Medical device recalls: September 2016
The FDA posted 131 medical device recall records in September 2016, from 59 recall events by 50 companies. 13 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Siemens Medical Solutions USA, Inc · 3 events, 6 records
- Mentor Texas, LP. · 2 events, 4 records
- Hologic, Inc · 2 events, 3 records
- Sorin Group USA, Inc. · 2 events, 3 records
- Stryker Howmedica Osteonics Corp. · 2 events, 3 records
- Alcon Research, Ltd. · 2 events, 2 records
- Beckman Coulter Inc. · 2 events, 2 records
- Merge Healthcare, Inc. · 2 events, 2 records
FDA root cause
- Process control · 47
- Nonconforming Material/Component · 15
- Software design · 12
- Use error · 11
- Device Design · 9
- Under Investigation by firm · 6
- Employee error · 6
- Labeling mix-ups · 4
Class I recalls (13)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2783-2016 · The Anspach Effort, Inc. · posted Sep 27, 2016 · Terminated
There was a potential for the Adaptor and Light Adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the products to burst.
- Z-2784-2016 · The Anspach Effort, Inc. · posted Sep 27, 2016 · Terminated
There was a potential for the Adaptor and Light Adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the products to burst.
- Z-2693-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2694-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2695-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2696-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2697-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2698-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2699-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2700-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2701-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2702-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
- Z-2703-2016 · MicroPort Orthopedics Inc. · posted Sep 8, 2016 · Terminated
ADVANCE(R) HA Coated tibial bases are being recalled due to a higher than expected revision rate as a result of component loosening.
Class II recalls (112)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2900-2016 · Beckman Coulter Inc. · posted Sep 28, 2016 · Terminated
Beckman Coulter is recalling the AQUIOS CL System because the use of the reagent carousel can lead to misalignment of the carousel and damage to the prep probe.
- Z-2890-2016 · EKOS Corporation · posted Sep 27, 2016 · Terminated
Casters on the EKOS carts may be missing washers which can cause premature wear and dislodgement of the casters/wheels leading to failure.
- Z-2891-2016 · Siemens Medical Solutions USA, Inc · posted Sep 27, 2016 · Terminated
Software upgrade to eliminate several issues
- Z-2892-2016 · Siemens Medical Solutions USA, Inc · posted Sep 27, 2016 · Terminated
Software upgrade to eliminate several issues
- Z-2893-2016 · Zyno Medical LLC · posted Sep 27, 2016 · Terminated
Administration Set potential filter leaking
- Z-2894-2016 · Zyno Medical LLC · posted Sep 27, 2016 · Terminated
Administration Set potential filter leaking
- Z-2895-2016 · Zyno Medical LLC · posted Sep 27, 2016 · Terminated
Administration Set potential filter leaking
- Z-2896-2016 · Zyno Medical LLC · posted Sep 27, 2016 · Terminated
Administration Set potential filter leaking
- Z-2885-2016 · Spacelabs Healthcare Inc · posted Sep 26, 2016 · Terminated
The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital.
- Z-2886-2016 · Spacelabs Healthcare Inc · posted Sep 26, 2016 · Terminated
The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital.
- Z-2887-2016 · R & D Systems, Inc. · posted Sep 26, 2016 · Terminated
R&D Systems, Inc. received two customer complaints that results of the Controls provided with the Quantikine IVD Human sTfR Immunoassay kit were out of the range high. Internal testing within R&D Systems confirmed a problem with Control results being out…
- Z-2884-2016 · Microgenics Corporation · posted Sep 23, 2016 · Terminated
Some lots of DRI Salicylate Serum Tox Assay negative patient samples are recovering higher than the product sensitivity claim of 4.4 mg/dL.
- Z-2870-2016 · Global Healthcare Inc · posted Sep 22, 2016 · Terminated
The multi-sample needle retractable sheath may not fully retract over the needle during the changes of the blood collection tubes.
- Z-2872-2016 · Mckesson Medical Imaging Group · posted Sep 22, 2016 · Terminated
The STAT priority icon was not displayed in the Study List when the study was opened.
- Z-2852-2016 · Siemens Medical Solutions USA, Inc · posted Sep 21, 2016 · Terminated
Software error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high.
- Z-2853-2016 · Siemens Medical Solutions USA, Inc · posted Sep 21, 2016 · Terminated
Software error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high.
- Z-2855-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2856-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2857-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2858-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2859-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2860-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2861-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2862-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2863-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2864-2016 · Abbott Medical Optics Inc. (AMO) · posted Sep 21, 2016 · Terminated
Abbott Medical Optics is recalling Intraocular Lenses (IOLs) because they were not being checked prior to release and this may result in the release of mislabeled IOLs.
