Medical device recalls: March 2017
The FDA posted 138 medical device recall records in March 2017, from 65 recall events by 49 companies. 4 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Merge Healthcare, Inc. · 7 events, 7 records
- Zimmer Biomet, Inc. · 4 events, 44 records
- Siemens Medical Solutions USA, Inc · 4 events, 5 records
- BioMerieux SA · 3 events, 14 records
- Beckman Coulter Inc. · 2 events, 4 records
- Siemens Healthcare Diagnostics, Inc. · 2 events, 2 records
- Centurion Medical Products Corporation · 1 event, 7 records
- Covidien LLC · 1 event, 7 records
FDA root cause
- Packaging · 35
- Under Investigation by firm · 12
- Storage · 12
- Software design · 12
- Nonconforming Material/Component · 11
- Packaging process control · 9
- Device Design · 8
- Process design · 7
Class I recalls (4)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-1898-2017 · ReFlow Medical · posted Mar 29, 2017 · Terminated
ReFlow Medical is recalling the Wingman 35 Crossing Catheter because it may be prone to tip detachment.
- Z-1417-2017 · Philips North America, LLC · posted Mar 24, 2017 · Terminated
HeartStart MRx Monitor/Defibrillator may may exhibit abnormal behavior when operating using a single battery as the sole source of power may fail to deliver therapy or monitoring
- Z-1497-2017 · Merit Medical Systems, Inc. · posted Mar 24, 2017 · Terminated
Merit Medical Systems, Inc. announces a voluntary field action for sheaths included with the 7F Prelude Short Sheath Introducers because the sheath marker tips may detach during use.
- Z-1257-2017 · Physio-Control, Inc. · posted Mar 3, 2017 · Terminated
The firm has received complaints that the LIFEPAK 1000 Defibrillator is unexpectedly powering off during device usage.
Class II recalls (132)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-1767-2017 · Smiths Medical ASD Inc. · posted Mar 30, 2017 · Terminated
Smiths Medical has become aware that the drawing on the Instructions for Use (IFU) on certain yellow striped CADD Administration Sets (part number 21-7339-24), incorrectly represent the actual filter size of the product. Although the text description on…
- Z-1824-2017 · Siemens Medical Solutions USA, Inc · posted Mar 30, 2017 · Terminated
Software error - As a result of a software fault in Artis zee systems with software version VD11 and an A100 generator, the possibility exists that following the failure of a tube assembly focus, the Artis system functions properly until the next…
- Z-1698-2017 · AGFA Healthcare Corp. · posted Mar 29, 2017 · Terminated
The DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored.
- Z-1853-2017 · Angiodynamics Inc. (Navilyst Medical Inc.) · posted Mar 29, 2017 · Terminated
The product has the potential to be non-sterile as it was not packaged in accordance to specification.
- Z-1455-2017 · Strata Skin Sciences, Inc. · posted Mar 27, 2017 · Terminated
When the Liquid Guide ( LLG) is not used according to the operator manual, the switch control wires may become loose and shorted. Shorted wires may lead to unintentional laser firing.
- Z-1685-2017 · Focus Diagnostics Inc · posted Mar 27, 2017 · Terminated
After a submission for CLIA database update from Focus Diagnostics to DiaSorin Molecular,the company found that their Legionella IFA device (IF0950) is not 510(k) exempt.
- Z-1686-2017 · Focus Diagnostics Inc · posted Mar 27, 2017 · Terminated
After a submission for CLIA database update from Focus Diagnostics to DiaSorin Molecular,the company found that their Legionella IFA device (IF0950) is not 510(k) exempt.
