Skip to content

ETEST Gentamicin GM256 FOAM packaging, Product Name: ETEST… recallZ-1530-2017 · BioMerieux SA · initiated Dec 15, 2016

Class II Terminated All recalls posted in March 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from BioMerieux SA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.

FDA root cause category: Storage

What the company is doing

biomerieux sent an Urgent Field Safety Notice dated December 15, 2016, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to identify the affected lots and immediately order the replacement products. Laboratories should continue to follow their current QC procedures. When replacement product is received, consignees should discontinue using any discard affected product. Consignees should distribute this information to all appropriate personnel in their laboratory, retain a copy in their files, and forward the information to all parties that may use this product, including others to whom they may have transferred our product. Consignees should contact their bioMerieux representative for product compensation. Consignees should also complete and return the Acknowledgment Form in Attachment A by Fax to confirm receipt of the notice. Consignees with questions should contact their local bioMerieux Clinical Customer Service representative. For questions regarding this recall call 314-731-8516.

Recall details

Recall number
Z-1530-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 15, 2016
Posted by the FDA
Mar 21, 2017
Recalling firm
BioMerieux SA
Quantity
unknown
Product code
JWY
FDA submissions
K913459
Distribution
Worldwide Distribution - U.S.( Nationwide); Foreign: Utd.Arab Emir., Argentina, Austria, Australia, Brazil, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Algeria, Spain, France, United Kingdom, Greece, Hong Kong, India, Italy, Japan, South Korea, Mexico, Netherlands, Poland, Portugal, Russian, Fed. Sweden, Singapore, Thailand, Turkey, South Africa, Indonesia, Peru, Saudi Arabia, Gabon, Kazakhstan, Philippines, Ivory Coast, Venezuela, Kenya, Serbia, Tunisia, Macedonia, Lebanon, Romania, Benin, Bosnia-Herz., Oman, Uruguay, Lithuania, Burkina Faso, Slovenia, Qatar, Kuwait, Costa Rica, Paraguay, Dominican Republic, Croatia, Panama, Congo, Dem. Rep. Congo, Honduras, Estonia, Latvia, Cambodia, Bulgaria, Vietnam, Dutch Antilles, Nicaragua, Israel, Georgia, Palestine, Egypt, Pakistan (includ.Afghanistan), Malta, Jordan Bahrain, El Salvador, Nigeria, Dutch Antilles, Iran, Belarus, Zimbabwe, Morocco, Mauretania, Anguilla, Philippines, Malaysia, Turkmenistan, Guam, Brunei, Bolivia, Guatemala, Ecuador, Belarus, Mongolia

Product description

ETEST Gentamicin GM256 FOAM packaging, Product Name: ETEST Gentamicin GM 0.016-256(low). ETEST is a quantitative technique for determining the antimicrobial susceptibility of Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in g/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. Current Antimicrobial Susceptibility Testing (AST) methods are based either on quantitative dilution techniques or qualitative diffusion procedures. Dilution methods are based on two-fold serial dilutions of antibiotics in broth or agar media. These methods generate the MIC value i.e. Minimum Inhibitory Concentration of a given antibiotic in g/mL that will inhibit the growth of a particular bacterium under defined experimental conditions

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Reference no. 512550, Lot # 1002001100 & 1002048740; Reference no. 512510, Lot #1001970170 & 1002154750; Reference no. 512558, Lot #1001564260, 1001676270, 1001733240, 1001793760, 1001873760, 1002113320, 1002230930, 1002559800, 1002823250, 1002113320, 1002230930, 1002559800, 1002823250, 1003495040, 1003883840, 1004248160, 1004252230, 1004277880, 1004567040, 1004946070, 1004980260, 1005146190, 1003495040, 1003883840, 1004248160, 1004252230, 1004277880, 1004567040, 1004946070, 1004980260 & 1005146190 and Reference no. 512518, Lot #1001595660, 1001707960, 1001803880, 1002094250, 1002229210, 1002233690, 1002559790, 1002684500, 1002823910, 1003164040, 1003191670, 1003279330, 1003604650, 1004047450, 1004157400, 1004247950, 1004252220, 1004528390 & 1004567000.

Other devices under the same FDA submission

2 devices cleared under K913459 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 76201).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page