- Z-2866-2016 · Sorin Group USA, Inc. · posted Sep 21, 2016 · Terminated
Sorin/LivaNova is initiating a field correction on the S5 Heart-lung machine because of the potential for failure of the Uninterruptible Power Supply.
- Sertera Biopsy Device, SterileClass IIZ-2869-2016 · Hologic, Inc · posted Sep 21, 2016 · Terminated
Incorrect label on the Sertera Biopsy Kit
- Z-2843-2016 · OrthoPediatrics Corp · posted Sep 19, 2016 · Terminated
OrthoPediatrics Corp.is voluntarily recalling multiple lots of the 3.5mm Bowed Locking Compression Femur 14-hole Plate and 18-hole Plate due to the devices being mislabeled.
- Z-2844-2016 · OrthoPediatrics Corp · posted Sep 19, 2016 · Terminated
OrthoPediatrics Corp.is voluntarily recalling multiple lots of the 3.5mm Bowed Locking Compression Femur 14-hole Plate and 18-hole Plate due to the devices being mislabeled.
- Z-2836-2016 · Ormco/Sybronendo · posted Sep 16, 2016 · Terminated
SybronEndo is recalling the Sealapex Xpress because it was labeled with an incorrect expiration date of 2017-11 while the syringe was labeled with the correct expiration date of 2016-11.
- Z-2841-2016 · Siemens Healthcare Diagnostics, Inc. · posted Sep 16, 2016 · Terminated
24 minutes of incubation is required prior to processing patient samples and for the pretreatment solution to reach full equilibration when using pretreatment solution when using the pretreatment solution (LGFA) contained in the kit.
- Z-2805-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2806-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2807-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2808-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2809-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2810-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2811-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2812-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2813-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2814-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2815-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2816-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2817-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2818-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2819-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2820-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2821-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2822-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2823-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2824-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2825-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2826-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2827-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2828-2016 · ICU Medical, Inc. · posted Sep 15, 2016 · Terminated
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
- Z-2829-2016 · AirStrip Technologies, Inc. · posted Sep 15, 2016 · Terminated
Belt clip may become detached causing electronic components to be exposed.
- Z-2830-2016 · C.R. Bard, Inc. · posted Sep 15, 2016 · Terminated
During an FDA inspection it was found out that the Practical Foley Catheters to be potentially nonsterile.
- Z-2831-2016 · C.R. Bard, Inc. · posted Sep 15, 2016 · Terminated
During an FDA inspection it was found out that the Practical Foley Catheters to be potentially nonsterile.
- LAUREATE World Phaco System ↗Class IIZ-2834-2016 · Alcon Research, Ltd. · posted Sep 15, 2016 · Terminated
Steps 16-25 are missing from the LAUREATE Multipak FMS Directions for Use (DFU)
- Z-2793-2016 · Siemens Medical Solutions USA, Inc · posted Sep 14, 2016 · Terminated
Siemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values. This suffix stands for on projection and indicates that a shown measurement value may be different…
- Z-2794-2016 · Siemens Medical Solutions USA, Inc · posted Sep 14, 2016 · Terminated
Siemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values. This suffix stands for on projection and indicates that a shown measurement value may be different…
- Z-2789-2016 · Stryker Howmedica Osteonics Corp. · posted Sep 13, 2016 · Terminated
Upon failure of the press-fit, the clamping sub components and pins could disassociate from the Scorpio Patella Assembly Instrument.
- Z-2791-2016 · Stryker Howmedica Osteonics Corp. · posted Sep 13, 2016 · Terminated
Stryker discovered that a Tibial Nail and femoral Nail were incorrectly labeled and distributed to the field.
- Z-2792-2016 · Stryker Howmedica Osteonics Corp. · posted Sep 13, 2016 · Terminated
Stryker discovered that a Tibial Nail and femoral Nail were incorrectly labeled and distributed to the field.
- Z-2785-2016 · Allergy Laboratories, Inc. · posted Sep 12, 2016 · Terminated
Potential for glass particles within the vials
- Z-2780-2016 · Mentor Texas, LP. · posted Sep 10, 2016 · Terminated
The CPX4 Tissue Expander may have been packaged in the Artoura Tissue Expander box.
- Z-2781-2016 · Mentor Texas, LP. · posted Sep 10, 2016 · Terminated
The CPX4 Tissue Expander may have been packaged in the Artoura Tissue Expander box.
- Z-2770-2016 · Sysmex America Inc · posted Sep 9, 2016 · Terminated
There is a potential that fluorescent platelet (PLT-F) and immature platelet fraction (IPF%) results performed using the recalled lots may exhibit lower than expected PLT-F and IPF values. The issue may be recognized by a large discrepancy between the…
- Z-2775-2016 · Thermo Finnigan LLC · posted Sep 9, 2016 · Terminated
Thermo Fisher has determined that the Endura MD mass spectrometer instrument control software versions 1.0 and 1.0 SP! have a software defect which affect data accuracy.