- Z-1688-2017 · Siemens Medical Solutions USA, Inc · posted Mar 27, 2017 · Terminated
Software changes now available to address several issues
- Z-1689-2017 · Siemens Medical Solutions USA, Inc · posted Mar 27, 2017 · Terminated
Software changes now available to address several issues
- POST-OP EYE TRAY ↗Class IIZ-1730-2017 · Centurion Medical Products Corporation · posted Mar 27, 2017 · Terminated
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- OPHTHALMIC DRESSING TRAY ↗Class IIZ-1731-2017 · Centurion Medical Products Corporation · posted Mar 27, 2017 · Terminated
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- OPHTHALMIC DRS TRAY ↗Class IIZ-1732-2017 · Centurion Medical Products Corporation · posted Mar 27, 2017 · Terminated
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- EYE KIT ↗Class IIZ-1733-2017 · Centurion Medical Products Corporation · posted Mar 27, 2017 · Terminated
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- Z-1734-2017 · Centurion Medical Products Corporation · posted Mar 27, 2017 · Terminated
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- OPHTHALMIC DRESSING TRAY ↗Class IIZ-1735-2017 · Centurion Medical Products Corporation · posted Mar 27, 2017 · Terminated
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- CUSTOM CADAVER KIT ↗Class IIZ-1736-2017 · Centurion Medical Products Corporation · posted Mar 27, 2017 · Terminated
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- Covidien Curity Eye Pad ↗Class IIZ-1739-2017 · Covidien LLC · posted Mar 27, 2017 · Terminated
Product sterility is compromised due to breach of sterile barrier
- Z-1740-2017 · Covidien LLC · posted Mar 27, 2017 · Terminated
Product sterility is compromised due to breach of sterile barrier
- Covidien Curity Eye Pad ↗Class IIZ-1741-2017 · Covidien LLC · posted Mar 27, 2017 · Terminated
Product sterility is compromised due to breach of sterile barrier
- Z-1742-2017 · Covidien LLC · posted Mar 27, 2017 · Terminated
Product sterility is compromised due to breach of sterile barrier
- Z-1743-2017 · Covidien LLC · posted Mar 27, 2017 · Terminated
Product sterility is compromised due to breach of sterile barrier
- Z-1744-2017 · Covidien LLC · posted Mar 27, 2017 · Terminated
Product sterility is compromised due to breach of sterile barrier
- Z-1745-2017 · Covidien LLC · posted Mar 27, 2017 · Terminated
Product sterility is compromised due to breach of sterile barrier
- MEDTECH ROSA Brain 3.0Class IIZ-1754-2017 · Zimmer Biomet, Inc. · posted Mar 27, 2017 · Terminated
Unapproved change made by the supplier.
- MEDTECH ROSA Spine 1.0.2 ↗Class IIZ-1755-2017 · Zimmer Biomet, Inc. · posted Mar 27, 2017 · Terminated
Unapproved change made by the supplier.
- various polyethylene implantsClass IIZ-1858-2017 · Zimmer Biomet, Inc. · posted Mar 27, 2017 · Terminated
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
- various polyethylene implantsClass IIZ-1859-2017 · Zimmer Biomet, Inc. · posted Mar 27, 2017 · Terminated
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
- various polyethylene implantsClass IIZ-1860-2017 · Zimmer Biomet, Inc. · posted Mar 27, 2017 · Terminated
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
- various polyethylene implantsClass IIZ-1861-2017 · Zimmer Biomet, Inc. · posted Mar 27, 2017 · Terminated
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
- various polyethylene implants ↗Class IIZ-1862-2017 · Zimmer Biomet, Inc. · posted Mar 27, 2017 · Terminated
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
- various polyethylene implants ↗Class IIZ-1863-2017 · Zimmer Biomet, Inc. · posted Mar 27, 2017 · Terminated
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
- Z-1679-2017 · Molded Products Inc · posted Mar 24, 2017 · Terminated
A customer relayed a complaint of a leaking dialysis tubing connector.
- Hitachi Oasis MRI System ↗Class IIZ-1540-2017 · Hitachi Medical Systems America Inc · posted Mar 23, 2017 · Terminated
Image data transferred from the MRI system to a workstation showed errors on the slice position reference image.
- Z-1541-2017 · Hitachi Medical Systems America Inc · posted Mar 23, 2017 · Terminated
Image data transferred from the MRI system to a workstation showed errors on the slice position reference image.
- Hitachi Echelon MRI System ↗Class IIZ-1542-2017 · Hitachi Medical Systems America Inc · posted Mar 23, 2017 · Terminated
Image data transferred from the MRI system to a workstation showed errors on the slice position reference image.
