Medical device recalls: January 2017
The FDA posted 100 medical device recall records in January 2017, from 65 recall events by 48 companies. 7 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Endologix · 7 events, 8 records
- The Binding Site Group, Ltd. · 5 events, 11 records
- GE OEC Medical Systems, Inc · 2 events, 5 records
- Bard Access Systems · 2 events, 4 records
- Beckman Coulter Inc. · 2 events, 3 records
- ICU Medical, Inc. · 2 events, 2 records
- Implant Direct Sybron Manufacturing, LLC · 2 events, 2 records
- Roche Diagnostics Operations, Inc. · 2 events, 2 records
FDA root cause
- Device Design · 23
- Nonconforming Material/Component · 15
- Under Investigation by firm · 13
- Process control · 7
- Software design · 5
- Use error · 4
- Component design/selection · 4
- Mixed-up of materials/components · 3
Class I recalls (7)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0950-2017 · CareFusion 303, Inc. · posted Jan 18, 2017 · Terminated
During an infusion, a false Air-in-Line (AIL) alarm would cause the infusion to be interrupted.
- Z-0968-2017 · Bard Peripheral Vascular, Inc · posted Jan 18, 2017 · Terminated
Complaints of sheath separation, kinking, and/or tip damage during use.
- Z-0969-2017 · Bard Peripheral Vascular, Inc · posted Jan 18, 2017 · Terminated
Complaints of sheath separation, kinking, and/or tip damage during use.
- Z-0970-2017 · Bard Peripheral Vascular, Inc · posted Jan 18, 2017 · Terminated
Complaints of sheath separation, kinking, and/or tip damage during use.
- Z-0976-2017 · BioMerieux SA · posted Jan 13, 2017 · Terminated
A drift of the performance with the BK test (BK virus, worst case application) was observed for some silica batches of Magnetic Silica (MagSil). This is the same issue as in recall Z-2342-2016.
- Z-0977-2017 · BioMerieux SA · posted Jan 13, 2017 · Terminated
A drift of the performance with the BK test (BK virus, worst case application) was observed for some silica batches of Magnetic Silica (MagSil). This is the same issue as in recall Z-2342-2016.
- Z-0926-2017 · Nurse Assist, Inc · posted Jan 3, 2017 · Terminated
Potential contamination with B. cepacia.
Class II recalls (92)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- SwishTapered ImplantClass IIZ-1112-2017 · Implant Direct Sybron Manufacturing, LLC · posted Jan 28, 2017 · Terminated
The extender product code (9037-03) that is packaged with the SwishTapered (Part Number 933308; Lot Number 73241) Implant was packaged in the incorrect position. This discrepancy may lead to the possibility of extender not seating properly on the implant,…
- Z-1114-2017 · Zimmer Biomet, Inc. · posted Jan 28, 2017 · Terminated
labeling for specific lots of the Standard OSS Modular Proximal Tibia incorrectly identifies the components as AVL Modular Proximal Tibia. The mislabeling could have the potential risks: A delay in surgery less than 30 minutes. Patient undergoes a revision…
- Z-1107-2017 · PerkinElmer Life and Analytical Sciences, Wallac, OY · posted Jan 27, 2017 · Terminated
The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the…
- Spare Disk Remover Module ↗Class IIZ-1108-2017 · PerkinElmer Life and Analytical Sciences, Wallac, OY · posted Jan 27, 2017 · Terminated
The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the…
- Z-1109-2017 · Immuno-Mycologics, Inc · posted Jan 27, 2017 · Terminated
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
- Z-1110-2017 · Biom'up · posted Jan 27, 2017 · Terminated
Lack of sterility assurance
- Z-1099-2017 · Roche Diabetes Care, Inc. · posted Jan 26, 2017 · Terminated
iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user. iOS only: Pairing and using multiple meters with the…
- NORMFlow H-110 Irrigation Fluid Warmer, Continental Europe, Model H-1100, Reorder H-1100-INT-230. ↗Class IIZ-1105-2017 · Smiths Medical ASD, Inc. · posted Jan 26, 2017 · Terminated
Smiths Medical became aware that that one (1) NORMOFLO H-1100 Irrigation Fluid Warmers was incorrectly labelled with Spanish language labels instead of English labels. Specifically five (5) labels were incorrect; the Operators Manual, Outer Base Box…