- Z-2776-2016 · Biosense Webster, Inc. · posted Sep 9, 2016 · Terminated
Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China without testing and approval by Biosense Webster Technical Services, as described in the Instructions for Use (IFU).
- Z-2750-2016 · Hill-Rom, Inc. · posted Sep 8, 2016 · Terminated
- Z-2752-2016 · Zimmer Biomet, Inc. · posted Sep 8, 2016 · Terminated
Zimmer Biomet is conducting a voluntary medical device recall of a single lot of the Coonrad/ Morrey Elbow Pin/ Bushing Replacement size Kit Extra Small because the lot was packaged with a missing humeral bushing. The missing bushing could result in a…
- Axiom REG implants Sterile ↗Class IIZ-2753-2016 · Anthogyr · posted Sep 8, 2016 · Terminated
Following a FDA inspection showing that the firm failed to validate the design and process of its implants.
- Axiom 2.8 implants Sterile ↗Class IIZ-2754-2016 · Anthogyr · posted Sep 8, 2016 · Terminated
Following a FDA inspection showing that the firm failed to validate the design and process of its implants.
- Z-2755-2016 · Aspen Surgical Puerto Rico Corp. · posted Sep 8, 2016 · Terminated
After receiving a complaint the firm disclosed that Conventional Stainless Steel Surgical Blade catalog number 371712 lot 0088464 was incorrectly identified in the single unit foil pack as 12 instead of 12B.
- Z-2756-2016 · Sorin Group USA, Inc. · posted Sep 8, 2016 · Terminated
Sorin Group USA, Inc. announces a voluntary field for the Aortic Arch Cannula and Coronary Artery Perfusion Cannula With Ballon because the tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of…
- Z-2757-2016 · Sorin Group USA, Inc. · posted Sep 8, 2016 · Terminated
The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.
- Z-2759-2016 · Teleflex Medical · posted Sep 8, 2016 · Terminated
Misbranded: Incorrect etching on the device.
- Z-2763-2016 · Blackrock Microsystems, LLC · posted Sep 8, 2016 · Terminated
Blackrock Microsystems, LLC announces a voluntary field action for the Blackrock NeuroPort Biopotential Signal Processing System because the Neural Signal Amplifier and the Patient Cable are incorrectly labeled as Type CF Applied Parts.
- Patient Cable. Part Numbers: 4460, 6735. Sterile within Chipboard Boxes within corrugated cartons. ↗Class IIZ-2764-2016 · Blackrock Microsystems, LLC · posted Sep 8, 2016 · Terminated
Blackrock Microsystems, LLC announces a voluntary field action for the Blackrock NeuroPort Biopotential Signal Processing System because the Neural Signal Amplifier and the Patient Cable are incorrectly labeled as Type CF Applied Parts.
- Z-2765-2016 · Hologic, Inc · posted Sep 8, 2016 · Terminated
Tip Breakage at the glue joint between the obturator body and tip
- Z-2766-2016 · Hologic, Inc · posted Sep 8, 2016 · Terminated
Tip Breakage at the glue joint between the obturator body and tip
- Z-2769-2016 · Innovision A/S · posted Sep 8, 2016 · Terminated
The purpose of the recall is to replace data transmission cable (USB-cable) connecting the lnnocor device with an external computer. The replacement USB cable will provide galvanic isolation of the lnnocor from the external computer via opto-couplers, as…
- Z-2740-2016 · Arrow International Inc · posted Sep 7, 2016 · Terminated
Shipping carton labeling error.
- Z-2741-2016 · Arrow International Inc · posted Sep 7, 2016 · Terminated
Shipping carton labeling error.
- Z-2747-2016 · Mentor Texas, LP. · posted Sep 7, 2016 · Terminated
Tissue expander product mix found: CPX4 350cc was placed in a carton for the CPX4 450cc
- Z-2748-2016 · Mentor Texas, LP. · posted Sep 7, 2016 · Terminated
Tissue expander product mix found: CPX4 350cc was placed in a carton for the CPX4 450cc
- Z-2749-2016 · Boston Scientific Corporation · posted Sep 7, 2016 · Terminated
Five Accolate MRI and Essentio MRI DR pulse generators with an alternate MICS module configuration were shipped to US customers and implanted. This alternate configuration did not receive FDA approval yet.
- Z-2722-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Z-2723-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Z-2724-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Z-2725-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Z-2726-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Z-2727-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Z-2728-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Z-2729-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Z-2730-2016 · Hospira Inc. · posted Sep 6, 2016 · Terminated
The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- PINNACLE Introducer Sheath ↗Class IIZ-2738-2016 · Terumo Medical Corporation · posted Sep 6, 2016 · Terminated
Terumo Medical is initiating this recall due to complaints of valve leakage that have been related to a manufacturing issue. If affected introducer sheaths are used, there is possible risk of incremental blood loss through the valve.
