Medical device recalls: July 2016
The FDA posted 107 medical device recall records in July 2016, from 50 recall events by 41 companies. 4 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Synthes (USA) Products LLC · 4 events, 20 records
- Merge Healthcare, Inc. · 3 events, 3 records
- Arrow International Inc · 2 events, 3 records
- Stryker Instruments Div. of Stryker Corporation · 2 events, 3 records
- Beckman Coulter Inc. · 2 events, 2 records
- Toshiba American Medical Systems Inc · 2 events, 2 records
- Stryker Neurovascular · 1 event, 28 records
- Pentax Medical Company · 1 event, 6 records
FDA root cause
- Process change control · 29
- Labeling False and Misleading · 18
- Device Design · 13
- Software design · 9
- Nonconforming Material/Component · 7
- Process control · 6
- Under Investigation by firm · 5
- Radiation Control for Health and Safety Act · 3
Class I recalls (4)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2096-2016 · Stryker Sustainability Solutions · posted Jul 22, 2016 · Terminated
Product defect; tip of two (2) Angiographic Catheters model # 10732203 broke off, prior to being used on the patient.
- Z-2212-2016 · Carefusion 2200 Inc · posted Jul 22, 2016 · Terminated
BD, formerly CareFusion, has identified a potential risk associated with a weld failure which could result in the wire protruding thru the tip of the instrument when articulated. If this failure were to occur while in use in a procedure it has the…
- Z-1887-2016 · Heartware, Inc. · posted Jul 8, 2016 · Terminated
Heartware Ventricular Assist System Battery Cell experiencing premature depletion.
- Z-1908-2016 · Hummingbird Med · posted Jul 5, 2016 · Terminated
Hummingbird Med Devices, Inc. is recalling the Hummi Micro Draw because of insertion difficulties.
Class II recalls (99)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2303-2016 · Merge Healthcare, Inc. · posted Jul 29, 2016 · Terminated
A migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.
- Z-2297-2016 · Smith & Nephew, Inc. · posted Jul 28, 2016 · Terminated
The labels correctly state that the product should be a 25mm screw, but the box contained a longer, 35mm screw.
- Z-2298-2016 · Siemens Healthcare Diagnostics, Inc. · posted Jul 28, 2016 · Terminated
Siemens Healthcare Diagnostics has confirmed the Syva EMIT 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may exhibit increased imprecision for…
- Z-2299-2016 · Siemens Healthcare Diagnostics, Inc. · posted Jul 28, 2016 · Terminated
Siemens Healthcare Diagnostics has confirmed the Syva EMIT 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may exhibit increased imprecision for…
- Z-2291-2016 · Moller Vital · posted Jul 26, 2016 · Terminated
Potential for the bolt on the suspension coupling on the Molift Air to come loose from the housing.
- Avenir Mller Stem 4 StandardClass IIZ-2292-2016 · Zimmer Gmbh · posted Jul 26, 2016 · Terminated
Product mix-up. The Avenir Mller Stem 6 lateral uncemented might be placed in the packaging of the Avenir Mller Stem 4 standard uncemented and vice versa. Potential for delay in surgery, and/or potential for use of incorrect implant which may in turn…
- Z-2290-2016 · Sunrise Medical (US) LLC · posted Jul 25, 2016 · Terminated
There is an observed failure of the backrest bracket which can fatigue over the life time of the product. When this occurs there is the possibility of the bracket fracturing causing the backrest to detach from the chair frame.
- Z-2236-2016 · GE Inspection Technologies, LP · posted Jul 23, 2016 · Terminated
Radiation emissions which exceeded the limit in the performance standard were detected during maintenance of one system. The identified cause was insufficient attachment of shielding in the door. Jostling during delivery was identified as the probable…
- Colonoscope ↗Class IIZ-2275-2016 · Pentax Medical Company · posted Jul 23, 2016 · Terminated
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
- Confocal GI Scope ↗Class IIZ-2276-2016 · Pentax Medical Company · posted Jul 23, 2016 · Terminated
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
- DuodenoscopeClass IIZ-2277-2016 · Pentax Medical Company · posted Jul 23, 2016 · Terminated
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
- Gastroscope ↗Class IIZ-2278-2016 · Pentax Medical Company · posted Jul 23, 2016 · Terminated
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
- Ultrasound Gastroscope ↗Class IIZ-2279-2016 · Pentax Medical Company · posted Jul 23, 2016 · Terminated
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
- Signmoidoscope ↗Class IIZ-2280-2016 · Pentax Medical Company · posted Jul 23, 2016 · Terminated
Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.
