Skip to content

ADVIA Chemistry (Syva Onboard) Theophylline_2 (THEO_2)… recallZ-2299-2016 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 17, 2016

Class II Terminated All recalls posted in July 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Siemens Healthcare Diagnostics has confirmed the Syva EMIT 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may exhibit increased imprecision for commercially available Bio-Rad Liquichek TDM Quality Control Levels and patient samples

FDA root cause category: Process control

What the company is doing

Siemens Healthcare Diagnostics mailed a letter to customers on June 17, 2016, asking that they discontinue use of and discard the affected kit lots identified in the recall notification letter. Customers were also asked to complete and return the Field Correction Effectiveness Check/Product Replacement Form that was mailed with the Recall Notification Letter.

Recall details

Recall number
Z-2299-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 17, 2016
Posted by the FDA
Jul 28, 2016
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
658 units
Product code
KLS
FDA submissions
K913123
Distribution
Distributed to: CA, CT, DE, FL, GA, IA, IN, KS, KY, LA, MA, MD, MI, MO, NC, NJ, NM, NY, OH, OK, PA, TN, VA, WA

Product description

ADVIA Chemistry (Syva Onboard) Theophylline_2 (THEO_2) Reagent is for in vitro diagnostic use in the quantitative analysis of theophylline in human serum or plasma on the ADVIA Chemistry Systems. Copies of IFUs included in submission.

Code information (models, lots, serial numbers)

ADVIA¿ Chemistry (Syva Onboard) Theophylline_2 (THEO_2) Reagent - SMN# 10377503, Lot # 334018

Other devices under the same FDA submission

1 device cleared under K913123 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 74524).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page