- Z-2739-2016 · Ge Healthcare It · posted Sep 6, 2016 · Terminated
There is a potential for data loss associated with empty path names resulting from HL-7 updates when the cache disk is almost full. In an unlikely event that the cache disk is almost full (<100MB) and HL-7 update events are received, the update action…
- Z-2714-2016 · Best Sanitizers Inc · posted Sep 2, 2016 · Terminated
Certain lots of the acid solution in Acecide-C are degrading faster than normal.
- Z-2715-2016 · Merge Healthcare, Inc. · posted Sep 2, 2016 · Terminated
When RadSuite is used with IPID (Issuer of Patient ID) as a part of the "Patient Identifier," it is possible in some circumstances that the demographics of one patient will be applied to a study or studies for another patient.
- Z-2716-2016 · Aribex Inc · posted Sep 2, 2016 · Terminated
Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway…
- Z-2717-2016 · Aribex Inc · posted Sep 2, 2016 · Terminated
Firm received customer complaints where end users reported that Nomad Pro and Pro 2 handsets had appeared to "overheat", "melt", "smoke" and/or "ignite." Investigation found that the internal lithium ion batteries had entered into a thermal runaway…
- Z-2719-2016 · Advanced Medical Solutions Ltd. · posted Sep 2, 2016 · Terminated
On 08 January 2015 AMS Sole Customer for the products notified AMS of the issue after discovering the misplaced IFUs during its review of packaging in its warehouse
- Z-2720-2016 · Advanced Medical Solutions Ltd. · posted Sep 2, 2016 · Terminated
On 08 January 2015 AMS Sole Customer for the products notified AMS of the issue after discovering the misplaced IFUs during its review of packaging in its warehouse
- Z-2732-2016 · Beckman Coulter Inc. · posted Sep 2, 2016 · Terminated
There is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID.
- Z-2733-2016 · Integra LifeSciences Corp. d.b.a. Integra Pain Management · posted Sep 2, 2016 · Terminated
Integra LifeSciences has been notified of a medical device recall by Adaptive Surgical, LLC involving their Riverlon" Nylon 3-0 Sutures that are packaged in Integras Cranial Access Kits. It was found that product seals may become open during extreme…
- Z-2682-2016 · Philips Electronics North America Corporation · posted Sep 1, 2016 · Terminated
Buttons or keys on the monitor allow the user to turn the volume setting down to 0", in which case the beep will not be heard at the conclusion of an exposure, as required 21 CFR 1020.31(j), and the Height-Adjustable Tabletop is missing the required labels…
- Z-2708-2016 · Alcon Research, Ltd. · posted Sep 1, 2016 · Terminated
The ULTRASERT Delivery System from certain lots have an inferior surface characteristic that could result in the IOL becoming lodged in the Delivery System
- Z-2709-2016 · Merge Healthcare, Inc. · posted Sep 1, 2016 · Terminated
A report can be confirmed with the incorrect patient demographics resulting in it being saved under the incorrect patient record in the Electronic Medical Record (EMR).
- SPECULAR MICROSCOPE CEM-530 ↗Class IIZ-2711-2016 · Nidek Inc · posted Sep 1, 2016 · Terminated
Software version 1.08 and 1.09 for the Specular Microscope CEM 530 included a change of analysis results feature that was not reviewed and approved by the FDA.
Class III recalls (6)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-2912-2016 · Siemens Healthcare Diagnostics Inc · posted Sep 30, 2016 · Terminated
Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge
- Z-2913-2016 · Siemens Healthcare Diagnostics Inc · posted Sep 30, 2016 · Terminated
Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge
- Z-2850-2016 · CooperSurgical, Inc. · posted Sep 20, 2016 · Terminated
Products contain international package insert and not the correct USA package insert
- Z-2851-2016 · CooperSurgical, Inc. · posted Sep 20, 2016 · Terminated
Products contain international package insert and not the correct USA package insert
- Trinity Biotech Captia Measles IgM, REF/Product Code 2326060,Test Kit for 96 Determinations, IVDClass IIIZ-2782-2016 · Clark Laboratories, Inc. (dba,Trinity Biotech USA) · posted Sep 12, 2016 · Terminated
The product lot failed to meet QC acceptance criteria during stability testing. The Calibrator OD was dropping causing the positive control to be invalid (running high out of range).
- VIDAS NT-pro-BNP ↗Class IIIZ-2779-2016 · Biomerieux France Chemin De L'Or · posted Sep 10, 2016 · Terminated
Due to a packaging error; The kit does not contain the correct number of C1 or C2 control vials.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.