Medical device recalls: December 2016
The FDA posted 71 medical device recall records in December 2016, from 45 recall events by 38 companies. 2 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Amendia, Inc · 3 events, 4 records
- PerkinElmer Health Sciences, Inc. · 3 events, 3 records
- Accuray Incorporated · 2 events, 2 records
- Medtronic Inc · 2 events, 2 records
- Synthes (USA) Products LLC · 2 events, 2 records
- Sterling Diagnostics, Inc. · 1 event, 9 records
- MicroPort Orthopedics Inc. · 1 event, 7 records
- Stryker Instruments Div. of Stryker Corporation · 1 event, 7 records
FDA root cause
- Device Design · 12
- Process change control · 11
- Process control · 9
- Storage · 7
- Packaging · 5
- Under Investigation by firm · 4
- Software design · 4
- Other · 3
Class I recalls (2)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0788-2017 · Medtronic Neuromodulation · posted Dec 19, 2016 · Terminated
Medtronic is following up to a May 2013 communication regarding the Priming Bolus function for the SynchroMed Infusion System. Medtronic is updating the Model 8870 software application card (to version AAU01) and the SynchroMed pump labeling to address…
- Z-0647-2017 · Centurion Medical Products Corporation · posted Dec 2, 2016 · Terminated
The kits contain Multi-Med Single Lumen Catheters that have a potential for excess material to remain at the tip of the catheter from the manufacturing process. It is possible that the excess material may separate from the catheter during use and pose a…
Class II recalls (49)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Specimen Gate Laboratory 5002-0180, Specimen Gate eReports 5002-0370, Software Version 1.0 ↗Class IIZ-0902-2017 · PerkinElmer Health Sciences, Inc. · posted Dec 23, 2016 · Terminated
Issue may cause eReports to consume an excessive amount of disk space causing the software to stop operating.
- Z-0886-2017 · Amendia, Inc · posted Dec 22, 2016 · Terminated
There is a problem with the mating feature between the plate and screw which could cause the plate opening not to interfere enough with the plate and head of the screw.
- Z-0887-2017 · Amendia, Inc · posted Dec 22, 2016 · Terminated
There is a problem with the mating feature between the plate and screw which could cause the plate opening not to interfere enough with the plate and head of the screw.
- Optimus Fixed Awl ↗Class IIZ-0891-2017 · Amendia, Inc · posted Dec 22, 2016 · Terminated
Complaints that the tip of the awl broke after impaction during surgery.
- Z-0892-2017 · Medtronic Inc · posted Dec 22, 2016 · Terminated
Medtronic identified an issue with specific models and lots of the PowerCross .018 Over the Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter. During an unrelated manufacturing inspection of scrapped products, an out of…
- AQURE SystemClass IIZ-0899-2017 · Radiometer America Inc · posted Dec 22, 2016 · Terminated
Design error when displaying additional information in the patient view window; error may result in misreading a parameter and its value.
- Z-0865-2017 · Accuray Incorporated · posted Dec 21, 2016 · Terminated
Accuray has become aware of a potential safety issue involving the possibility of electric shock during maintenance activities from the lon Pump HV Power Supply at the back of the modulator cabinet. This issue is limited to certain G3, G4, and VSI models…
- Z-0867-2017 · Synthes (USA) Products LLC · posted Dec 21, 2016 · Terminated
The screw rack plate was assembled incorrectly which could cause the screws in the screw rack to sit low and touch the surface beneath the screw rack.
