Skip to content

The OEC 9900 Elite mobile fluoroscopy system is designed to… recallZ-0975-2017 · GE OEC Medical Systems, Inc · initiated Dec 19, 2016

Class II Terminated All recalls posted in January 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE OEC Medical Systems, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.

FDA root cause category: Use error

What the company is doing

GE Healthcare sent an Urgent Medical Device Correction letter dated December 19, 2016, to all affected consignees. GE informed consignees that GE will correct all affected products. If consignnes have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative. For questions regarding this recall call 801-536-4952.

Recall details

Recall number
Z-0975-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 19, 2016
Posted by the FDA
Jan 9, 2017
Recalling firm
GE OEC Medical Systems, Inc
Product code
OXO
FDA submissions
K120613, K122234
Distribution
Worldwide Distribution - US including U.S. States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PUERTO RICO, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Internationally to Canada, Mexico, and Costa Rica

Product description

The OEC 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedure. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedure. The system may be used for other imaging applications at the physician's discretionOEC 9900 Elite. MDL Numbers: D148942, D155043

Code information (models, lots, serial numbers)

Manufactured between 1/1/16 to 11/1/16

Other devices under the same FDA submission

1 device cleared under K120613, K122234 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.