Artis zee/zeego, Angiographic x-ray system recallZ-0602-2017 · Siemens Medical Solutions USA, Inc · initiated Nov 1, 2016
Class II Terminated All recalls posted in November 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The possibility exists that the semi-automatic focus switch may not work as intended when pressing the footswitch multiple times.
FDA root cause category: Software in the Use Environment
What the company is doing
Siemens mailed a Customer Safety Advisory Notice to affected customers to inform them of a corrective action they plan to perform to prevent a possible hazard to persons and equipment. The notice identified the associated risks, underlying issues, impact on system operation and actions to be taken.
Recall details
- Recall number
- Z-0602-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 1, 2016
- Posted by the FDA
- Nov 17, 2016
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 4,657 systems Worldwide
- Product code
- OWB
- FDA submissions
- K141574
- Distribution
- US Nationwide Distribution.
Product description
Artis zee/zeego, Angiographic x-ray system Product Usage: The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego and Q/ Q.zen include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee / zeego and Q/ Q.zen can also support the acquisition of position triggered imaging for spatial data synthesis.
Code information (models, lots, serial numbers)
Model Numbers: 10094135, 10094137, 10094139, 10094141, 10280959
Other devices under the same FDA submission
7 devices cleared under K141574 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Artis zee floor 10094135Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls2 deaths94 reports
- Artis zee ceiling 10094137Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls91 reports
- Artis zeego 10280959Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system4 recalls1 death24 reports
- Artis zee biplane 10094141Siemens Healthcare GmbH · System, x-ray, angiographic3 recalls63 reports
- Artis zee multi-purpose 10094139Siemens Healthcare GmbH · System, x-ray, tomography, computed3 recalls26 reports
- Artis zee biplane MN 10094143Siemens Healthcare GmbH · System, x-ray, tomography, computed2 recalls3 reports
- Artis zee floor MN 10094142Siemens Healthcare GmbH · System, x-ray, angiographic2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 75606).
- Z-0601-2017 · TerminatedArtis Q, Angiographic x-ray system Product Usage: The Artis systems are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and intervent…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page