Skip to content

DRX-Revolution Mobile X-Ray System recallZ-2651-2017 · Carestream Health Inc · initiated May 25, 2017

Class II Terminated All recalls posted in June 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Carestream Health Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.

FDA root cause category: Process control

What the company is doing

Carestream Health sent the URGENT MEDICAL DEVICE CORRECTION Letters and Response Forms to the US consignees on May 25, 2017, via Certified Mail, Certified Receipt Requested. Letters to foreign accounts were distributed via e-mail or courier about the same time. Customers with questions were instructed to call 1-800-328-2910.

Recall details

Recall number
Z-2651-2017
Recall class
Class II
Status
Terminated
Initiated by the company
May 25, 2017
Posted by the FDA
Jun 19, 2017
Recalling firm
Carestream Health Inc
Quantity
Domestic: 1449 units; Foreign: 1416 units
Product code
IZL
FDA submissions
K120062
Distribution
Worldwide Distribution - US (nationwide) Internationally to Australia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, India, Japan, Israel, Korea, Malaysia, Maldives, Mexico, New Zealand, Peru, Puerto Rico, Singapore, Switzerland, Taipei, Thailand, Turkey, and UAE.

Product description

DRX-Revolution Mobile X-Ray System; MODEL DRXR-1; Catalog # 1019397 (all other countries) and Catalog 1060177 (China only) -- COMMON/USUAL NAME: DRX-Revolution --- The DRX-Revolution is a mobile system used to generate and control X-Rays for diagnostic procedures

Code information (models, lots, serial numbers)

Serial Numbers: Worldwide except China: 101 through 2623; China only: 800101 through 800482

Other devices under the same FDA submission

1 device cleared under K120062 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page