RingLoc Hip System recallZ-1024-2017 · Zimmer Biomet, Inc. · initiated Dec 19, 2016
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product complaints indicated that the part and lot numbers on the packaging differed from the physical product inside the packaging. Risk of delay in surgery.
FDA root cause category: Under Investigation by firm
What the company is doing
Zimmer Biomet sent an Urgent Medical Device Recall (REMOVAL) LOT SPECIFIC notifications dated December 19, 2016, to the affected distributors and Risk Managers and Distributors via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Risk Manager Responsibilities: 1. Review this notification and ensure affected personnel are aware of the contents. 2. Assist your Zimmer Biomet sales representative quarantine all affected product. 3. Your Zimmer Biomet sales representative will remove the affected product from your facility. 4. Complete Attachment 1 Certificate of Acknowledgement. a. Return a digital copy to [email protected]. b. Retain a copy of the Acknowledgement Form with your recall records in the event of a compliance audit of your facilities documentation. 5. If after reviewing this notice you have further questions or concerns please call 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of the call center operating hours will receive a prompt to record a voicemail or be transferred to an on-call representative in the case of an emergency. Alternatively, your questions may be sent by email to [email protected]. Distributor Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form a. Return a digital copy to [email protected] within three (3) days. 4. Immediately return all affected product from your distributorship and affected hospitals within your territory along with a completed Attachment 1 Inventory Return Certification
Recall details
- Recall number
- Z-1024-2017
- Status
- Terminated
- Initiated by the company
- Dec 19, 2016
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 36
- Product code
- LPH
- FDA submissions
- K110555
- Distribution
- USA (nationwide) and Internationally to ARGENTINA
Product description
RingLoc ArComXL Highly Crosslinked Polyethylene Liners ArComXL High Wall +3mm Liners Size 23 item XL-108323 ArComXL MaxRom +3mm Liners Size 23 item XL-108223
Code information (models, lots, serial numbers)
Item XL-108323 Lot 761680 UDI (01)00880304651524(17)210713(10)761680 Item XL-108223 Lot 419460 UDI (01)00880304651494(17)210713(10)419460
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- RingLoc® Hip System XL-108223Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 recall0 reports
- RingLoc® Hip System XL-108323Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 recall0 reports
Product lines
Other devices under the same FDA submission
16 devices cleared under K110555 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Active Articulation™ Dual Mobility Hip System XL-200152Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall0 reports
- Active Articulation Hip System EP-200148Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented19 reports
- Active Articulation Hip System EP-200146Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED18 reports
- Active Articulation Hip System EP-200152Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented17 reports
- Active Articulation Hip System EP-200150Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented14 reports
- Active Articulation Hip System EP-200144Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented9 reports
- Active Articulation™ Dual Mobility Hip System XL-200150Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented3 reports
- Active Articulation 010000991Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented3 reports
- Active Articulation Hip System EP-200154Biomet Orthopedics, LLC · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Active Articulation Hip System EP-200156Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented2 reports
- Active Articulation™ Dual Mobility Hip System XL-200156Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented2 reports
- Active Articulation Hip System EP-200160Biomet Orthopedics, LLC · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 75970).
- Z-1025-2017 · TerminatedActive Articulation ArComXL Polyethylene Bearings Active Articulation ArComXL Bearings item XL-200152
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page