- Z-1090-2017 · Roche Diagnostics Corporation · posted Jan 24, 2017 · Terminated
- Z-1092-2017 · Roche Diagnostics Operations, Inc. · posted Jan 24, 2017 · Terminated
Tina-quant Cardiac C-reactive Protein (Latex) High Sensitive (CRPHS) lot number 604450 (catalog number 11972855216) shows an under-recovery of up to -25% at concentrations below 5 mg/L on the MODULAR ANALYTICS P module. this issue can lead to erroneous…
- Z-1067-2017 · The Binding Site Group, Ltd. · posted Jan 23, 2017 · Terminated
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
- Z-1068-2017 · The Binding Site Group, Ltd. · posted Jan 23, 2017 · Terminated
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
- Z-1069-2017 · The Binding Site Group, Ltd. · posted Jan 23, 2017 · Terminated
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
- Z-1070-2017 · The Binding Site Group, Ltd. · posted Jan 23, 2017 · Terminated
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
- Z-1071-2017 · The Binding Site Group, Ltd. · posted Jan 23, 2017 · Terminated
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
- Z-1072-2017 · The Binding Site Group, Ltd. · posted Jan 23, 2017 · Terminated
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
- MininephThl Human lgG4 Kit ↗Class IIZ-1073-2017 · The Binding Site Group, Ltd. · posted Jan 23, 2017 · Terminated
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
- Cobas b 123 POC system ↗Class IIZ-1077-2017 · Roche Diagnostics Operations, Inc. · posted Jan 23, 2017 · Terminated
under specific settings, an issue may occur during simultaneous Sensor Cartridge and Fluid Pack change on the cobas b 123 <2> POC system and cobas b 123 <4> POC system. The issue occurs when the software function [AutoQC as follow-up] is configured to run…
- Z-1078-2017 · Sekisui Diagnostics P.E.I. Inc. · posted Jan 23, 2017 · Terminated
Certain lots are showing an atypical decrease in optical density (OD) over time, which may result in an impact to performance at the high end of the linear range.
- Z-0922-2017 · Philips Electronics North America Corporation · posted Jan 20, 2017 · Terminated
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
- Z-0923-2017 · Philips Electronics North America Corporation · posted Jan 20, 2017 · Terminated
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
- Z-0924-2017 · Philips Electronics North America Corporation · posted Jan 20, 2017 · Terminated
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
- Z-0925-2017 · Philips Electronics North America Corporation · posted Jan 20, 2017 · Terminated
Due to a production error, screws were not tightened to specified torque. These screws may come loose from the C-arm shaft causing the C-arm to rotate uncontrolled.
- Z-1057-2017 · On-X Life Technologies, Inc. · posted Jan 20, 2017 · Terminated
Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plastic particulate was found within sealed primary packages, 3) Loose hair was discovered within four…
- Z-1059-2017 · GE OEC Medical Systems, Inc · posted Jan 20, 2017 · Terminated
GE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component replacement frequencies.
- Z-1060-2017 · GE OEC Medical Systems, Inc · posted Jan 20, 2017 · Terminated
GE Healthcare Surgery announces a voluntary field action for the OEC 9800 and OEC 9900 Elite due to additional supplementation to the user documentation to include areas for enhanced planned maintenance and component replacement frequencies.
- Z-1039-2017 · Endologix · posted Jan 19, 2017 · Terminated
Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
- Z-1047-2017 · Endologix · posted Jan 19, 2017 · Terminated
Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
- AFX Endovascular AAA SystemClass IIZ-1048-2017 · Endologix · posted Jan 19, 2017 · Terminated
Endologix updated information on the rates of Type III endoleaks and suggestions for patient surveillance and treatment.
