Transpac IV Trifurcated Monitoring Kit with 84" SafeSet… recallZ-2813-2016 · ICU Medical, Inc. · initiated Aug 10, 2016
Class II Terminated All recalls posted in September 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
FDA root cause category: Process control
What the company is doing
An Urgent Medical Device Recall Notification letter dated 8/10/16 was sent to all their customers who purchased the Transpac IV with SafeSet reservoir kit and without SafeSet Reservoir, and with Transpac IV with marvelous valve without SafeSet reservoir. The letter informs the customers that ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers. The recall letter list the affected products and lot numbers. Customers are informed of the actions to be taken. Customers with questions are instructed to contact ICU Customer Service Monday through Friday between the hours of 8:30 AM and 4:00 PM Pacific time: 866-829-9025 and select option 8, or e-mail the following address: [email protected]. Customers are instructed to complete the attached Medical Device Recall Response Form and fax to (801) 264-1755 or e-mail to [email protected].
Recall details
- Recall number
- Z-2813-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 10, 2016
- Posted by the FDA
- Sep 15, 2016
- Recalling firm
- ICU Medical, Inc.
- Quantity
- 90 units
- Product code
- DRS
- FDA submissions
- K061573, K932188
- Distribution
- US Nationwide Distribution
Product description
Transpac IV Trifurcated Monitoring Kit with 84" SafeSet Reservoir, 03 ml Squeeze Flush and Needleless Valves, Item No. 42650-66 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure monitor. The SafeSet Reservoir system is a blood sampling system assembled directly to the patient monitoring line (Transpac units). The SafeSet reservoir is used on patients where periodic blood sampling is required.
Code information (models, lots, serial numbers)
lot number 3279181
Other devices under the same FDA submission
26 devices cleared under K061573, K932188 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ICU Medical 46104-07Icu Medical, Inc. · Transducer, blood-pressure, extravascular7 reports
- ICU Medical 42650-06Icu Medical, Inc. · Transducer, blood-pressure, extravascular5 reports
- ICU Medical 42648-06Icu Medical, Inc. · Transducer, blood-pressure, extravascular4 reports
- ICU Medical 42598-01Icu Medical, Inc. · Transducer, blood-pressure, extravascular3 reports
- ICU Medical 46115-65Icu Medical, Inc. · Transducer, blood-pressure, extravascular3 reports
- ICU Medical 42585-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular2 reports
- ICU Medical 42632-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular2 reports
- ICU Medical 42590-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular2 reports
- ICU Medical 852-011-46104-58Icu Medical, Inc. · Transducer, blood-pressure, extravascular2 reports
- ICU Medical 42408-11Icu Medical, Inc. · Transducer, blood-pressure, extravascular1 report
- Transpac 42587-05Icu Medical, Inc. · Catheter, continuous flush1 report
- ICU Medical 42586-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 74966).
- Z-2805-2016 · TerminatedTranspac IV Monitoring Kit with SafeSet 84inch Arterial Pressure Tubing, Reservoir, Squeeze Flush and 2 Needleless Valves, Item No. 011-42646-66 Usage: The Transpac Disposable S…
- Z-2806-2016 · TerminatedTranspac IV Trifurcated Kit With Safeset Reservoir And 2 Blood Sampling Ports, 84 inch Tubing, 3 3 ml Intraflo Flushes, 3 Disposable Transducers, Macrodrip, Item No. 42500-04 Us…
- Z-2807-2016 · TerminatedTranspac IV Bifurcated Kit with SafeSet Reservoir, 03 ml Squeeze Flush and Needleless Valve, Item No. 42500-66 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT…
- Z-2808-2016 · TerminatedTranspac IV Monitoring Kit with 2 SafeSet Blood Sampling Ports, 60" TUBING, Disposable Transducer, 3 ML Intraflo Flush, Macrodrip (Pole Mount), Item No. 42619-08 Usage: The Tran…
- Z-2809-2016 · TerminatedTranspac IV Monitoring Kit with 2 SafeSet Reservoir and 2 Blood Sampling Ports, 84" Tubing, Disposable Transducer, 3 mL Squeeze Flush, Macrodrip (Pole Mount), Item No. 42646-06 …
- Z-2810-2016 · TerminatedTranspac IV Monitoring Kit with 2 SafeSet 84" Arterial Pressure Tubing, Reservoir, 03 mL Squeeze Flush and 2 Needleless Valves, Item No. 42646-66 Usage: The Transpac Disposable …
- Z-2811-2016 · TerminatedTranspac IV Monitoring Kit with SafeSet Reservoir and Blood Sampling Port, 24" Tubing, Disposable Transducer, 03 mL Squeeze Flush, Macrodrip (Patient Mount), Item No. 42648-06 P…
- Z-2812-2016 · TerminatedTranspac IV Trifurcated Monitoring Kit with SafeSet Reservoir and 2 Blood Sampling Ports, 84", 3.3mL Squeeze Flushes, 3 Disposable Transducers, Macrodrip (Pole Mount), Item No. 426…
- Z-2814-2016 · TerminatedSingle Monitoring Kit with 30mL Flush Device, Item No. 46073-38 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transduce…
- Z-2815-2016 · TerminatedMonitoring Kit with 30mL Flush Device, Item No. 46076-15 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that …
- Z-2816-2016 · TerminatedTriple Line SafeSet Kit with 03mL Flush Device, Item No. 46076-27 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transdu…
- Z-2817-2016 · TerminatedDouble Kit with SafeSet and 03mL Flush Device, Item No. 46080-37 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transduc…
- Z-2818-2016 · TerminatedMonitoring Kit with 03mL Flush Device, Item No. 46080-57 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that …
- Z-2819-2016 · TerminatedSafeSet Kit with 30 mL Flush Device, Item No. 46083-42 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that in…
- Z-2820-2016 · TerminatedBifurcated SafeSet Kit with 03mL Flush Device, Item No. 46091-75 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transduc…
- Z-2821-2016 · TerminatedPediatric Kit with 30mL Flush Device, Item No. 46099-10 Product Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that i…
- Z-2822-2016 · Terminated5 Port OFF Manifold with Transpac IV, Item No. 46100-29 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interface…
- Z-2823-2016 · TerminatedMonitoring Kit with TP4, 30 ml Squeeze Flush and Needleless Valve, Item No. 46103-22 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular press…
- Z-2824-2016 · TerminatedTranspac IV Monitoring Kit with SafeSet 84" Arterial Pressure Tubing, Reservoir, Squeeze Flush and Needleless Valve, Item No. 46104-53 Usage: The Transpac Disposable Straight Pre…
- Z-2825-2016 · TerminatedTP4 Kit with 10cc SafeSet Reservoir, Needleless Valve, Filter Millex and Admin Set, Item No. 46104-65 Usage: The Transpac Disposable Straight Pressure Transducer (DSPT) is an e…
- Z-2826-2016 · TerminatedTranspac IV Monitoring Kit with 03 ml Squeeze Flush, SafeSet Reservoir, 2 Needleless Valves, Mounts and Bracket, Item Nos. 46107-35, 46110-07 Usage: The Transpac Disposable Stra…
- Z-2827-2016 · TerminatedTranspac IV Trifurcated Monitoring Kit with SafeSet Reservoir, Port and 03 ml Squeeze Flush Devices, Item No. 46110-84 Usage: The Transpac Disposable Straight Pressure Transduc…
- Z-2828-2016 · TerminatedMonitoring Kit with TP4, 30 ml Squeeze Flush Device, 10 cc Contamination Syringe and Needleless Valve, Item No. 46112-65 Product Usage: The Transpac Disposable Straight Pressure…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page