Medical device recalls: December 2017
The FDA posted 13 medical device recall records in December 2017, from 9 recall events by 8 companies. 3 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips Electronics North America Corporation · 2 events, 2 records
- Stryker Neurovascular · 1 event, 3 records
- BioMerieux SA · 1 event, 2 records
- Physio-Control, Inc. · 1 event, 2 records, 2 Class I
- Euro Diagnostica AB · 1 event, 1 record
- Maquet · 1 event, 1 record
- Medacta Usa Inc · 1 event, 1 record
- Sterilmed Inc · 1 event, 1 record, 1 Class I
FDA root cause
- Process control · 3
- Labeling Change Control · 3
- Under Investigation by firm · 2
- Software design · 1
- Nonconforming Material/Component · 1
- Incorrect or no expiration date · 1
- Error in labeling · 1
- Device Design · 1
Class I recalls (3)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0213-2018 · Sterilmed Inc · posted Dec 29, 2017 · Terminated
Inadequate seal due to lack of adhesive glue
- Z-0186-2018 · Physio-Control, Inc. · posted Dec 12, 2017 · Terminated
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
- Z-0187-2018 · Physio-Control, Inc. · posted Dec 12, 2017 · Terminated
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
Class II recalls (9)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0266-2018 · Medacta Usa Inc · posted Dec 20, 2017 · Terminated
Medacta has observed a worldwide occurrence rate of post-operative insert screw back-out of approximately 0.1%. For this reason, we stress the importance of using the 3.5 Nm torque limiting driver if the screw is used, as directed in our surgical techniques.
- Z-0306-2018 · Euro Diagnostica AB · posted Dec 20, 2017 · Terminated
Euro Diagnostica has identified a deviation in one of the components of the DIASTAT ANA (FANA200), lot TS 2076. The kit might not perform correctly and so potentially yield false positive samples.
- Z-0490-2018 · Maquet · posted Dec 20, 2017 · Terminated
55 kits were distributed after their expiration date of September 2015.
- ETEST OXACILLIN OX 256 US F100Class IIZ-0264-2018 · BioMerieux SA · posted Dec 19, 2017 · Terminated
QC failures (MIC out of range high) for S.aureus ATCC 29213 strain on ETEST OX 256 (OXACILLIN) Foam packaging (Ref. 520558, 520518) reported. The product behavior could lead to False Resistant results.
- ETEST OXACILLIN OX 256 WW F100Class IIZ-0265-2018 · BioMerieux SA · posted Dec 19, 2017 · Terminated
QC failures (MIC out of range high) for S.aureus ATCC 29213 strain on ETEST OX 256 (OXACILLIN) Foam packaging (Ref. 520558, 520518) reported. The product behavior could lead to False Resistant results.
- Azurion 3 M12, Azurion 3 M15, Allura R9 7 M12, Allura R9 M20, Azurion 7 M12, Azurion 7 M20 ↗Class IIZ-0255-2018 · Philips Electronics North America Corporation · posted Dec 18, 2017 · Terminated
When using the downscale option with the Allura R9 system or Azurion R1.1 system, the measurements performed using the QA Basic Measurement tool will not be correctly exported to the external DICOM destination.
- Z-0275-2018 · Stryker Neurovascular · posted Dec 5, 2017 · Terminated
Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the pack.
- Z-0276-2018 · Stryker Neurovascular · posted Dec 5, 2017 · Terminated
Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the pack.
- Z-0277-2018 · Stryker Neurovascular · posted Dec 5, 2017 · Terminated
Stryker Neurovascular has become aware that some 3-Pack Stroke Fast Packs and Trevo Procedure Packs were manufactured using a carton sleeve where the sleeve label contents did not match the physical contents within the pack.
Class III recalls (1)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0268-2018 · Philips Electronics North America Corporation · posted Dec 20, 2017 · Terminated
Some SureSigns VSi, VS2+ products have the Date of Manufacture on its serial number label as %DoM , instead of the specific date in the format of YYYY-MM.