- Z-0868-2017 · Stryker Instruments Div. of Stryker Corporation · posted Dec 21, 2016 · Terminated
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
- Z-0869-2017 · Stryker Instruments Div. of Stryker Corporation · posted Dec 21, 2016 · Terminated
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
- Z-0870-2017 · Stryker Instruments Div. of Stryker Corporation · posted Dec 21, 2016 · Terminated
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
- Z-0871-2017 · Stryker Instruments Div. of Stryker Corporation · posted Dec 21, 2016 · Terminated
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
- Z-0872-2017 · Stryker Instruments Div. of Stryker Corporation · posted Dec 21, 2016 · Terminated
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
- Z-0873-2017 · Stryker Instruments Div. of Stryker Corporation · posted Dec 21, 2016 · Terminated
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
- Z-0874-2017 · Stryker Instruments Div. of Stryker Corporation · posted Dec 21, 2016 · Terminated
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
- Z-0878-2017 · Merge Healthcare, Inc. · posted Dec 21, 2016 · Terminated
Measurements in the software are changing after saving a plan with a measurement result, which may result in potential patient injury or delay in diagnosis or treatment.
- Z-0826-2017 · Medtronic Inc · posted Dec 20, 2016 · Terminated
Medtronic is initiating a voluntary product recall for specific lot numbers of the Affinity Fusion Oxygenators with Balance1 Biosurface. These are distributed as stand-alone devices, or as a combination unit with the Affinity Fusion Cardiotomy/Venous…
- Zeus-P Lumbar Interbody Device ↗Class IIZ-0827-2017 · Amendia, Inc · posted Dec 20, 2016 · Terminated
Labeling error where three boxes in inventory had conflicting patient and primary labels. Evaluation indicated that the primary label and patient label did not match. The primary label matched the label on the inner pouch, but the patient label was incorrect.
- Z-0789-2017 · Smiths Medical ASD Inc. · posted Dec 19, 2016 · Terminated
Smiths Medical became aware that three (3) lots of Acuvance2 Safety Catheter, product number 1720-AI, contain devices with missing or incorrectly positioned side port valves.
- Z-0820-2017 · Stanmore Implants Worldwide Ltd. · posted Dec 19, 2016 · Terminated
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- Z-0821-2017 · Stanmore Implants Worldwide Ltd. · posted Dec 19, 2016 · Terminated
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- Femoral Shaft L=120mm ↗Class IIZ-0822-2017 · Stanmore Implants Worldwide Ltd. · posted Dec 19, 2016 · Terminated
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- Z-0823-2017 · Stanmore Implants Worldwide Ltd. · posted Dec 19, 2016 · Terminated
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- Z-0824-2017 · Stanmore Implants Worldwide Ltd. · posted Dec 19, 2016 · Terminated
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- Ingenia MR-OR ↗Class IIZ-0825-2017 · Philips Electronics North America Corporation · posted Dec 19, 2016 · Terminated
Possibility that a patient or user may be exposed to a very small level of touch current.
- Algovita Spinal Cord Stimulation System, Percutaneous Leads, 8-Electrode Percutaneous Lead Kit ↗Class IIZ-0781-2017 · Nuvectra · posted Dec 17, 2016 · Terminated
Nuvectra is conducting a recall due to two clinical risks that are associated with certain percutaneous leads currently being used in the Algovita Spinal Cord Stimulation System. It has been reported that in some percutaneous leads, the stylet protruded…
- IVEA 500A ↗Class IIZ-0767-2017 · Firefly Medical, Inc. · posted Dec 15, 2016 · Terminated
The dowel pin could potentially slip out of its press fit, resulting in the product potentially slumping to one side which could cause the patient to stumble.
- Z-0768-2017 · Beckman Coulter Inc. · posted Dec 15, 2016 · Terminated
Beckman Coulter has received and confirmed reports of an intermittent and unexpected behavior when loading new panels into a WalkAway instrument using LabPro Data Management System version 4.42. The issue could cause workflow interruption with a potential…
- Z-0769-2017 · Natus Medical Incorporated · posted Dec 15, 2016 · Terminated
The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by…
- Z-0758-2017 · Halyard Health · posted Dec 14, 2016 · Terminated
Defects: Product may be discolored and brittle resulting in the inability of the user to don the gloves without tearing.
