Saw Sagittal Catalog Number: ES 300 recallZ-2085-2016 · Medtronic Sofamor Danek Usa, Inc - Dallas Distribution · initiated May 13, 2016
Class II Terminated All recalls posted in June 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Sofamor Danek Usa, Inc - Dallas Distribution or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic Powered Surgical Solutions testing of the Sagittal Saw Rocker Arm Seal identified a potential for leakage of fluid into the motor during use.
FDA root cause category: Device Design
What the company is doing
Recall notification letters were sent to consignees on 05/13/2016.
Recall details
- Recall number
- Z-2085-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 13, 2016
- Posted by the FDA
- Jun 28, 2016
- Recalling firm
- Medtronic Sofamor Danek Usa, Inc - Dallas Distribution
- Quantity
- 368 devices
- Product code
- HBE
- FDA submissions
- K081475
- Distribution
- Worldwide Distribution-Including the states of AL, AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, MA, MD, NC, NH, NY, SC, TN, TX, and UT.
Product description
Saw Sagittal Catalog Number: ES 300; Saw Sagittal Refurbished Catalog Number: ES300-R; Saw Saggital, Finger Catalog Number: ES310; Saw Saggital, Finger Refurbished Catalog Number: ES310-R; Used to dissect bone in surgical applications.
Code information (models, lots, serial numbers)
All Serial Numbers.
Other devices under the same FDA submission
30 devices cleared under K081475 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Indigo™ 1845000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)165 reports
- Indigo 1845020Medtronic Xomed, Inc. · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE105 reports
- Midas Rex® EM210Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)64 reports
- Midas Rex® EC300RFMedtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall1 report
- NA EC300RFMedtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 40R2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45R2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 40A2500Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 40R2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 40S2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45A2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
- N/A 45R2000Medtronic Xomed, Inc. · DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page