Skip to content

VIDAS Estradiol II, Ref 30431-01, 60 tests recallZ-1801-2017 · BioMerieux SA · initiated Mar 17, 2017

Class II Terminated All recalls posted in March 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from BioMerieux SA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.

FDA root cause category: Device Design

What the company is doing

bioMerieux sent an Important Product Safety Notice A Product Safety Notice to all affected customers on March accounts on March 17, 2017, informing them not to use VIDAS Estradiol II test to test estradiol level in patients under Fulvestrant therapy. The letter did not request return of any of the product, and requested that they return an acknowledgement form acknowledging receipt of the notice and that they have followed the instructions and implemented the actions in the field notice. Customers with questions were instructed to contact their bioMerieux Customers Service representative. For questions regarding this recall call 314-731-8805.

Recall details

Recall number
Z-1801-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 17, 2017
Posted by the FDA
Mar 20, 2017
Recalling firm
BioMerieux SA
Quantity
834 kits
Product code
CHP
FDA submissions
K955647
Distribution
Nationwide Distribution including AZ, CA, CO, FL, GA, IN, IA, LA, MI, MS, NY, NC, OH, OK, OR, PA, TX, WA

Product description

VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) to the determination of total estradiol concentration in human serum or plasma (heparin).

Code information (models, lots, serial numbers)

Lots 1005081700, 1005172850 & 1005359170

Other devices under the same FDA submission

1 device cleared under K955647 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.