Siemens RAPIDPoint 500 v2.2.2A Software Upgrade Kit recallZ-2601-2016 · Siemens Healthcare Diagnostics Inc · initiated Jul 5, 2016

Class III Terminated All recalls posted in August 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some v2.2.2 upgrade kits include a dialysate mode which not cleared/approved for shipment in the United States

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens sent a Customer Notification dated July 2016 to all affected domestic (US) customers via Federal Express on July 5, 2016. Customers were asked to take the following actions: Please do not use the dialysate mode. If you have enabled this feature, please disable it by following these instructions: 1. At the System screen, select Setup. 2. Select Secured Options > Analysis Options. 3. Unselect Dialysate. 4. Select the Continue button 3 times. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check attached to this letter within 7 days. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens Technical Support representative. For questions regarding this recall call 781-269-3000.

Recall details

Recall number
Z-2601-2016
Recall class
Class III
Status
Terminated
Initiated by the company
Jul 5, 2016
Posted by the FDA
Aug 19, 2016
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
76
Product code
KHP
FDA submissions
K113216
Distribution
Nationwide Distribution to AL, CA, MA, NJ, OH, TX, WA, PA

Product description

Siemens RAPIDPoint 500 v2.2.2A Software Upgrade Kit; SMN 11066719, Software Version 2.2A The RAPIDPoint 500 system is designed for professional use in a point-of care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili

Code information (models, lots, serial numbers)

Software Version 2.2A

Other devices under the same FDA submission

1 device cleared under K113216 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.