PowerCross recallZ-0892-2017 · Medtronic Inc · initiated Nov 10, 2016
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic identified an issue with specific models and lots of the PowerCross .018 Over the Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Dilatation Catheter. During an unrelated manufacturing inspection of scrapped products, an out of specification condition was observed. The catheter outer shafts were pinched and deformed at the proximal balloon bond. Bench testing of affected devices showed that this resulted in either the inability to inflate the device, a slow deflation, or the inability to deflate the device post expansion.
FDA root cause category: Process control
What the company is doing
Medtronic verbally communicated about the issue on November 20. 2016, to affected customers. Additionally, consignees were hand delivered by Field Sales Representatives a Medtronic "Urgent Medical Device Recall" letter dated November 2016. The letter was addressed to Risk Manger or Health Care Professional. The letter described the problem and the product involved in the recall. Advised consignees to immediately identify, quarantine the affected product, and return it to Medtronic. Requested consignees to complete and email the Customer Confirmation Certificate to [email protected] or fax it to Medtronic at 651-367-0612, attention: Customer Focused Quality. For questions, contact your Medtronic representative. If you require assistance in contacting your representative, please contact Medtronics National Answering Service at 800-633-8766.
Recall details
- Recall number
- Z-0892-2017
- Status
- Terminated
- Initiated by the company
- Nov 10, 2016
- Posted by the FDA
- Dec 22, 2016
- Recalling firm
- Medtronic Inc
- Quantity
- 315
- Product code
- LIT
- FDA submissions
- K093286
- Distribution
- NY, NC, WA
Product description
ev3, PowerCross .018 OTW PTA Dilatation Catheter, Sterile EO. The PowerCross .018 OTW PTA Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infrapopliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.
Code information (models, lots, serial numbers)
UDI/Product Lot/Serial Product Number UPN/GTIN Number 00821684046647 A358626 AB18W030100150 00821684046883 A357386 AB18W060100150 00821684046722 A357070 AB18W040100150 00821684046777 A358043 AB18W050040150 00821684046722 A356360 AB18W040100150 00821684046777 A357348 AB18W050040150 00821684046562 A359219 AB18W025040150
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- PowerCross AB18W060100150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 recall1 report
- PowerCross AB18W025040150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 recall0 reports
- PowerCross AB18W030100150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 recall0 reports
- PowerCross AB18W040100150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 recall0 reports
- PowerCross AB18W050040150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 recall0 reports
Product lines
Other devices under the same FDA submission
7 devices cleared under K093286 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- PowerCross AB18W060060150Covidien LP · Catheter, angioplasty, peripheral, transluminal2 reports
- PowerCross AB18W020100090Covidien LP · Catheter, angioplasty, peripheral, transluminal1 report
- PowerCross AB18W040040090Covidien LP · Catheter, angioplasty, peripheral, transluminal1 report
- PowerCross AB18W030060150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 report
- PowerCross AB18W030120150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 report
- PowerCross AB18W050150150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 report
- PowerCross AB18W050200150Covidien LP · Catheter, angioplasty, peripheral, transluminal1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page