Medical device recalls: June 2016
The FDA posted 110 medical device recall records in June 2016, from 40 recall events by 35 companies. 12 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Synthes (USA) Products LLC · 4 events, 8 records
- Siemens Medical Solutions USA Inc. · 2 events, 37 records
- Alcon Research, Ltd. · 2 events, 2 records
- Pentax Medical Company · 1 event, 16 records
- Carefusion 211 Inc dba Carefusion · 1 event, 12 records, 12 Class I
- B. Braun Medical, Inc. · 1 event, 2 records
- GE Healthcare, LLC · 1 event, 2 records
- Nephron Pharmaceuticals Corp. · 1 event, 2 records
FDA root cause
- Nonconforming Material/Component · 39
- Device Design · 21
- Under Investigation by firm · 20
- Process control · 9
- Process design · 5
- Software design · 4
- Radiation Control for Health and Safety Act · 2
- Other · 2
Class I recalls (12)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1888-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1889-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1890-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1891-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1892-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1893-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- TCA Board, 16542AClass IZ-1894-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1895-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1896-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1897-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1898-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
- Z-1899-2016 · Carefusion 211 Inc dba Carefusion · posted Jun 13, 2016 · Terminated
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
Class II recalls (98)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2108-2016 · Beckman Coulter Inc. · posted Jun 30, 2016 · Terminated
Demonstrate an upward shift in patient results.
- Z-2088-2016 · Siemens Medical Solutions USA, Inc · posted Jun 29, 2016 · Terminated
Software error in previous software versions in which two references for the same image may exist in the database.
- Z-2089-2016 · CMP Industries, Llc · posted Jun 29, 2016 · Terminated
Possible presence of foreign material (rust).
- BronchoscopeClass IIZ-2065-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Choledocho-Nephroscope ↗Class IIZ-2066-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Choledochoscope ↗Class IIZ-2067-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- ColonoscopeClass IIZ-2068-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Confocal GI Scope ↗Class IIZ-2069-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- CystoscopeClass IIZ-2070-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- GastroscopeClass IIZ-2071-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Intubation Scope ↗Class IIZ-2072-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Laryngo Stroboscope ↗Class IIZ-2073-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Naso-Pharyngo-LaryngoscopeClass IIZ-2074-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- SignmoidoscopeClass IIZ-2075-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Ultrasound BronchoscopeClass IIZ-2076-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Ultrasound Gastroscope ↗Class IIZ-2077-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Ureteroscope ↗Class IIZ-2078-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Esophagoscope ↗Class IIZ-2079-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- Small Bowel Scope ↗Class IIZ-2080-2016 · Pentax Medical Company · posted Jun 28, 2016 · Terminated
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
- CENTURION Vision System ↗Class IIZ-2081-2016 · Alcon Research, Ltd. · posted Jun 28, 2016 · Terminated
The grounding pin within the Power Entry Switch on certain systems was susceptible to detachment when the AC power cable is removed.
- Z-2085-2016 · Medtronic Sofamor Danek Usa, Inc - Dallas Distribution · posted Jun 28, 2016 · Terminated
Medtronic Powered Surgical Solutions testing of the Sagittal Saw Rocker Arm Seal identified a potential for leakage of fluid into the motor during use.
- Z-2053-2016 · Nephron Pharmaceuticals Corp. · posted Jun 27, 2016 · Terminated
Lack of Assurance of Sterility: Internal dye ingress test failure to assure the integrity of the vial.
- Z-2054-2016 · Nephron Pharmaceuticals Corp. · posted Jun 27, 2016 · Terminated
Lack of Assurance of Sterility: Internal dye ingress test failure to assure the integrity of the vial.
- Z-2049-2016 · Baxter Healthcare Corp. · posted Jun 24, 2016 · Terminated
Presence of leaks near top of the PrismaSate bags
- Z-2005-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2006-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2007-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2008-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2009-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2010-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2011-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2012-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2013-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2014-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2015-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2016-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2017-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2018-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2019-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2020-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2021-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2022-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2023-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2024-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2025-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2026-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2027-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2028-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2029-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2030-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2031-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2032-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2033-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2034-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2035-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2036-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2037-2016 · Siemens Medical Solutions USA Inc. · posted Jun 21, 2016 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
- Z-2040-2016 · Immuno-Mycologics, Inc · posted Jun 20, 2016 · Terminated
The device was found to have bacterial contamination.
- Z-2004-2016 · Musculoskeletal Transplant Foundation, Inc. · posted Jun 16, 2016 · Terminated
AFT Diverted Tubes are being recalled due to a customer complaint involving one unit which had an unsealed Tyvek pouch. The non-sterile foil pouch may introduce contamination into the sterile field.
