AFX Endovascular AAA System, Endoleak Type IIIB recallZ-1038-2017 · Endologix · initiated Dec 27, 2016
Class II Terminated All recalls posted in January 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Endologix or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
FDA root cause category: Other
What the company is doing
A Dear Physician letter dated 6/2/15 was sent to customers to inform them that the 2014 Clinical Update on the Endologix Endovascular Systems for AAA Repair. The letter provides update reports for the cumulative clinical data and information on the Powerlink System and the AFX Endovascular AAA System (AFX System). On January 2016, a Dear Physician letter was sent to customers to inform them of the 2015 Clinical Update covering the Endologix Endovascular Systems for AAA Repair. The letter provided update reports on the cumulative clinical data and commercial experience (i.e., post-market surveillance) of the Powerlink System and the AFX Endovascular AAA system and contains important information and recommendations regarding the continued safe and effective use of these devices. A Dear Physician letter dated 12/27/16 will be sent to inform customers of the important information related to the AFX Endovascular AAA System (AFX System). The letter provides a series of updates to the Instructions for Use (IFU) and modifications to the product were implemented, including introduction of a graft material processing improvement known as Duraply. Customers with any questions are instructed to contact their Endologix representative.
Recall details
- Recall number
- Z-1038-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 27, 2016
- Posted by the FDA
- Jan 18, 2017
- Recalling firm
- Endologix
- Quantity
- 40967
- Product code
- MIH
- FDA submissions
- P040002
- Distribution
- US and worldwide: Austria, Belgium, Czech Republic, France, Germany, Greece, Ireland, Israel, Italy, Latvia, Luxembourg, Monaco, Netherlands, Poland, Portugal, Spain, Sweden, Switzerland, Turkey, United Kingdom.
Product description
AFX Endovascular AAA System, Endoleak Type IIIB; The AFX Endovascular AAA System is intended for endovascular treatment of patients with abdominal aortic or aortoiliac aneurysms (AAA).
Code information (models, lots, serial numbers)
The correction is not lot specific. It applies to AFX procedures conducted after July 2015
Other devices under the same FDA submission
97 devices cleared under P040002 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- AFX2 Bifurcated Endograft System BEA28-100/I20-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths107 reports
- AFX2 Bifurcated Endograft System BEA28-90/I20-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths67 reports
- Vela Proximal Endograft System A28-28/C95-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment2 deaths66 reports
- AFX2 Bifurcated Endograft System BEA25-100/I16-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths62 reports
- AFX2 Bifurcated Endograft System BEA25-90/I16-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death53 reports
- AFX Endovascular AAA System BA28-90/I20-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death50 reports
- AFX2 Bifurcated Endograft System BEA28-100/I16-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths50 reports
- AFX2 Bifurcated Endograft System BEA25-100/I20-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths50 reports
- AFX2 Bifurcated Endograft System BEA22-90/I16-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths48 reports
- AFX2 Bifurcated Endograft System BEA28-90/I16-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment7 deaths47 reports
- AFX2 Bifurcated Endograft System BEA28-120/I20-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths45 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page