Siemens RAPIDPoint 400 Measurement Cartridge recallZ-2912-2016 · Siemens Healthcare Diagnostics Inc · initiated Sep 2, 2016

Class III Terminated All recalls posted in September 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens Healthcare issued An Urgent Field Safety Notice via e-mail to affected Siemens Healthcare Diagnostics Regional offices on September 2, 2016 both in the United States and outside of the United States, for communication with affected customers e: the bias of the chloride values. US accounts sent hard copy via Federal Express and requested to: Complete and return the Field Correction Effectiveness Check form attached to this letter within 7 days. Review this letter with your Medical Director. Questions:Contact your Siemens Customer Care Center or your local Siemens technical support representative.

Recall details

Recall number
Z-2912-2016
Recall class
Class III
Status
Terminated
Initiated by the company
Sep 2, 2016
Posted by the FDA
Sep 30, 2016
Recalling firm
Siemens Healthcare Diagnostics Inc
Quantity
689 cartridges
Product code
JFS
FDA submissions
K002738
Distribution
Nationwide: Foreign: Canada, Bahamas,Germany, Japan, Mexico.Belgium, Italy,Spain, Korea, France, Austria, Korea, France, Poland, Australia, china, AE, Bangladesh, Columbia, Czechkoslovakia, Bosnia

Product description

Siemens RAPIDPoint 400 Measurement Cartridge Siemens SMN numbers:10329756, 10341161

Code information (models, lots, serial numbers)

Serial Numbers:2610900214 through 2621001325 Expiry Dates: Sept 13, 2016 through Dec. 6, 2016

Other devices under the same FDA submission

1 device cleared under K002738 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.