various polyethylene implants recallZ-1858-2017 · Zimmer Biomet, Inc. · initiated Feb 17, 2017
Class II Terminated All recalls posted in March 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Endotoxin levels higher than process maximum limits were discovered in the identified polyethylene components.
FDA root cause category: Process design
What the company is doing
On February 17, 2017 URGENT MEDICAL DEVICE RECALL (REMOVAL) - LOT SPECIFIC notifications were sent to the affected consignees via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Distributors, Sales Representatives, and Distributor Operation Managers Your Responsibilities: 1. Review this notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form a. Return a digital copy to [email protected] within three (3) days. 4. Return all affected product from your distributorship and affected hospitals within your territory along with a completed Attachment 1 Inventory Return Certification Form to Zimmer Biomet within 2 weeks. a. Request a Return Authorization Number via email to [email protected] or through FAST/SMS. Specify RECALL as the RGA type. b. For each return, send a copy of Attachment 1 to [email protected]. c. Include a hardcopy of Attachment 1 with your shipment for immediate processing. d. Clearly mark RECALL on the outside of return boxes. 5. Hospitals that received direct shipments of the affected products from Zimmer Biomet will be sent a copy of the Risk Manager Recall Notice directly. Surgeons will also be notified. It is important that you review the list of hospitals and surgeons included with the email notification sent to your facility to identify additional accounts Zimmer Biomet has not notified. Using the Additional Accounts Form provided with the email notice, return contact information for any additional hospitals and surgeons that may have received or used the affected product. If there are no additional
Recall details
- Recall number
- Z-1858-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 17, 2017
- Posted by the FDA
- Mar 27, 2017
- Recalling firm
- Zimmer Biomet, Inc.
- Quantity
- 28,253 in total
- Product code
- JDS
- FDA submissions
- K800144
- Distribution
- Worldwide - US Nationwide in the ARGENTINA AUSTRALIA CANADA CHILE CHINA COSTA RICA ECUADOR GSCC INDIA ITALY JAPAN KOREA, THE REP OF (SOUTH) MALAYSIA MEXICO PANAMA Singapore SWITZERLAND THAILAND UK
Product description
various polyethylene implants BIO PLUG 10MM Product Usage: Non-weight bearing accessory. Sterile, optional speciality components that are non-bioabsorbable, implantable devices designed to be inserted into the medullary canal of a bone during orhtopedic surgery to prevent cement progression in the diaphysis and, therefore, facilitate cement pressurization during the introduction of an implant.
Code information (models, lots, serial numbers)
Item 163005 Lot 309030
Other devices under the same FDA submission
1 device cleared under K800144 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- INTRAMEDULLARY BONE PLUG 130615Biomet Orthopedics, LLC · NAIL, FIXATION, BONE2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 76507).
- Z-1859-2017 · Terminatedvarious polyethylene implants Extremities, Shoulders and Elbows BioModular Shoulder System; Comprehensive Shoulder System; Comprehensive Reverse Shoulder System Product Usage: Sh…
- Z-1860-2017 · Terminatedvarious polyethylene implants Hips Active Articulation ArComXL Bearings Active Articulation E1 Bearings Arcos Taper Cap BIOLOX delta Option Head E1 RingLoc Bi-Polar Freedom Constr…
- Z-1861-2017 · Terminatedvarious polyethylene implants Knees Revision Tibial Tray Systems Oxford Partial Knee Orthopedic Salvage System Vanguard Complete Knee System Vanguard SSK Revision System Biomet Pa…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.