Skip to content

SONNET Mini Battery Pack CableProduct Usage recallZ-1395-2017 · MED-EL Elektromedizinische Gereate, Gmbh · initiated Oct 25, 2016

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MED-EL Elektromedizinische Gereate, Gmbh or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Unit Defect: Firm inspection revealed units did not have the compartment lock firmly welded to the housing.

FDA root cause category: Device Design

What the company is doing

MED-EL is notifying all consignees to whom affected products have been directly shipped by Headquarters (based on first delivery shipment records) on 10/25/2016.

Recall details

Recall number
Z-1395-2017
Status
Terminated
Initiated by the company
Oct 25, 2016
Posted by the FDA
Mar 2, 2017
Recalling firm
MED-EL Elektromedizinische Gereate, Gmbh
Quantity
538 units
Product code
ISM
FDA submissions
P000025S087
Distribution
Worldwide Distribution - US Nationwide in the states of CA, NC and TX; and the countries of Belgium, Belarus, Canada, France, Germany, Hong Kong, Israel, Netherlands, Philippines, Poland, Portugal, Turkey, United Kingdom, and Uruguay.

Product description

SONNET Mini Battery Pack CableProduct Usage: The SONNET Mini Battery Pack Cable is part of the SONNET Mini Battery Pack, which is an external battery pack used to power one MED-EL SONNET control unit. It can be worn on the upper body or the hip using different SONNET Mini Battery Pack Cable lengths to connect the SONNET Mini Battery Pack to one processor unit.

Code information (models, lots, serial numbers)

Part No. 32291, Kit 33372, Kit UDI - Primary ID: 09008737333720 (anthracite 50); Part No. 32292, Kit 33374, Kit UDI - Primary ID: 09008737333744 (ebony 50); Part No. 32293, Kit 33375, Kit UDI - Primary ID: 09008737333751 (black 50); Part No. 32294, Kit 33376, Kit UDI - Primary ID: 09008737333768 (white 50); Part No. 32295, Kit 33377, Kit UDI - Primary ID: 09008737333775 (beige 50) and Part No. 32296, Kit 33378, Kit UDI - Primary ID: 09008737333782 (grey 50).

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates