AFX Endovascular AAA System recallZ-1048-2017 · Endologix · initiated Jan 4, 2017
Class II Terminated All recalls posted in January 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Endologix or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Endologix updated information on the rates of Type III endoleaks and suggestions for patient surveillance and treatment.
FDA root cause category: Device Design
What the company is doing
Endologix sent a Dear Physician letter on 12/30/16 to inform them information related to the AFX Endovascular AAA System (AFX System). The letter provides updated information on the rates of Type III endoleaks and suggestions for patient surveillance and treatment. Customers were informed of the voluntary recall of the small remaining quantity of original AFX with Strata graft material, and the larger diameter sizes of AFX2. Customers with any questions are instructed to contact their Endologix representative. Customers with questions were instructed to call their Endologix representative. On 1/17/17, Endologix issued a press release to inform consumers that firm is resuming shipments of all sizes of AFX2 Endovascular AAA systems and appointed Laura Nagel as Vice President, Global Quality. Consumers are informed that the large diameter sizes of AFX2 were placed on temporary hold on 12/27/16 to investigate a manufacture issue.
Recall details
- Recall number
- Z-1048-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 4, 2017
- Posted by the FDA
- Jan 19, 2017
- Recalling firm
- Endologix
- Quantity
- 34 units
- Product code
- MIH
- FDA submissions
- P040002
- Distribution
- Worldwide Distribution - US (nationwide)and Internationally to Austria, Belgium, France, Germany, Greece, Italy, Ireland, Monaco, Poland Portugal, Romania, Spain, Sweden, Turkey, Israel, Argentina, Florida, Mexico, Chile, Brazil, Japan, Singapore, Malaysia, and Thailand.
Product description
AFX Endovascular AAA System Item Number Item Description F0070305 A2525/C75 V F0070308 A2828/C75 V F0070312 A3434/C80 V f00553 I1616/C55F SA F00556 I2013/C70F SA F00555 I2020/C55F SA F00558 IS2025/C55 SA F00559 IS2025/C65 SA F0072605 A2525/C75O20 V F0072606 A2525/C95O20 V f00552 I1616/C88 SA F00557 I2013/C88F SA The AFX Endovascular AAA System is intended for endovascular treatment of patients with abdominal aortic or aortoiliac aneurysms (AAA
Code information (models, lots, serial numbers)
Lot # Lot/Serial 1164318 1164318019 1164390 1164390020 1164389 1164389009 1164388 1164388021 1262039 1262039002 1246596 1246596024 1164477 1164477026 1164470 1164470015 1055838 1055838026 1055838 1055838022 1250434 1250434002 1249743 1249743025 1257058 1257058017 1055877 1055877010 1249363 1249363009 1249358 1249358025 1267536 1267536008 1047170 1047170016 1267536 1267536025 1257553 1257553020 1214252 1214252011 1229346 1229346007 1274322 1274322003 1257977 1257977011 1250424 1250424016 1250424 1250424008 1250424 1250424007 1250399 1250399024 1047135 1047135007 1275167 1275167014 1275166 1275166031 1274322 1274322023 1250399 1250399011 1056308 1056308054
Other devices under the same FDA submission
97 devices cleared under P040002 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- AFX2 Bifurcated Endograft System BEA28-100/I20-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths107 reports
- AFX2 Bifurcated Endograft System BEA28-90/I20-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths67 reports
- Vela Proximal Endograft System A28-28/C95-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment2 deaths66 reports
- AFX2 Bifurcated Endograft System BEA25-100/I16-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths62 reports
- AFX2 Bifurcated Endograft System BEA25-90/I16-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death53 reports
- AFX Endovascular AAA System BA28-90/I20-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment1 death50 reports
- AFX2 Bifurcated Endograft System BEA28-100/I16-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths50 reports
- AFX2 Bifurcated Endograft System BEA25-100/I20-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths50 reports
- AFX2 Bifurcated Endograft System BEA22-90/I16-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment4 deaths48 reports
- AFX2 Bifurcated Endograft System BEA28-90/I16-30Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment7 deaths47 reports
- AFX2 Bifurcated Endograft System BEA28-120/I20-40Endologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths45 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 76256).
- Z-1049-2017 · TerminatedAFX2 Endovascular AAA System Item No. Item Description F00820-01 BEA28-120/I20-40 F00820-02 BEA28-100/I20-40 F00820-03 BEA28-80/I20-40 F00820-06 BEA28-120/I16-40 F00820-07 B…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page