SOMATOM Force, System x-ray, tomography, computed recallZ-0692-2017 · Siemens Medical Solutions USA, Inc · initiated Oct 20, 2016

Class II Terminated All recalls posted in December 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Siemens is providing software update version VA50A_SP3 to address the software bugs that were identified through normal field monitoring and the Global Complaint Handling Process. Correction for the problems are as follows: 1. Correction to volumetric misrepresentations of high contrast objects when using ADMIRE. 2. Correction to highly sporadic scan aborts due to temporarily tube currents at 0mA. 3. Correction to missing Microsoft Hotfixes (MS16-001, MS15-135, MS15-088, MS15-048). 4. Correction to highly sporadic image artifacts caused by error signals from high voltage chain at extremely high tube currents. 5. Correction to highly sporadic incomplete image reconstruction due to temporary slow hard disk write speed. 6. Correction to highly sporadic scan aborts due to temporarily high disk load of system disk. 7. Correction to highly sporadic image artifacts generated by scatter correction in case of tube arcings.

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens sent a Customer Information Letter dated November 2, 2016, to customers affected by the recall to inform them of a release of a software update that will provide bug-fixes to improve system performance. Customers will be contacted to arrange a date for onsite installation. For further questions, please call (610) 448-6471.

Recall details

Recall number
Z-0692-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 20, 2016
Posted by the FDA
Dec 2, 2016
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
48 units distributed in U.S.
Product code
JAK
FDA submissions
K133589
Distribution
Nationwide Distribution.

Product description

SOMATOM Force, System x-ray, tomography, computed

Code information (models, lots, serial numbers)

Model Number 10742326, Serial Numbers: 75708,75709,75607,75617,75550,75526 75512,75567,75536,75711,75514,75600,75563,75510,75454,75540,75571,75500 75527,75475,75535,75481,75482,75710,75460,75476,75588,75493,75458,75528 75467,75602,75487,75464,75478,75604,75513,75532,75555,75524,75515,75559 75610,75501,75570,75616,75576,75605

Other devices under the same FDA submission

1 device cleared under K133589 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.