- Z-2251-2016 · Best Theratronics Ltd. · posted Jul 22, 2016 · Terminated
It was discovered that both the new device and the predicate device failed to comply with performance standard.
- UniCel DxC Synchron Clinical Systems and UniCel DxC Synchron Access Integrated Clinical Systems ↗Class IIZ-2272-2016 · Beckman Coulter Inc. · posted Jul 22, 2016 · Terminated
Possible leaking of No Foam solution from the No Foam Bottle Assembly on the UniCel DxC Synchron Systems.
- Z-2246-2016 · DeVilbiss Healthcare LLC · posted Jul 20, 2016 · Terminated
A small percentage of units of 7305 Series Vacu-Aide Suction Units are experiencing premature failure during operation.
- Z-2247-2016 · Beckman Coulter Inc. · posted Jul 20, 2016 · Terminated
Beckman Coulter has identified an incompatibility between No Foam DxC reagent (REF B64130) and the DxC hydropneumatic parts associated with No Foam delivery.
- 2.3MM Tapered Router, ↗Class IIZ-2238-2016 · Stryker Instruments Div. of Stryker Corporation · posted Jul 19, 2016 · Terminated
A variation in flute depth on the routers was observed. The variation in flute depth may cause the routers to be more susceptible to breakages. Risk to Health: Injury to critical soft tissue requiring surgical intervention may occur.
- Z-2239-2016 · Hologic, Inc · posted Jul 19, 2016 · Terminated
Cracks ATEC Canister Lids.
- Z-2240-2016 · Biomet, Inc. · posted Jul 19, 2016 · Terminated
The screw may fracture at the junction of the shaft and the head during implantation. The patient may retain a foreign body if the fractured screw cannot be removed. Soft tissue irritation will not occur as the retained portion of the screw is below the…
- Z-2242-2016 · Encore Medical, Lp · posted Jul 19, 2016 · Terminated
Reported failure of the instrument. The retaining ring component of a neck trial became disassembled from the body of the neck.
- Z-2243-2016 · Parker Medical · posted Jul 19, 2016 · Terminated
Parker Medical announces a voluntary field action for the Parker Trach-Vac Endotracheal Tubes - (Endotracheal tube,with Parker Flex-Tip,Sub-Glottic Suction, High Volume low Pressure) due to tube kinking during use.
- Z-2188-2016 · Toshiba American Medical Systems Inc · posted Jul 18, 2016 · Terminated
During a procedure the Peak Skin Dose (PSD) value was displayed on the DTS larger than the DTS expected value. It was found that the software incorrectly calculated the PSD value when the x-ray condition of the frontal plane and the lateral plane are the…
- Z-2217-2016 · Carestream Health Inc · posted Jul 18, 2016 · Terminated
A software issue related to the generic volume measurement functionality could result in the volume of the subject anatomy being overestimated. For example, measurement results of the anatomy may appear enlarged or distended when it actually is not.
- Z-2219-2016 · WalkMed Infusion, LLC · posted Jul 18, 2016 · Terminated
WalkMed Infusion, LLC Announces a Voluntary Field Action of the Triton Infusion Pump (model 300000) and Triton FP Infusion Pump (model 400000) Due to Unapproved Changes to the Software and Specifications of the Triton Infusion Pump (model 300000) and…
- Z-2220-2016 · WalkMed Infusion, LLC · posted Jul 18, 2016 · Terminated
WalkMed Infusion, LLC Announces a Voluntary Field Action of the Triton Infusion Pump (model 300000) and Triton FP Infusion Pump (model 400000) Due to Unapproved Changes to the Software and Specifications of the Triton Infusion Pump (model 300000) and…
- Z-2214-2016 · Synthes (USA) Products LLC · posted Jul 15, 2016 · Terminated
DePuy Synthes is initiating a voluntary medical device recall due to Blade/Screw Guide Sleeves (Part # 03.037.017) may have been manufactured with a slightly larger than specified outer diameter which could prevent the sleeve from going through the TFNA…
- Z-2204-2016 · Stryker Instruments Div. of Stryker Corporation · posted Jul 14, 2016 · Terminated
Stryker Instruments initiated a voluntary recall of specific lots of Round Fluted Burs, due to tarnishing or corrosion which may be present on the recalled burs and could result in a foreign body reaction (inflammation) necessitating surgical intervention.