- Ceramax Skin Barrier Cream: ↗Class IIZ-0763-2017 · Oculus Innovative Sciences, Inc. · posted Dec 14, 2016 · Terminated
Firm became aware that professional samples of Ceramax are not meeting stability standards set for the product.
- Z-0765-2017 · Stryker Sustainability Solutions · posted Dec 14, 2016 · Terminated
An increase in reports indicating that the reprocessed H12LP Trocar adjustable plug may break in the packaging or during use.
- Cellfina Prep Pack, Part No. CP1Class IIZ-0749-2017 · Ulthera Inc · posted Dec 13, 2016 · Terminated
It was discovered that a non-sterile vacuum tube is supplied within the pack, although the Instructions for Use (IFU) describe the use of a sterile vacuum tube as part of the procedure setup.
- Z-0751-2017 · Integra York PA, Inc · posted Dec 13, 2016 · Terminated
The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.
- Z-0752-2017 · Integra York PA, Inc · posted Dec 13, 2016 · Terminated
The Chevron seals of 6x10 Tyvek peel pouch used to package the sterile instruments may separate causing the packages to open and compromise sterility of the instrument.
- Z-0733-2017 · Cook Inc. · posted Dec 12, 2016 · Terminated
Cook Medical has received reports of the trays containing the incorrect needle, which in some cases could result in the inability to pass a wire guide through the needle. Potential adverse events that may occur as a result of utilizing needles of incorrect…
- Z-0734-2017 · Accuray Incorporated · posted Dec 12, 2016 · Terminated
Accuray has become aware of a potential safety issue involving unexpected treatment robot motion when removing an accessory. This issue is limited to certain CyberKnife Systems with software version 1 0.6.
- BD Spinal Anesthesia Tray ↗Class IIZ-0740-2017 · Becton Dickinson & Company · posted Dec 12, 2016 · Terminated
BD is conducting a voluntary product recall of the BD Spinal Anesthesia Tray, Cat (Ref) 405671, lot B01K089D since a very low number of trays are incorrectly labeled as BD Single Shot Epidural Tray, Cat (Ref) 406069.
- Z-0741-2017 · Cypress Medical Products · posted Dec 12, 2016 · Terminated
A container of McKesson vinyl gloves that failed the FDA leak test was inadvertently distributed.
- Z-0746-2017 · Steris Corporation · posted Dec 12, 2016 · Terminated
The Drager Pendula spring arm device has a set of screws located on the spring arm that were not properly secured with Loctite and may become loose.
- Halyard C-Section Drape with Clear Screen and Full Incise (Non-Sterile), Product Code 44965 NS ↗Class IIZ-0724-2017 · Halyard Health, Inc · posted Dec 8, 2016 · Terminated
Drapes within the affected lots may include a manufacturing variation that prevents convenient removal of the coated release paper, which inhibits easy access to the surgical site.
- Z-0727-2017 · Amd-Ritmed, Inc. · posted Dec 8, 2016 · Terminated
The lot of transparent dressing was found to be contaminated (not sterile) based on FDA sampling and analysis.
- Z-0701-2017 · 3M Company - Health Care Business · posted Dec 3, 2016 · Terminated
3M is recalling Universal Electrosurgical Pads because a report of an unintended material (process liner) in the product. This may prevent or impede the safe return of electrosurgical current following Electrosurgical Unit (ESU) activation. This defect…
- Z-0703-2017 · DePuy Orthopaedics, Inc. · posted Dec 3, 2016 · Terminated
DePuy Orthopaedics, Inc. is voluntarily recalling one lot of the LCS Complete Knee Revision System VVC Insert Small 22.5mm, due to the insert being manufactured outside of specifications, which prevents the insert from properly seating in the mating…
- Z-0704-2017 · Ultradent Products, Inc. · posted Dec 3, 2016 · Terminated
There was a labeling error with product, UltraSeal XT Plus Clear Refill, Part Number (PN) 565, Lot Number (Lot) BD79T. The package was mislabeled as PQ1, PN 641; while the barcode, IFU and products were labeled for UltraSeal XT Plus Clear Refill, PN 565,…
- Z-0692-2017 · Siemens Medical Solutions USA, Inc · posted Dec 2, 2016 · Terminated
Siemens is providing software update version VA50A_SP3 to address the software bugs that were identified through normal field monitoring and the Global Complaint Handling Process. Correction for the problems are as follows: 1. Correction to volumetric…
- Z-0681-2017 · Synthes (USA) Products LLC · posted Dec 1, 2016 · Terminated
Package may be incorrectly labeled as containing 0.76 mm Drill Bit, but in fact contain a 1.1 mm diameter Drill Bit.
