SCREW 02.7-12MM LOCKING, REF W27 ST112: a) DRP, b) NEO… recallZ-0952-2017 · In2Bones, SAS · initiated Dec 19, 2016
Class II Terminated All recalls posted in January 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from In2Bones, SAS or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Dimensional non-conformity of the thread on the screw head.
FDA root cause category: Component design/selection
What the company is doing
IN2BONES en a Field Safety Corrective Action letter dated December 15, 2016, to the sole US distributor. The letter identified the product, the problem, and the action to be taken by the customer. We therefore recommend you to follow the instructions here below: 1. Identify all the devices with the batches subject of this recall that might still be in your inventory and quarantine them 2. Inform and distribute this Recall Notification to all relevant persons within your organization 3. To distributors only: Identify all the devices with the batches subject of this recall that were delivered to your customers, and if relevant, instruct them to also follow these instructions (identification and quarantine) 4. Fill in and return the fax back form enclosed. With this form, you will certify that you have received this Recall Notification and intend to comply with the recommendations listed. This fax back form will enable In2Bones to conduct effectiveness checks. Customers with questions were instructed to contact the Quality and Regulatory Affairs team at +33 4 72 29 26 26 / +33 7 61 88 44 30 or by email: [email protected]. For questions regarding this recall call 844-602-6637.
Recall details
- Recall number
- Z-0952-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 19, 2016
- Posted by the FDA
- Jan 6, 2017
- Recalling firm
- In2Bones, SAS
- Quantity
- 19 units
- Product code
- HRS
- FDA submissions
- K160995
- Distribution
- Worldwide Distribution - US TN only and Internationally to France and Switzerland
Product description
SCREW 02.7-12MM LOCKING, REF W27 ST112: a) DRP, b) NEO, In2Bones, 28 chemin du petit bois, 69130 Ecully - France The NEOVIEW Plating System is intended for fixation of intra-articular and extraarticular fractures of the distal radius and reconstruction of the distal radius
Code information (models, lots, serial numbers)
Lot Numbers: a) 1602064, b) 1606067
Other devices under the same FDA submission
1 device cleared under K160995 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NEOVIEW® W27 ST014In2bones · Plate, Fixation, Bone2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 76104).
- Z-0953-2017 · TerminatedNEO SCREW 02.7-14MM LOCKING, REF W27 ST114, In2Bones, 28 chemin du petit bois, 69130 Ecully - France The NEOVIEW Plating System is intended for fixation of intra-articular and ext…
- Z-0954-2017 · TerminatedNEOVIEW(R) SCREWS SET, REF W27 ST900, STERILE R The NEOVIEW Plating System is intended for fixation of intra-articular and extraarticular fractures of the distal radius and recons…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page