- AFX2 Endovascular AAA SystemClass IIZ-1049-2017 · Endologix · posted Jan 19, 2017 · Terminated
Endologix updated information on the rates of Type III endoleaks and suggestions for patient surveillance and treatment.
- Human Lambda Free SPAPlus Kit ↗Class IIZ-1050-2017 · The Binding Site Group, Ltd. · posted Jan 19, 2017 · Terminated
Calibration curve activity has increased over time in the kit lots listed.
- Z-1051-2017 · Great Basin Scientific, Inc. · posted Jan 19, 2017 · Terminated
The Great Basin Staph ID/R Panel is being recalled because the S. aureus probe could potentially result in a false positive.
- Z-1052-2017 · Carestream Health Inc · posted Jan 19, 2017 · Terminated
Software error; Carestream Health Inc, received a complaint stating that when a user accidentally obtains a measurement value of 0 and corrects the value in the report, the resulting measurement unit is not displayed, i.e., centimeters or millimeters. As…
- Z-1053-2017 · The Binding Site Group, Ltd. · posted Jan 19, 2017 · Terminated
A change in the calibration curve causing an increase in false prozone flags.
- Z-1054-2017 · Endologix · posted Jan 19, 2017 · Terminated
Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
- Z-1055-2017 · The Binding Site Group, Ltd. · posted Jan 19, 2017 · Terminated
If a control ring is marked after reading, the software will not flag results that are out of the specified QC range.
- Z-1056-2017 · Baxter Healthcare Corporation · posted Jan 19, 2017 · Terminated
Customer complaints received for the presence of leaks
- Z-1035-2017 · Endologix · posted Jan 18, 2017 · Terminated
Endologix initiated a correction by updating reports for the cumulative clinical data and information on the Powerlink System and the AFX Endovascular AAA System (AFX System).
- Z-1037-2017 · Endologix · posted Jan 18, 2017 · Terminated
Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
- Z-1038-2017 · Endologix · posted Jan 18, 2017 · Terminated
Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
- Z-1043-2017 · The Binding Site Group, Ltd. · posted Jan 18, 2017 · Terminated
Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when…
- Ambra PACS ↗Class IIZ-1045-2017 · Dicom Grid Inc · posted Jan 18, 2017 · Terminated
A software error caused the window/level to become the same in one series regardless if the image had different levels; image results have a washed-out grey appearance.
- Z-1033-2017 · Meridian Bioscience Inc · posted Jan 17, 2017 · Terminated
Internal testing has demonstrated that the indicated kit lots produce false positive results when tested with some broth specimens.
- Z-1034-2017 · Ultradent Products, Inc. · posted Jan 17, 2017 · Terminated
Ultradent Products, Inc. announces a voluntary field action for the NanoTips 30g 20pk, 100pk because the cannula attached to the tip hub may become loose or detach from the blue base that connects to the syringe.
- Z-1023-2017 · Medacta Usa Inc · posted Jan 13, 2017 · Terminated
Damaged packaging, lack of sterility assurance
- Z-1007-2017 · Medtronic Navigation, Inc. · posted Jan 11, 2017 · Terminated
Medtronic Navigation, Inc. announces a voluntary field action for the AxIEM Touch N Go Pointer Probe because the patient registration accuracy may compromise navigation accuracy when using the StealthStation System with Touch N Go Pointer Probes for…
- Z-1010-2017 · Symmetry Surgical, Inc. · posted Jan 11, 2017 · Terminated
Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
- Z-1011-2017 · Symmetry Surgical, Inc. · posted Jan 11, 2017 · Terminated
Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
- Z-1012-2017 · Symmetry Surgical, Inc. · posted Jan 11, 2017 · Terminated
Sealant on the base trim of sterilizer container may be deteriorated or absent which could provide a path for contaminants to enter the container under certain conditions post-sterilization.