- STCKERT HeaterCooler System 3T ↗Class IIZ-0682-2017 · Sorin Group USA, Inc. · posted Dec 1, 2016 · Terminated
On August 1, 2014 Sorin Group USA, Inc. informed consignees via letter to review the disinfection and monitoring practices and to strictly adhere to the instructions in the Instructions for Use (IFU).
- Z-0690-2017 · Domain Surgical, Inc. · posted Dec 1, 2016 · Terminated
Domain Surgical, Inc. announces a voluntary field action for the FMsealer Open Shears due to fixed jaw fractures during clincal usage.
Class III recalls (20)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0929-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0930-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0931-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0932-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0933-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0934-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0935-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0936-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0937-2017 · Sterling Diagnostics, Inc. · posted Dec 29, 2016 · Terminated
Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits.
- Z-0879-2017 · OrthoPediatrics Corp · posted Dec 22, 2016 · Terminated
Labeled as a 48mm Cortical Screw but measuring at 50mm
- EnLite Neonatal TREC Kit ↗Class IIIZ-0897-2017 · PerkinElmer Health Sciences, Inc. · posted Dec 22, 2016 · Terminated
Unique Device Identifier (UDI) is missing from outer kit box label.
- Z-0779-2017 · PerkinElmer Health Sciences, Inc. · posted Dec 16, 2016 · Terminated
A DELFIA hFSH kit lot failed real time stability testing. The real time stability data demonstrated a decrease in hFSH concentrations between -4.8% to -11.3% for control sera. The -11.3% bias was observed at an hFSH target concentration of 101 IU/L. The…
- Z-0722-2017 · OraSure Technologies, Inc. · posted Dec 8, 2016 · Terminated
Shipping error. Cocaine metabolite Microplate distributed instead of Barbiturate Microplate due to incorrect selection of Microplate during manufacturing.
- Lineage Liners: ↗Class IIIZ-0683-2017 · MicroPort Orthopedics Inc. · posted Dec 1, 2016 · Terminated
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected…
- Dynasty Liners: ↗Class IIIZ-0684-2017 · MicroPort Orthopedics Inc. · posted Dec 1, 2016 · Terminated
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected…
- Procotyl (Rim- Lock) Liners: ↗Class IIIZ-0685-2017 · MicroPort Orthopedics Inc. · posted Dec 1, 2016 · Terminated
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected…
- Z-0686-2017 · MicroPort Orthopedics Inc. · posted Dec 1, 2016 · Terminated
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected…
- Z-0687-2017 · MicroPort Orthopedics Inc. · posted Dec 1, 2016 · Terminated
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected…
- Z-0688-2017 · MicroPort Orthopedics Inc. · posted Dec 1, 2016 · Terminated
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected…
- Cobalt Chromium Heads ↗Class IIIZ-0689-2017 · MicroPort Orthopedics Inc. · posted Dec 1, 2016 · Terminated
Revision rate trends from within the MicroPort Orthopedics Inc. database evaluated for the modular head Metal on Metal THA systems show an increasing overall trend from 2009 to present and it was found that there was a specific hazard/harm for suspected…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.