- Z-1993-2016 · Physio-Control, Inc. · posted Jun 15, 2016 · Terminated
The firm became aware that when using EtC02 in the kPa or % setting and in a situation where the reading is above 9.9 kPa, the display of the LIFEPAK 15 respiratory rate may partially obscure a portion of the leading digit of the EtC02 value. This affects…
- Z-1995-2016 · FSC Laboratories, Inc · posted Jun 15, 2016 · Terminated
Product had torn/loose seal around the mouthpiece.
- Z-1996-2016 · Ethicon, Inc. · posted Jun 15, 2016 · Terminated
Ethicon is recalling EVARREST Fibrin Sealant Patch because the characters "0-bad" were printed in the expiration date field on the foil pouch labels of the Fibrin Patch units instead of the valid expiration date (2016-12-28).
- Z-1997-2016 · Siemens Medical Solutions USA Inc. · posted Jun 15, 2016 · Terminated
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
- Z-1998-2016 · Siemens Medical Solutions USA Inc. · posted Jun 15, 2016 · Terminated
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
- Z-1999-2016 · Siemens Medical Solutions USA Inc. · posted Jun 15, 2016 · Terminated
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
- Z-2000-2016 · Siemens Medical Solutions USA Inc. · posted Jun 15, 2016 · Terminated
The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
- Z-1989-2016 · Excelsior Medical Corp · posted Jun 14, 2016 · Terminated
Certain lots of Heparin Lock Flush, USP Syringes (5 ml Fill in 10 ml Syringe) failed Anti-Factor IIa assay for potency during testing of retention samples for heparin potency.
- Z-1991-2016 · CarboFix Orthopedics, Ltd. · posted Jun 14, 2016 · Terminated
Firm received complaints where the Piccolo Composite Ball Tip Guide Wire penetrated the internal package, and thus compromised the product sterility.
- Z-1945-2016 · Sysmex Corporation · posted Jun 10, 2016 · Terminated
There is a potential for elevated hemoglobin (Hgb) and impedance platelet count (PLT-I) values.
- Z-1949-2016 · Synthes (USA) Products LLC · posted Jun 10, 2016 · Terminated
Product incorrectly packaged. The labels on the outside of the box and on the LCP Dia-Meta Volar Distal Radius Plate inside the box are correct; however, the preprinted box which contains the LCP-Dia-Meta Volar Distal Radius Plate is incorrect. The box…
- Z-1938-2016 · Teleflex Medical · posted Jun 9, 2016 · Terminated
The labeling of some units indicates that it contains a size 7mm tracheal tube, but may contain a size 7.5mm tracheal tube.
- Maestro Air Motor ↗Class IIZ-1939-2016 · Stryker Instruments Div. of Stryker Corporation · posted Jun 9, 2016 · Terminated
Due to the buildup of grease within the hose and motor, the product may require multiple attempts to start. The grease may also migrate into the attachment end of the motor. There is a negligible possibility that the orange grease may leak from the…
- Z-1940-2016 · Alcon Research, Ltd. · posted Jun 9, 2016 · Terminated
Alcon is conducting this Voluntary Medical Device Removal for specific lots of its CENTURION FMS packs due to the possible presence of a molding irregularity found within the aspiration luer which may result in a reduction of vacuum within the aspiration line.
- Z-1942-2016 · Synthes (USA) Products LLC · posted Jun 9, 2016 · Terminated
DePuy Synthes is initiating a voluntary medical device recall of unexpired and unopened part and lot numbers for the Norian Drillable Injects-Sterile due to the rotary pouch within the referenced lots and associated part numbers may potentially be labeled…
- Z-1943-2016 · Synthes (USA) Products LLC · posted Jun 9, 2016 · Terminated
DePuy Synthes is initiating a voluntary medical device recall of unexpired and unopened part and lot numbers for the Norian Drillable Injects-Sterile due to the rotary pouch within the referenced lots and associated part numbers may potentially be labeled…
- Z-1944-2016 · Synthes (USA) Products LLC · posted Jun 9, 2016 · Terminated
DePuy Synthes is initiating a voluntary medical device recall of unexpired and unopened part and lot numbers for the Norian Drillable Injects-Sterile due to the rotary pouch within the referenced lots and associated part numbers may potentially be labeled…
- Z-1927-2016 · GE Healthcare, LLC · posted Jun 8, 2016 · Terminated
The gas venting may not occur properly. A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. Cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users and patients.
- Z-1928-2016 · GE Healthcare, LLC · posted Jun 8, 2016 · Terminated
The gas venting may not occur properly. A site reported a magnet quenching with subsequent venting of the cryogen gas into the magnet room. Cryogen gas that vents into the magnet room could cause critical deprivation of oxygen to users and patients.