- Z-1013-2017 · Beckman Coulter Inc. · posted Jan 11, 2017 · Terminated
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. This is due…
- Z-1014-2017 · Beckman Coulter Inc. · posted Jan 11, 2017 · Terminated
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. This is due…
- Z-1016-2017 · Hospira Inc. · posted Jan 11, 2017 · Terminated
Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the…
- Z-0980-2017 · ICU Medical, Inc. · posted Jan 10, 2017 · Terminated
ICU Medical Inc. has identified an isolated event involving a 15 drop drip chamber assembled on a device which should contain a 60 drop drip chamber.
- Z-0981-2017 · Bard Access Systems · posted Jan 10, 2017 · Terminated
Bard Access Systems announces a voluntary field action for the StatLock Catheter Stabilization Devices because some single unit pouches have been released without unique product identifiers (product code, lot number, expiration date, etc.).
- Z-0982-2017 · Bard Access Systems · posted Jan 10, 2017 · Terminated
Bard Access Systems announces a voluntary field action for the StatLock Catheter Stabilization Devices because some single unit pouches have been released without unique product identifiers (product code, lot number, expiration date, etc.).
- Algerbrush-II ↗Class IIZ-0983-2017 · Bausch & Lomb, Inc. · posted Jan 10, 2017 · Terminated
According to firm ( Bausch & Lomb, Inc. ) on June 17, 2016 a packaging operator notified his supervisor he had not placed the operating instruction into the last repackaging of SKU 0816 C (Pterygium Burr Medium Diamond). Further investigation found that…
- Z-0985-2017 · Bard Access Systems · posted Jan 10, 2017 · Terminated
Bard Access Systems announces a voluntary field action for the SafeStep Huber Needle Set with Y-Injection Site and MiniLoc Safety Infusion Set Winged without Y-Injection Site devices because sets may contain one or more units where the needle length and…
- Z-0986-2017 · Bard Access Systems · posted Jan 10, 2017 · Terminated
Bard Access Systems announces a voluntary field action for the SafeStep Huber Needle Set with Y-Injection Site and MiniLoc Safety Infusion Set Winged without Y-Injection Site devices because sets may contain one or more units where the needle length and…
- Z-0988-2017 · ICU Medical, Inc. · posted Jan 10, 2017 · Terminated
ICU Medical Inc. has identified a potential risk of leaking with certain Tego Connector devices.
- i.v. Station and i.v. Station ONCO Automated Drug Compounding System, Pharmacy Compounding System ↗Class IIZ-0989-2017 · Health Robotics S.R.L. · posted Jan 10, 2017 · Terminated
The firm has initiated a product correction due to a bag carousel motor error causing misalignment and resulting in a mislabeling of final IV bag preparations.
- Z-0990-2017 · Varian Medical Systems, Inc. · posted Jan 10, 2017 · Terminated
Modifications in version 13.6MR2 for Contouring, SmartAdapt, and SmartSegmentation workspaces resulted in contours not being saved consistently in Eclipse. Treatment Planning System. The issue only occurs if certain conditions are fulfilled.
- Z-0991-2017 · ConforMIS, Inc. · posted Jan 10, 2017 · Terminated
iView report displayed femoral cut measurements in the incorrect unit of measure (inches vs. mm).
- Z-0992-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-0993-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-0994-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-0995-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-0996-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-0997-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-0998-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-0999-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-1000-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-1001-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- Z-1002-2017 · Siemens Medical Solutions USA Inc. · posted Jan 10, 2017 · Terminated
Siemens Medical Solutions, Molecular Imaging has become aware, through our supplier that some laser cover windows may fall off.
- GoDirect Screw-Retaining TopClass IIZ-0961-2017 · Implant Direct Sybron Manufacturing, LLC · posted Jan 9, 2017 · Terminated
Consignees received product 6060-62, Replant Overdenture Abutment with the label for 4500-61 GoDirect Screw-Retaining Top.
