZ-1675-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Z-1676-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Zimmer Plates and Screws (ZPS)Class IIZ-1677-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Z-1678-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Z-1679-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Z-1680-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Z-1681-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Z-1682-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Z-1683-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Trilogy self-tapping bone screwClass IIZ-1684-2016 · Zimmer Manufacturing B.V. · posted May 16, 2016 · Terminated
A small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
- Lumenis M22 System Model Number: GA-0005400 (M22 IPL + YAG Module) with Acne Filter (KT-1014971). ↗Class IIZ-1669-2016 · Lumenis Ltd · posted May 13, 2016 · Terminated
Lumenis Ltd Announces a Field Action of the M22 IPL Acne Filters for the Lumenis M22 IPL Hand Piece due to the Risk of Superficial Burns When Using the Device.
- Z-1637-2016 · Bradshaw Medical Inc. · posted May 10, 2016 · Terminated
A component failure could cause a situation where the applied torque was significantly higher than the calibrated torque setting.
- Distal Femoral Replacement (patient specific, custom made orthopedic implant device). PIN 19960 ↗Class IIZ-1623-2016 · Stanmore Implants Worldwide Ltd. · posted May 9, 2016 · Terminated
Incorrect component used to manufacture distal femoral replacement.
- Z-1624-2016 · Hamilton Medical, Inc. · posted May 9, 2016 · Terminated
After performing the suctioning maneuver, including disconnecting the patient, suctioning , and reconnecting the patient, the preset pattern of ventilation many not continue as expected.
- Z-1613-2016 · Novarad Corporation · posted May 6, 2016 · Terminated
The SUV values that are being calculated in the PET/CT fusion tool are incorrect.
- Z-1617-2016 · C.R. Bard, Inc. · posted May 6, 2016 · Terminated
The product may contain electronic components that lead to a premature drainage of the internal control panel coin cell battery responsible for maintaining the system close and Static Random Access Memory (SRAM). This premature coin cell battery drainage…
- Z-1618-2016 · Siemens Healthcare Diagnostics, Inc. · posted May 6, 2016 · Terminated
Siemens Healthcare Diagnostics confirmed that Syva Emit Methotrexate Application (DVMTX.1) for Dimension Vista System does not meet On-Board Stability claim of 20 hours. Testing using reagent at the end of the 20 hours On-Board Stability has shown that…
- Z-1619-2016 · Covidien LLC · posted May 6, 2016 · Terminated
Potential of an allergic reaction in a patient with nickel sensitivity.
- Z-1620-2016 · Covidien LLC · posted May 6, 2016 · Terminated
Potential of an allergic reaction in a patient with nickel sensitivity.
- Z-1621-2016 · Applied Medical Resources Corp · posted May 6, 2016 · Terminated
Increased customer complaints indicating inconsistent clip application; may lead to unoccluded vessels.
- Syngo Dynamics ↗Class IIZ-1601-2016 · Siemens Medical Solutions USA, Inc · posted May 4, 2016 · Terminated
Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section…
- Z-1602-2016 · Medtronic Sofamor Danek Usa, Inc - Dallas Distribution · posted May 4, 2016 · Terminated
Incorrect color coding and attachment reference were included on the labeling of 10BA40D dissecting tools with lot number H5237248.
- Z-1607-2016 · Toshiba American Medical Systems Inc · posted May 4, 2016 · Terminated
Potential degradation of detector performance.
- Z-1596-2016 · Biomerieux Inc · posted May 3, 2016 · Terminated
Product may lead to false susceptible results of Staphylococcus spp., Enterococcus spp., and S. agalactiae to Minocycline.
- Z-1597-2016 · Alcon Research, Ltd. · posted May 3, 2016 · Terminated
The recalling firm received complaints related to leakage from the tubing near the white luer of the irrigation line for the INFINITI Ultrasound Packs.
- Z-1598-2016 · Siemens Medical Solutions USA, Inc · posted May 3, 2016 · Terminated
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
- Z-1600-2016 · Nuvectra · posted May 3, 2016 · Terminated
If used, the optional Swap feature copies programming parameters from the external pulse generator (EPG) as well as unintended calibration data to the implantable pulse generator (lPG). This may result in the user being unable to recharge the lPG. This…
- Z-1593-2016 · Carestream Health Inc · posted May 2, 2016 · Terminated
Carestream Health received a complaint related to CARESTREAM Image Suite 4 from a foreign hospital stating that the annotation on the overlay is displayed as "L (Left)", when it should be "R(Right)".
- CARESTREAM Image Suite V4 ↗Class IIZ-1594-2016 · Carestream Health Inc · posted May 2, 2016 · Terminated
Carestream Health received a complaint related to CARESTREAM Image Suite 4 from a foreign hospital stating that the annotation on the overlay is displayed as "L (Left)", when it should be "R(Right)".
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1615-2016 · Siemens Healthcare Diagnostics, Inc. · posted May 6, 2016 · Terminated
The firm identified an issue in sample processing when using EZee-Nest Insert Cups due to two inaccuracies that exist in the ADVIA Centaur CP Operators Guide.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.