- Liko Universal SlingBars ↗Class IIZ-1916-2016 · Hill-Rom, Inc. · posted Jun 7, 2016 · Terminated
The center bolt of the sling bar, which connects the bar to the patient lift, has failed during use. If the sling bar is not used as intended the bolt may be weakened. If the bolt is weakened there is a potential risk for breakage with the result of a…
- Persona Trabecular Metal TibiaClass IIZ-1917-2016 · Zimmer Biomet, Inc. · posted Jun 7, 2016 · Terminated
Zimmer Inc. is initiating a recall of specific Persona" Trabecular Metal" Tibial Plate Instruments and Modular Brackets, due to incidents of radiolucent lines and loosening in implants.
- Z-1918-2016 · Synthes (USA) Products LLC · posted Jun 7, 2016 · Terminated
Labeling errors. The outer packge label expiration date may exceed the expiration date of the subcomponents.
- Z-1919-2016 · Synthes (USA) Products LLC · posted Jun 7, 2016 · Terminated
Labeling errors. The outer packge label expiration date may exceed the expiration date of the subcomponents.
- Z-1920-2016 · Synthes (USA) Products LLC · posted Jun 7, 2016 · Terminated
Labeling errors. The outer packge label expiration date may exceed the expiration date of the subcomponents.
- Z-1924-2016 · Medical Components, Inc dba MedComp · posted Jun 7, 2016 · Terminated
This recall has been initiated due to the product labeled with the incorrect expiration date.
- Z-1925-2016 · Berchtold Corp. · posted Jun 7, 2016 · Terminated
The Surgical Table Cushion (HybriGel Pads) could potentially contain staples that's secure two underlying fabrics, which would be captured on certain scans on patients, such as x-ray.
- Z-1926-2016 · Medtronic Neuromodulation · posted Jun 7, 2016 · Terminated
Loss of therapy during recharging of device. Reportedly, a charging session was terminated prior to obtaining recharge threshold voltage of 3.615, which triggered rapid battery depletion state. As a result, the device depleted to 1.925 volts in one to two…
- Z-1912-2016 · Synthes (USA) Products LLC · posted Jun 6, 2016 · Terminated
Devices did not pass the biological safety evaluation for cytotoxicity following exposure to test conditions. The high growth inhibition levels observed during testing could be attributed to corrosion of the device at solder points. This could…
- Z-1695-2016 · Philips Healthcare · posted Jun 3, 2016 · Terminated
The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image.
- Z-1696-2016 · Philips Healthcare · posted Jun 3, 2016 · Terminated
The detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image.
- Z-1910-2016 · Stryker Corporation · posted Jun 3, 2016 · Terminated
Complaints were received for the Stryker FlowPort II Adapter, and investigation found that the scope sat in the adapter 180 degrees in the wrong direction
- Z-1911-2016 · Integrated Science Systems Inc · posted Jun 3, 2016 · Terminated
Customer complaints of very weak reactions with the kit, or false negative results with different Staph aureus cultures.
- Z-1906-2016 · B. Braun Medical, Inc. · posted Jun 2, 2016 · Terminated
BBMI recently identified a potential for a tear in the outer blister packaging on a limited number of units. The tears have been identified in the blister package paper, where the unit label writing is located.
- Z-1907-2016 · B. Braun Medical, Inc. · posted Jun 2, 2016 · Terminated
BBMI recently identified a potential for a tear in the outer blister packaging on a limited number of units. The tears have been identified in the blister package paper, where the unit label writing is located.
- Z-1878-2016 · Natus Medical Incorporated · posted Jun 1, 2016 · Terminated
The replacement HEPA Filters have been manufactured with sealant that may not be properly cured. Improperly cured sealant can emit a toxic gas when heated, and can drip down during use, contaminating the interior chamber of the Sterile Drier and its contents.
- STAT-Check Resuscitation Bag ↗Class IIZ-1880-2016 · Ventlab LLC · posted Jun 1, 2016 · Terminated
Stat-check CO2 indicators are non-functional. The indicator is yellow when not in use and should be purple when not in use. The yellow indicates a 5% CO2 level, cardiac massage and ventiliation being performed correctly. The color will not change during…
- Z-1881-2016 · Orthofix, Inc · posted Jun 1, 2016 · Terminated
There is a possibility that the subject set screw driver could disengage from the torque limiting handle when the counter torque wrench is allowed to contact the counter torque handle.
- Z-1885-2016 · Siemens Healthcare Diagnostics Inc. · posted Jun 1, 2016 · Terminated
Discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare confirmed a software defect which, in a very specific set of circumstances, results in the Dimension Vista System omitting an aliquot probe rinse between sample…
- Z-1886-2016 · Siemens Healthcare Diagnostics Inc. · posted Jun 1, 2016 · Terminated
Discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare confirmed a software defect which, in a very specific set of circumstances, results in the Dimension Vista System omitting an aliquot probe rinse between sample…