- Z-0965-2017 · Advanced Sterilization Products · posted Jan 9, 2017 · Terminated
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor System because the hose may come loose and cause a leak.
- Z-0966-2017 · Amendia, Inc · posted Jan 9, 2017 · Terminated
Absent tantalum market pin.
- Z-0973-2017 · GE OEC Medical Systems, Inc · posted Jan 9, 2017 · Terminated
GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
- Z-0974-2017 · GE OEC Medical Systems, Inc · posted Jan 9, 2017 · Terminated
GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
- Z-0975-2017 · GE OEC Medical Systems, Inc · posted Jan 9, 2017 · Terminated
GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
- Z-0978-2017 · Alcon Research, Ltd. · posted Jan 9, 2017 · Terminated
The intraocular lens could become lodged within the cartridge
- Z-0963-2017 · Stryker Sustainability Solutions · posted Jan 7, 2017 · Terminated
A breach in the sterile packaging may have occurred resulting in potentially non-sterile product.
- Z-0964-2017 · Stryker Sustainability Solutions · posted Jan 7, 2017 · Terminated
A breach in the sterile packaging may have occurred resulting in potentially non-sterile product.
- Z-0952-2017 · In2Bones, SAS · posted Jan 6, 2017 · Terminated
Dimensional non-conformity of the thread on the screw head.
- NEO SCREW 02.7-14MM LOCKING, REF W27 ST114, In2Bones, 28 chemin du petit bois, 69130 Ecully - FranceClass IIZ-0953-2017 · In2Bones, SAS · posted Jan 6, 2017 · Terminated
Dimensional non-conformity of the thread on the screw head.
- Z-0954-2017 · In2Bones, SAS · posted Jan 6, 2017 · Terminated
Dimensional non-conformity of the thread on the screw head.
- Z-0955-2017 · Voco GmbH · posted Jan 6, 2017 · Terminated
Labeling mix-up: Futurabond M+ DCA (Dual Cure Activator) was marked labeled as Futurabond M+ Universal Adhesive.
- Z-0956-2017 · Zimmer Biomet, Inc. · posted Jan 6, 2017 · Terminated
Zimmer Biomet is initiating a removal of a single lot of Oxford Fixed Lateral Bearings due to mislabeling. Zimmer Biomet received product complaints reporting that when the package labeled as a Right was opened it actually contained a Left Bearing.
- Z-0960-2017 · Trilliant Surgical Ltd. · posted Jan 6, 2017 · Terminated
Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.
- Z-0949-2017 · Merit Medical Systems, Inc. · posted Jan 5, 2017 · Terminated
Recalling dilators included with the 6F PreludeEASE Hydrophilic Sheath Introducer because the dilator tips were manufactured with an undersized inner diameter and therefore will not allow passage of an 0.021 (0.53 mm) guide wire.
- Z-0941-2017 · Interventional Spine Inc · posted Jan 3, 2017 · Terminated
During implantation, the proximal end of the device could separate if the angle between the device and insertion instrument is too steep. This separation will prevent the device from expanding.
- Z-0942-2017 · Pride Mobility Products Corp · posted Jan 3, 2017 · Terminated
The interface between the TB3 back and tilt base can separate when tilted beyond the specified angle.
- PROFEMUR(R) R RASP HANDLE, REF PPW38078, 1 EACH, NON-STERILE, MicroPort Orthopedics, Arlington, TN ↗Class IIZ-0943-2017 · MicroPort Orthopedics Inc. · posted Jan 3, 2017 · Terminated
The recall is being initiated because all broach handles in the affected lot had the same non-conformity with the connection feature. The broach handles may have machining issues that do not allow the handle to properly mate with an associated broach.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1040-2017 · Beckman Coulter Inc. · posted Jan 18, 2017 · Terminated
Beckman Coulter initiated a design change to update the Automate PC image to accommodate the operating system change to Windows 10.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.