Skip to content

Euro Diagnostica DIASTAT ANA (anti-nuclear antibody) ELISA recallZ-0306-2018 · Euro Diagnostica AB · initiated Feb 14, 2017

Class II Terminated All recalls posted in December 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Euro Diagnostica AB or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Euro Diagnostica has identified a deviation in one of the components of the DIASTAT ANA (FANA200), lot TS 2076. The kit might not perform correctly and so potentially yield false positive samples.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Consignees were notified by phone or email beginning 2/15/2017. The Urgent Field Safety Notice identified affected product, described the issue, and stated affected product should be destroyed. A response form was asked to be returned. Questions can be directed to Malin Bornschein at [email protected]

Recall details

Recall number
Z-0306-2018
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 14, 2017
Posted by the FDA
Dec 20, 2017
Recalling firm
Euro Diagnostica AB
Quantity
534 kits
Product code
LKJ
FDA submissions
K932876
Distribution
MN, NE, TX

Product description

Euro Diagnostica DIASTAT ANA (anti-nuclear antibody) ELISA DIASTAT ANA (anti-nuclear antibody) A positive ANA test result provides presumptive evidence for systemic rheumatic disease; further definition of specific antibody profiles is a valuable aid in the diagnostic process.

Code information (models, lots, serial numbers)

Lot: TS 2076

Other devices under the same FDA submission

1 device cleared under K932876 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page