- Proteus 235 ↗Class IIZ-1695-2017 · Ion Beam Applications S.A. · posted Mar 23, 2017 · Terminated
The Proton Therapy System can allow gantry rotation even if the snout is not properly locked.
- Z-1763-2017 · Roche Diagnostics Corporation · posted Mar 23, 2017 · Terminated
"Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (URF) modules with software version 02-xx, only when rack buffering is activated and…
- Z-1764-2017 · Roche Diagnostics Corporation · posted Mar 23, 2017 · Terminated
Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (BRF) modules with software version 02-xx, only when rack buffering is activated and…
- Z-1769-2017 · Datascope Corporation · posted Mar 23, 2017 · Terminated
A complaint was received on January 1, 2017 stating the pigtail on line 2 of the oxygenator was assembled to the port where the yellow cap and the clear cap remained on the center port.
- Z-1781-2017 · Boston Scientific Corporation · posted Mar 23, 2017 · Terminated
One lot of Imager II Angiographic Catheters is being recalled due to packaging mislabeling. The product may be labeled as Imager II Contra 2 curve catheters, when in fact the package contains Contra curve catheters.
- LipiFlow Thermal Pulsation System/LipiFlow, System Activator (Disposable), Rx Only, Sterile ↗Class IIZ-1535-2017 · Tearscience, Inc · posted Mar 22, 2017 · Terminated
Labeling: The labeled storage temperature is not consistent with the temperature conditions tested for the entire four year shelf life duration.
- Z-1518-2017 · Revision Optics Inc · posted Mar 21, 2017 · Terminated
ReVision Optics has updated the instructions for use (IFU) for the Raindrop Near Vision Inlay to emphasize that only Balanced Salt Solution (BSS) may be used to irrigate under the flap and that topical medications or lubricants should not be administered…
- Z-1521-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1522-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1523-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1524-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1525-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1526-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1527-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1528-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1529-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1530-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1531-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1532-2017 · BioMerieux SA · posted Mar 21, 2017 · Terminated
Product Stability issues: The current shelf-life claims of the ETEST products are not supported by internal testing.
- Z-1487-2017 · Crossroads Extremity Systems Llc · posted Mar 20, 2017 · Terminated
Seven 18mm x 18mm x 18 mm implants had been placed in kits, packed and labeled as 18mm x 14mm x 14mm.
- Z-1496-2017 · Merge Healthcare, Inc. · posted Mar 20, 2017 · Terminated
For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and medical record number in…
- Z-1683-2017 · National Biological Corp · posted Mar 20, 2017 · Terminated
The wiring may allow the lamps to turn on with the key switch rather than the timer.
- Z-1768-2017 · Nihon Kohden America Inc · posted Mar 20, 2017 · Terminated
The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.
- Z-1801-2017 · BioMerieux SA · posted Mar 20, 2017 · Terminated
Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.
- Z-1814-2017 · Aesculap Implant Systems LLC · posted Mar 20, 2017 · Terminated
Aesculap Implant Systems LLC is recalling the Minop Trocar due to the possibility it may have sharp edges on the distal end which may lead to the abrasion of the insulation when removing the electrode.
- Z-2549-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2550-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2551-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2552-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2553-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2554-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2555-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2556-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2557-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2558-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2559-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2560-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2561-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2562-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2563-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2564-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2565-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2566-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2567-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2568-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2569-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2570-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2571-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2572-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2573-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2574-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2575-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2576-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2577-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2578-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2579-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2580-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2581-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2582-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2583-2017 · Zimmer Biomet, Inc. · posted Mar 20, 2017 · Terminated
Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-1534-2017 · Siemens Medical Solutions USA, Inc · posted Mar 19, 2017 · Terminated
In Artis systems with A100G generators, a component can be affected by aging which could result in the potential failure of a module in the high-voltage generator.
