Skip to content

Raindrop Near Vision Inlay, Model# RD1-1 recallZ-1518-2017 · Revision Optics Inc · initiated Feb 7, 2017

Class II Terminated All recalls posted in March 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Revision Optics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

ReVision Optics has updated the instructions for use (IFU) for the Raindrop Near Vision Inlay to emphasize that only Balanced Salt Solution (BSS) may be used to irrigate under the flap and that topical medications or lubricants should not be administered until centration of the inlay and proper flap positioning has been confirmed at the slit lamp.

FDA root cause category: Use error

What the company is doing

A Dear Doctor letter was sent to inform customers that ReVision Optics wanted to share a surgical update regarding the Raindrop Near Vision Inlay procedure. Customers are informed that no ophthalmic medications or lubricants should be used immediately prior to or during placement of the Raindrop Near Vision Inlay. Only Balanced Salt Solution (BSS) should be used to irrigate under the flap during the Raindrop procedure. Customers are informed that ophthalmic medications or lubricants should be administered only after verifying proper inlay centration and flap positioning as observed at the slit lamp. This ensures that no further inlay or flap manipulation is required. ReVision Optics has updated the instructions for use (IFU) for the Raindrop Near Vision Inlay to emphasize that only BSS may be used to irrigate under the flap and that topical medications or lubricants should not be administered until centration of the inlay and proper flap positioning has been confirmed at the slit lamp. Customers with questions are instructed to contact Luis Vargas regarding the surgical procedure update.

Recall details

Recall number
Z-1518-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 7, 2017
Posted by the FDA
Mar 21, 2017
Recalling firm
Revision Optics Inc
Quantity
1,279 devices
Product code
LQE
FDA submissions
P150034
Distribution
US: AL, AZ, CA, FL, GA, HI, IL, IN, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, SD, TN, TX, UT, WA

Product description

Raindrop Near Vision Inlay, Model# RD1-1

Code information (models, lots, serial numbers)

multiple lots since 08/01/2016

Other devices under the same FDA submission

1 device cleared under P150034 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page