Oxford Fixed Lateral Bearing Size F Right, Part Number… recallZ-0956-2017 · Zimmer Biomet, Inc. · initiated Dec 6, 2016

Class II Terminated All recalls posted in January 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Zimmer Biomet is initiating a removal of a single lot of Oxford Fixed Lateral Bearings due to mislabeling. Zimmer Biomet received product complaints reporting that when the package labeled as a Right was opened it actually contained a Left Bearing.

FDA root cause category: Employee error

What the company is doing

Zimmer Biomet sent an Urgent Medical Device Recall letter dated December 6, 2016. to all affected customers for a single lot of Oxford Fixed Lateral Bearings via e-mailed letter. Zimmer Biomet received two product complaints reporting that when the package labeled as a Right was opened it actually contained a Left Bearing. Risks: If another bearing is not available, a delay greater than 30 minutes may occur. If the product of the incorrect orientation is implanted, premature revision may be required. Customers were instructed to review the recall notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of the Inventory Return Certification Form a. Return a digital copy within three (3) days via email to [email protected]. 4. Immediately return all affected product from your distributorship and affected hospitals within your territory along with a completed Inventory Return Certification Form to Zimmer Biomet. Request a Return Authorization Number via email at [email protected] . Be sure to specify RECALL as the RGA type. b. For each return, send a copy of Inventory Return Certifictaion Form via email to [email protected]. c. Include a hardcopy of Inventory Return Certifictaion Form to Zimmer Biomet with your shipment for immediate processing. d. Mark the outside of the returns box(es) clearly with RECALL. 5. If after reviewing this notice you have further questions or concerns please call 1-877-946-2761 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of business hours will receive a prompt to record a voicemail and will be returned upon receipt. Alternatively, your questions may be sent by email to [email protected]. Customer Service Zimmer Biomet, Inc. P.O. Box 708 Warsaw, I

Recall details

Recall number
Z-0956-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 6, 2016
Posted by the FDA
Jan 6, 2017
Recalling firm
Zimmer Biomet, Inc.
Quantity
12 units distributed
Product code
HRY
FDA submissions
K133940
Distribution
Domestic: None Froeign: Netherlands & Switzerland VA/DOD: None

Product description

Oxford Fixed Lateral Bearing Size F Right, Part Number 154375 Lot 744260, Sterile, The product is placed into an Aluminum Oxide coated PET pouch and vacuum sealed with an Argon flush. The vacuum sealed part is then placed into a coated 1073B Tyvek/film pouch and heat sealed. This pouch acts as the sterile barrier. The pouch is then labeled and placed into a carton with an IFU. The carton is then labeled and shrink wrapped. The pouch and carton labels also contain sterilization indicator dots. Partial replacement of the articulating surfaces of the knee when only one side of the knee joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous partial arthroplasty in the affected compartment. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implanted with bone cement. The Oxford Femoral components are indicated for use in the lateral compartment for a fixed bearing application. They are intended to be implanted with bone cement

Code information (models, lots, serial numbers)

Oxford Fixed Lateral Bearing Size F Right, Part Number 154375 Lot 744260, Date of Mfg. 26 Apr 2015

Other devices under the same FDA submission

11 devices cleared under K133940 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.