- Z-2205-2016 · Stryker Instruments Div. of Stryker Corporation · posted Jul 14, 2016 · Terminated
Stryker Instruments initiated a voluntary recall of specific lots of Round Fluted Burs, due to tarnishing or corrosion which may be present on the recalled burs and could result in a foreign body reaction (inflammation) necessitating surgical intervention.
- Z-2207-2016 · Terumo Cardiovascular Systems Corporation · posted Jul 14, 2016 · Terminated
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
- Sarns TCM ↗Class IIZ-2208-2016 · Terumo Cardiovascular Systems Corporation · posted Jul 14, 2016 · Terminated
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
- Sarns TCM II ↗Class IIZ-2209-2016 · Terumo Cardiovascular Systems Corporation · posted Jul 14, 2016 · Terminated
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
- Z-2210-2016 · Terumo Cardiovascular Systems Corporation · posted Jul 14, 2016 · Terminated
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Inter chlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
- Z-2211-2016 · Devicor Medical Products Inc · posted Jul 14, 2016 · Terminated
A 10 Gauge needle, 9cm long was shipped in an 8 Gauge needle, 15 cm package.
- Z-2189-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2190-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2191-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2192-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2193-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2194-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2195-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2196-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2197-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Z-2198-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Part # 313.250, 1.1mm Drill Bit, Mini QC with 4mm Stop/44.5mm, lot # 4326989, mfg. 9/28/2001Class IIZ-2199-2016 · Synthes (USA) Products LLC · posted Jul 12, 2016 · Terminated
DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.
- Femoral Arterial LineClass IIZ-2201-2016 · Arrow International Inc · posted Jul 12, 2016 · Terminated
Labeling error. Some shipping cartons labeled as ASK-04018-VUH contained ASK-04020-VUH kits; and some shipping cartons labeled as ASK-04020-VUH contained ASK-04018-VUH kits.
- Femoral Arterial LineClass IIZ-2202-2016 · Arrow International Inc · posted Jul 12, 2016 · Terminated
Labeling error. Some shipping cartons labeled as ASK-04018-VUH contained ASK-04020-VUH kits; and some shipping cartons labeled as ASK-04020-VUH contained ASK-04018-VUH kits.
- Z-2149-2016 · Varian Medical Systems, Inc. · posted Jul 11, 2016 · Terminated
Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different couch rotation from the plan couch…
- Z-2150-2016 · Varian Medical Systems, Inc. · posted Jul 11, 2016 · Terminated
Varian Medical Systems has received reports that unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up. This can occur when a user sets a different couch rotation from the plan couch…
- 6.5 mm Cannulated Screw, Product code HWC, Device Listing Number D028949, Screw, Fixation BoneClass IIZ-2152-2016 · Synthes (USA) Products LLC · posted Jul 11, 2016 · Terminated
Labeling does not match the cleared indications for use in the United States and Canada.
- 6.5 mm Cancellous Screw, Product code HWC, Device Listing Number D029043, Screw, Fixation BoneClass IIZ-2153-2016 · Synthes (USA) Products LLC · posted Jul 11, 2016 · Terminated
Labeling does not match the cleared indications for use in the United States and Canada.
- Z-2154-2016 · Synthes (USA) Products LLC · posted Jul 11, 2016 · Terminated
Labeling does not match the cleared indications for use in the United States and Canada.
- Z-2155-2016 · Synthes (USA) Products LLC · posted Jul 11, 2016 · Terminated
Labeling does not match the cleared indications for use in the United States and Canada.
- Z-2156-2016 · Synthes (USA) Products LLC · posted Jul 11, 2016 · Terminated
Labeling does not match the cleared indications for use in the United States and Canada.
- Z-2157-2016 · Synthes (USA) Products LLC · posted Jul 11, 2016 · Terminated
Labeling does not match the cleared indications for use in the United States and Canada.
- Z-2158-2016 · Synthes (USA) Products LLC · posted Jul 11, 2016 · Terminated
Labeling does not match the cleared indications for use in the United States and Canada.
- Z-2159-2016 · Merge Healthcare, Inc. · posted Jul 11, 2016 · Terminated
Potential data loss occurs as a result of product archiving not working properly.
- Target Detachable Coils: ↗Class IIZ-2160-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2161-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2162-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2163-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2164-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2165-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2166-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2167-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2168-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2169-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2170-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2171-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2172-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2173-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2174-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2175-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2176-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2177-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2178-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2179-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2180-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2181-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2182-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2183-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2184-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2185-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2186-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Target Detachable Coils: ↗Class IIZ-2187-2016 · Stryker Neurovascular · posted Jul 11, 2016 · Terminated
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased…
- Z-2144-2016 · Merge Healthcare, Inc. · posted Jul 8, 2016 · Terminated
Potential data loss occurs as a result of the software archiving not working properly.
