Aortic Arch Cannula and Coronary Artery Perfusion Cannula… recallZ-2756-2016 · Sorin Group USA, Inc. · initiated Aug 12, 2016
Class II Terminated All recalls posted in September 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Sorin Group USA, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.
FDA root cause category: Process control
What the company is doing
LivaNova issued an Urgent Safety Alert letter via certified mail on August 12, 2016 to all affected customers. The letter instructed customers to check their inventory to determine if the Cannulae belongs to the product code and lots listed. If the customer does not have any of the products described in the communication, please complete the response form below and return it per the instructions indicated on the form; otherwise, please check the tip angle to determine if it is conforming to specifications. If specification is not met, please file a complaint through your normal process whether you decide to keep it and use it or return the product to Sorin Group; if you are not willing to use it, return the affected parts to Sorin Group USA. Please contact Customer Service 1-S00-650-2623 or email [email protected] to have a Return Material Authorization (RMA) issued. Complete the Customer Response Form attached to this letter to acknowledge receipt of this letter. Please return this form no later than September 15, 2016. For questions regarding this recall call 303-467-6306.
Recall details
- Recall number
- Z-2756-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 12, 2016
- Posted by the FDA
- Sep 8, 2016
- Recalling firm
- Sorin Group USA, Inc.
- Quantity
- 200 units worldwide
- Product code
- DWF
- FDA submissions
- K943934
- Distribution
- Worldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.
Product description
Aortic Arch Cannula and Coronary Artery Perfusion Cannula with Ballon (part number NA-55X6). The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect. The Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip
Code information (models, lots, serial numbers)
Lot Numbers: 1526500032, 1529300026, 1602600042.
Other devices under the same FDA submission
7 devices cleared under K943934 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Arterial cardiopulmonary bypass cannula RA-xxxxxLivanova USA, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS3 reports
- Arterial cardiopulmonary bypass cannula RA-xxxxxLivanova USA, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 reports
- Venous cardiopulmonary bypass cannula LR-xxxxxLivanova USA, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
- Venous cardiopulmonary bypass cannula LR-xxxxxLivanova USA, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
- Arterial cardiopulmonary bypass cannula NA-xxxxxLivanova USA, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
- Arterial cardiopulmonary bypass cannula RA-xxxxxLivanova USA, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
- Venous cardiopulmonary bypass cannula RT-xxxxxLivanova USA, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 74934).
- Z-2757-2016 · TerminatedCoronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page