- Z-1703-2017 · Siemens Healthcare Diagnostics, Inc. · posted Mar 19, 2017 · Terminated
Siemens Healthcare Diagnostics has confirmed that in isolated cases when ECREA is processed immediately after the weekly automated Acid Clean routine during Probe Test, there is the remote potential for an elevation of greater than 15 percent in the ECREA…
- Z-1713-2017 · BioMerieux SA · posted Mar 19, 2017 · Terminated
Problem with colored eluates for whole blood extractions
- Z-1479-2017 · Philips Medical Systems (Cleveland) Inc · posted Mar 16, 2017 · Terminated
Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan
- Z-1480-2017 · Philips Medical Systems (Cleveland) Inc · posted Mar 16, 2017 · Terminated
Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan
- Z-1481-2017 · Philips Medical Systems (Cleveland) Inc · posted Mar 16, 2017 · Terminated
Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan
- Z-1483-2017 · Acumed LLC · posted Mar 16, 2017 · Terminated
Guide wires sized as .045 X 6 were mis-packaged as .054 X 7 guide wires and those sized as .054 X 7 were mis- packaged as .045 X 6 guide wires. There is a possibility that using the incorrect guide wire size can cause health hazards, such as mal or…
- Z-1477-2017 · C2 Therapeutics, Inc. · posted Mar 15, 2017 · Terminated
The Controller Cap of the C2 CyroBalloon Ablation System may crack as it is tightened onto the controller. That could compromise the integrity and result in damage to the Controller Cap.
- Z-1543-2017 · Medtronic Vascular, Inc. · posted Mar 15, 2017 · Terminated
Medtronic is initiating a voluntary Urgent Medical Device Recall for a subset of Endurant/ Endurant II Bifurcated Stent Graft Systems of specific models and serial numbers as this specific subset of stent grafts has greater susceptibility to fabric…
- Z-1709-2017 · Beckman Coulter Inc. · posted Mar 15, 2017 · Terminated
Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the…
- Z-1710-2017 · Beckman Coulter Inc. · posted Mar 15, 2017 · Terminated
Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the…
- Z-1711-2017 · Beckman Coulter Inc. · posted Mar 15, 2017 · Terminated
Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the…
- Z-1458-2017 · Terumo Cardiovascular Systems Corporation · posted Mar 14, 2017 · Terminated
Terumo CVS initiated a voluntary recall for the Level Sensor II Pads and Level Sensor Gel Pads due to non-compliant labeling because the product expiration date is displayed in a format that may not be recognizable to all users.
- Z-1459-2017 · Terumo Cardiovascular Systems Corporation · posted Mar 14, 2017 · Terminated
Terumo CVS initiated a voluntary recall for the Level Sensor II Pads and Level Sensor Gel Pads due to non-compliant labeling because the product expiration date is displayed in a format that may not be recognizable to all users.
- Z-1454-2017 · Corentec CO., Ltd · posted Mar 13, 2017 · Terminated
Lot may contain the wrong instruction (Instruction for Chinese customer) instead of correct US instructions for use.
- Z-1495-2017 · Zimmer Biomet, Inc. · posted Mar 13, 2017 · Terminated
Digital templates were created with the incorrect files.
- Z-1681-2017 · Merge Healthcare, Inc. · posted Mar 13, 2017 · Terminated
The system will completely stop allowing the downloading or acquisition of new images due to incompatible software application versions.
- Specimen Gate Laboratory ↗Class IIZ-1761-2017 · PerkinElmer Life and Analytical Sciences, Wallac, OY · posted Mar 13, 2017 · Terminated
Potential for errors in patient results generated by the Specimen Gate Laboratory Software.
- Z-1470-2017 · Merge Healthcare, Inc. · posted Mar 10, 2017 · Terminated
The software produced a number of "do not route" exceptions, which may result in potential patient injury or delay in diagnosis or treatment.
- Z-1533-2017 · Siemens Medical Solutions USA, Inc · posted Mar 10, 2017 · Terminated
Software updates
- Z-1699-2017 · Stryker Communications · posted Mar 10, 2017 · Terminated
The potential hazard that can arise for the combination of the missing screws and the overloading by the end user is insufficient mount force that causes the equipment to fall.
- Z-1426-2017 · Merge Healthcare, Inc. · posted Mar 9, 2017 · Terminated
Reports were missing demographic header information when they were faxed to referring physicians.
