Why it was recalled

After a submission for CLIA database update from Focus Diagnostics to DiaSorin Molecular,the company found that their Legionella IFA device (IF0950) is not 510(k) exempt.

FDA root cause category: No Marketing Application

What the company is doing

Focus Diagnostics/DiaSorin Molecular is issuing a medical device correction for the Legionella IFA (Immuno Fluoresence Assay). On February 1, 2017, the firm was notified that the Legionella IFA was not properly classified according to 21 CFR Part 807. The product is being withdrawn from the market. Use of the product should be immediately discontinued, and any remaining inventory should be immediately destroyed. The firm sent notice to its consignees, which should be reviewed with the facility's laboratory or mediclal director. Patient results obtained with all lots of the affected product should be evaluated by the clinician and if recommended, retested by another method.

Recall details

Recall number
Z-1686-2017
Status
Terminated
Initiated by the company
Feb 15, 2017
Posted by the FDA
Mar 27, 2017
Recalling firm
Focus Diagnostics Inc
Quantity
1,400 units
Product code
LHL
Distribution
Product was distributed to three testing laboratories in California, New Jersey and Virginia

Product description

Legionella IFA Substrate Slide (IF0951) intended for qualitatively detecting and semi-quantitating human serum antibodies to Legionella pneumophila.

Code information (models, lots, serial numbers)

IF0951: 10816101020055, lot numbers 29428, 31871