Pressure Injectable Arrowg+ard Blue Plus Multi-Lumen… recallZ-1879-2017 · Arrow International Inc · initiated Mar 17, 2017

Class II Terminated All recalls posted in April 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Arrow International is recalling due to possible voids in the catheter hub, which could lead to an interlumen crossover.

FDA root cause category: Process control

What the company is doing

Arrow International mailed an Urgent Medical Device Notification Letter dated April 4, 2017, to affected customers to inform them of the issue. Customers were asked to discontinue use immediately and return product by completing the enclosed Recall Acknowledgement Form then emailing it to [email protected] or faxing it to 1-855-419-8507. A Customer Service Representative will provide a Return Goods Authorization Number for the products' return. For questions regarding this recall call 866-246-6990.

Recall details

Recall number
Z-1879-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 17, 2017
Posted by the FDA
Apr 10, 2017
Recalling firm
Arrow International Inc
Quantity
310 units
Product code
FOZ
FDA submissions
K071538
Distribution
US Distribution to NJ, IL, and PA

Product description

Pressure Injectable Arrowg+ard Blue Plus Multi-Lumen Central Venous Catheterization Kit. Permits venous access to central circulation by way of subclavian, jugular and femoral veins. The ARROWg+ard technology is intended to provide protection against catheter-related bloodstream infections (CRBSIs). It is not intended to be used as a treatment for existing infections nor is it indicated for long-term use (>30 days). Clinical effectiveness of the ARROW g+ard Blue Plus catheter in preventing CRBSIs compared to the original ARROWg+ard Blue catheter has not been studied

Code information (models, lots, serial numbers)

Product Codes ASK-45703-PRWJ1, ASK-45703-PUAB1, and ASK-45703-PUPM  Lot # 13F16L0105, 23F16K0180 and 13F16L0199