- Z-2146-2016 · Hardy Diagnostics · posted Jul 8, 2016 · Terminated
Discrepant culture and gram stain results, which could potentially lead to misdiagnosis or improper treatment of a patient.
- Z-2147-2016 · Sorin Group USA, Inc. · posted Jul 8, 2016 · Terminated
Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.
- Z-2138-2016 · Synthes (USA) Products LLC · posted Jul 7, 2016 · Terminated
DePuy Synthes is voluntarily initiating a recall due to affected devices being distributed and or labeled with MR Safety information that does not meet the current ASTM standard.
- Z-2129-2016 · Medtronic Navigation · posted Jul 6, 2016 · Terminated
Medtronic Navigation is recalling certain components of the PoleStar system (N20 / N30) because of misplaced ground pins which may cause electrical shock to the user.
- Z-2130-2016 · Custom Ultrasonics, Inc. · posted Jul 6, 2016 · Terminated
To warn customers that Custom Ultrasonics, Inc. System 83 Plus AERs should not be used for cleaning and/or high-level disinfection of duodenoscopes until further notice.
- Z-2132-2016 · Lifeline First Aid LLC · posted Jul 6, 2016 · Terminated
First aid kits or rescue kits are recalled because they contains the CPR Shield with One Way Valve and Barrier Filter (mask) and this mask is cleared only for prescription use and is not cleared for over-the-counter (OTC) use.
- Z-2133-2016 · Lifeline First Aid LLC · posted Jul 6, 2016 · Terminated
Cadillac First Aid Kit is recalled because it contains the CPR Shield with One Way Valve and Barrier Filter (mask) and this mask is cleared only for prescription use and is not cleared for over-the-counter (OTC) use.
- Z-2134-2016 · Karl Storz Endoscopy · posted Jul 6, 2016 · Terminated
Potential electrical interference issue when the Image1 SPIES X-LINK and Image1 SPIES D1 camera head are used in conjunction with the DCI Video Mediastinocope and an electro-surgery generator (ESG).
- Z-2135-2016 · Toshiba American Medical Systems Inc · posted Jul 6, 2016 · Terminated
A potential problem associated to the gradient coil (ASGC) in the Titan 3T MRI System.
- Z-2136-2016 · Arrow International Inc · posted Jul 6, 2016 · Terminated
Arrow is recalling these products due to the possibility that although the kit is labeled as containing a 20 cm catheter, a 16 cm catheter may be packaged in the kit.
- Z-2124-2016 · Esaote S.p.A. · posted Jul 5, 2016 · Terminated
The device is being recalled because dynamic loading forces from larger patients can fracture the extremity portion of the patient table. Additionally, a weight limit is added to the patient seat for weight-bearing examinations.
- Z-2125-2016 · Medtronic Inc., Cardiac Rhythm and Heart Failure · posted Jul 5, 2016 · Terminated
Recently, a new software version was automatically sent to a subset of Model 24950 MyCareLink Monitors. After release, Medtronic identified an issue with the software that prevents implanted device data from being available to clinicians on the CareLink…
Class III recalls (4)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Instructions for Use (IFU) for Progesterone 3 Calibrator Set, repackaged as one set/package. ↗Class IIIZ-2282-2016 · Tosoh Bioscience, Inc. · posted Jul 25, 2016 · Terminated
Instructions For Use had incorrect lot number and calibrator values
- Z-2203-2016 · Roche Molecular Systems, Inc. · posted Jul 13, 2016 · Terminated
Kit labeled with the incorrect expiration date.
- Breathe Right Clear, 30 ct. carton., UPC 7-57145-00246-7 (US), UPC 9-300673-871749 (AU/NZ) ↗Class IIIZ-2148-2016 · GSK Consumer Healthcare · posted Jul 11, 2016 · Terminated
Some cartons used in the secondary packaging of Breathe Right Clear, Large, 30 ct. Nasal Strips were incorrectly labeled for cartons intended for the AU/NZ market.
- Z-2143-2016 · Ortho-Clinical Diagnostics · posted Jul 8, 2016 · Terminated
If the user selects the Comment or Interpretation option (in the enGEN .gsb file), and the numerical results fall between Negative and Reactive, the IM incorrectly displays the interpretive text as Borderline and sends the incorrect text to the LIS. The…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.