UniCel DxH 800 Coulter Cellular Analysis System, Catalog… recallZ-1013-2017 · Beckman Coulter Inc. · initiated Dec 19, 2016
Class II Terminated All recalls posted in January 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. This is due to limitations in the available technology as well as sample limitations.
FDA root cause category: Under Investigation by firm
What the company is doing
Beckman Coulter sent an Urgent Medical Device Recall letter dated December 19, 2016, to all affected customers to inform them that Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood samples. The letter informs the customers of the impact, actions to be taken, and resolution. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Customer Support Center: http://www.beckmancoulter.com/customersupport/support, 800-526-7694 in the United States and Canada, and outside the United States and Canada, to contact their local Beckman Coulter Representative.
Recall details
- Recall number
- Z-1013-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 19, 2016
- Posted by the FDA
- Jan 11, 2017
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 4,657 units total (2,067 units in US)
- Product code
- GKZ
- FDA submissions
- K081930, K120771, K140911
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the countries of : Australia, Belarus, Belgium, Bosnia and Herzegovina, Brunei Darussalam, Canada, China, Columbia, Croatia, Czech Republic, France, Germany, Gibraltar, Ecuador Hong Kong, India, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Macao, Malaysia, Mayotte, Mexico, Morocco, Myanmar, New Zealand, Oman, Panama, Philippines, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Republic, United Kingdom, Uruguay and Vietnam.
Product description
UniCel DxH 800 Coulter Cellular Analysis System, Catalog No. 629029, B24465, B24802, B68304
Code information (models, lots, serial numbers)
Serial Numbers: All
Other devices under the same FDA submission
1 device cleared under K081930, K120771, K140911 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DxH 900 Hematology System C11478Beckman Coulter, Inc. · Counter, Differential Cell1 recall17 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 76044).
- Z-1014-2017 · TerminatedDxH 600 Coulter Cellular Analysis System, Catalog No. B23858
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page