- CELL-DYN Ruby System: ↗Class IIZ-1438-2017 · Abbott Laboratories · posted Mar 9, 2017 · Terminated
The Pump Relay Printed Circuit Board Assembly (PCBA) used in the CELL-DYN Ruby instrument may prematurely fail and lead to instrument stoppage
- Merge Cardio software using EchoIMS. The firm name on the label is Merge Healthcare, Hartland, WI.Class IIZ-1415-2017 · Merge Healthcare, Inc. · posted Mar 8, 2017 · Terminated
A situation can occur allowing two physicians to access the same study report in EchoIMS when launched from the Cardio Study List without receiving the read-only notification prompt.
- Z-1512-2017 · Devicor Medical Products Inc · posted Mar 8, 2017 · Terminated
The device did not have Revolve Stereo Probe labeling, no product code indicating the size and length of the probe, no sterilization sticker or lot number.
- Z-1536-2017 · Medical Components, Inc dba MedComp · posted Mar 8, 2017 · Terminated
The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit contains a 10F valved peelable introducer. The kit is packaged with a 9F valved peelable introducer. The port lumen will not fit through the introducer…
- Z-1684-2017 · Streck · posted Mar 8, 2017 · Open, Classified
Complaints of the 2.0ml tube stopper coming loose during phlebotomy when small diameter tube adapters were not utilized. Such adapters aid with centering small diameter tubes to properly pierce the middle of the stoppers.
- Z-1705-2017 · Arrow International Inc · posted Mar 8, 2017 · Terminated
Arrow International, Inc. notified its customers who received the affected Antimicrobial/Antithrombogenic PICC Kits, that the kits were being recalled due to a labeling error. The lidstock on some of the kits reads CDC-45541-HPK1A, while the banner card on…
- Z-1485-2017 · Purewick Corporation · posted Mar 6, 2017 · Terminated
PureWick is recalling external catheter for women Wicks because the labeling did not contain the phrase "This product contains dry natural rubber" and/or erroneously included the phrase "Latex-free." The white gauze component of the PureWick contains dry…
- Z-1399-2017 · Parker Medical · posted Mar 3, 2017 · Terminated
The product was not functioning as intended. The ventilation to the patient was not disrupted, but the single lung isolation and ventilation did not function as intended.
- Z-1402-2017 · Metrex Research, LLC. · posted Mar 3, 2017 · Terminated
Metrex is recalling the CaviWipes Extra Large because they may have been contaminated during the packaging process.
- Z-1403-2017 · Merge Healthcare, Inc. · posted Mar 3, 2017 · Terminated
When taking measurements from images on the Cardio workstation or from the US cart, numbers are not crossing to the report in the correct unit of measure.
- MicroAire K-WiresClass IIZ-1407-2017 · MicroAire Surgical Instruments, LLC · posted Mar 3, 2017 · Terminated
Mislabeling on three lots of K-Wires. The incorrect product description " dual trocar" instead of "single trocar" was on the product label on the following parts: 1600-9355NS (lot 61005), 1600-9625NS (lot 60297), and 1604-162NS (lot 62856).
- Z-1385-2017 · Merge Healthcare, Inc. · posted Mar 2, 2017 · Terminated
The software is not adding the correct tomo image.
- Z-1395-2017 · MED-EL Elektromedizinische Gereate, Gmbh · posted Mar 2, 2017 · Terminated
Unit Defect: Firm inspection revealed units did not have the compartment lock firmly welded to the housing.
- Z-1384-2017 · Beckman Coulter Inc. · posted Mar 1, 2017 · Terminated
The probe misalignment or bending has the potential to lead to delay in reporting results or incorrect patient results.
- Z-1692-2017 · United Orthopedic Corporation · posted Mar 1, 2017 · Terminated
The locking lever pin component of the femoral driver may become deformed and break, resulting in malfunction during surgery and causing the device to become inoperable.
Class III recalls (2)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-1712-2017 · Siemens Healthcare Diagnostics, Inc. · posted Mar 15, 2017 · Terminated
Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry…
- Roth Net Polyp Retrieval ↗Class IIIZ-1416-2017 · US Endoscopy Group Inc · posted Mar 8, 2017 · Terminated
The firm was notified that the expiration date on the outer carton is different than the expiration date on the inner package. The expiration date is shorter than the actual expiration.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.