Medical device recalls: June 2017
The FDA posted 68 medical device recall records in June 2017, from 44 recall events by 34 companies. 7 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Medtest Holdings, Inc. · 4 events, 4 records
- Acumedia Manufacturers, Inc. · 3 events, 3 records
- Exactech, Inc. · 3 events, 3 records
- CareFusion 303, Inc. · 2 events, 2 records
- Cook Inc. · 2 events, 2 records
- O-Two Medical Technologies, Inc. · 2 events, 2 records
- Cell Marque Corporation · 1 event, 12 records
- ICU Medical, Inc. · 1 event, 9 records
FDA root cause
- Nonconforming Material/Component · 12
- Labeling mix-ups · 12
- Device Design · 12
- Under Investigation by firm · 6
- Process control · 6
- Error in labeling · 5
- Component design/selection · 4
- Other · 2
Class I recalls (7)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2736-2017 · Maquet Datascope Corp - Cardiac Assist Division · posted Jun 23, 2017 · Terminated
The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
- Z-2737-2017 · Maquet Datascope Corp - Cardiac Assist Division · posted Jun 23, 2017 · Terminated
The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
- Z-2738-2017 · Maquet Datascope Corp - Cardiac Assist Division · posted Jun 23, 2017 · Terminated
The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.
- Z-2702-2017 · Penumbra Inc. · posted Jun 22, 2017 · Terminated
Penumbra has identified an issue in these four lots involving a raw material component of the delivery wire. This issue could result in breakage of the delivery wire, which could potentially lead to serious patient injury or death.
- Z-2225-2017 · Magellan Diagnostics, Inc. · posted Jun 5, 2017 · Terminated
Underestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
- Z-2226-2017 · Magellan Diagnostics, Inc. · posted Jun 5, 2017 · Terminated
Underestimates the lead concentration of venous blood samples when the sample is analyzed with the LeadCare systems
- Z-2151-2017 · Abbott Vascular · posted Jun 1, 2017 · Terminated
Abbott Vascular is initiating a recall of the Absorb Bioresorbable Vascular Scaffold (BVS) System due to studies showing elevated rates of major adverse events, specifically, myocardial infraction and scaffold thrombosis when compared to patients treated…
Class II recalls (55)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- VIDAS Testosterone, Ref 30418 ↗Class IIZ-2339-2017 · bioMerieux, Inc. · posted Jun 29, 2017 · Terminated
Complaints were received from customers observing falsely overestimate results or external quality control higher results than expected when performing tests with VIDAS Testosterone.
- Z-2645-2017 · Boston Scientific Corporation · posted Jun 27, 2017 · Terminated
A field report indicated some units within this lot were missing the bottom pouch seal. The compromised seal is completely missing and is obvious to the user.
- Z-2698-2017 · Arrow International Inc · posted Jun 27, 2017 · Terminated
Arrow is recalling additional lots that were identified as part of an active recall. Arrow is recalling the affected product due to the possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel…
- Z-2699-2017 · Arrow International Inc · posted Jun 27, 2017 · Terminated
Arrow is recalling additional lots that were identified as part of an active recall. Arrow is recalling the affected product due to the possibility that the catheter peel-away component hub tabs may prematurely detach when the practitioner begins to peel…
- Z-2770-2017 · Biomerieux Inc · posted Jun 27, 2017 · Terminated
The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to…
- Z-2867-2017 · Sentinel CH SPA · posted Jun 27, 2017 · Terminated
Reagent 1 component may result in Quality Control values shifting out of acceptable range, which is not resolved by recalibration.
- Z-2714-2017 · Arjo, Inc. dba ArjoHuntleigh · posted Jun 26, 2017 · Terminated
Reports of the device failure involving delivering continuous pressure to the calf garment and not deflating as intended, without triggering an alarm.
- 18 cm (7") PUR Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, NanoClave, Rotating Luer.Class IIZ-2716-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- Z-2717-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- 28 cm (11") PUR Bifuse Ext Set w/6-Port NanoClave Manifold, Check Valve, 2 NanoClave, Rotating Luer.Class IIZ-2718-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- Z-2719-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- Z-2720-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- Z-2721-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- Z-2722-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- Z-2723-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- Z-2724-2017 · ICU Medical, Inc. · posted Jun 26, 2017 · Terminated
The connection between the Female Luer and the NanoClave manifold has the potential for developing a leak.
- Z-2746-2017 · Medtronic Navigation, Inc. · posted Jun 26, 2017 · Terminated
Software issue related to the StealthStation S7 system and the Synergy Spine application Version 2.1 configured with Spine Tool Install CD version 25. Issue may result in user being unable to navigate the Navigated Elevate Inserter with the StealthAiR…
- Z-2705-2017 · O-Two Medical Technologies, Inc. · posted Jun 22, 2017 · Terminated
Malfunction of an O-Two Adult Ventilation Timer, where the flash rate of the malfunctioning device was reported to be twice that of the rate noted on the label.
- Z-2731-2017 · O-Two Medical Technologies, Inc. · posted Jun 22, 2017 · Terminated
For lack of evidence to validate effectiveness of sterilization dose, O-Two Medical Technologies Inc. hereby announces a voluntary medical device recall of the suspected batches of Burn Relief products to prevent potential contamination. .
- Proteus 235 ↗Class IIZ-2592-2017 · Ion Beam Applications S.A. · posted Jun 19, 2017 · Terminated
It has been observed during testing that there could be clinically non-negligible amount of dark current (a well-known physics effect observed in cyclotron) in Pencil Beam Scanning (PBS) treatment mode when system is not irradiating between layers.
- Z-2621-2017 · Omega Medical Imaging, Inc. · posted Jun 19, 2017 · Terminated
Elevating Monitor Suspension column dropping and will not rise back up.
- Z-2644-2017 · Exactech, Inc. · posted Jun 19, 2017 · Terminated
The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if…
- Z-2651-2017 · Carestream Health Inc · posted Jun 19, 2017 · Terminated
Carestream found a single unit where four screws in the assembly were not applied in the manufacturing process. This could result in separation of the tube and yoke assembly.
- Logic Fit Tibial Tamp Head ↗Class IIZ-2662-2017 · Exactech, Inc. · posted Jun 19, 2017 · Terminated
The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if…
- Alaris PC Unit, Model 8015Class IIZ-2671-2017 · CareFusion 303, Inc. · posted Jun 19, 2017 · Terminated
BD initiated the recall of Alaris PC unit model 8015 after the firm identified five scenarios which can result in the occurrence of Systems Error Code 255-16-275 and can potentially result in interrupted infusions.
- Alaris Pump Module model 8100Class IIZ-2700-2017 · CareFusion 303, Inc. · posted Jun 19, 2017 · Terminated
There is a potential risk that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100.
- Z-2704-2017 · Exactech, Inc. · posted Jun 19, 2017 · Terminated
Missing Screw in carton. Nine (9) serial numbers were identified as being inadvertently packaged as one per carton when two per carton are required.
- Z-2740-2017 · Flowonix Medical, Inc. · posted Jun 19, 2017 · Terminated
Flowonix Medical received a report of a patient implanted with the Prometra II Programmable Pump who may have received a fatal drug overdose during an MRI procedure. The Prometra II pump has an FDA-approved design feature intended to permit safe exposure…
- Z-2533-2017 · Applied Medical Resources Corp · posted Jun 14, 2017 · Terminated
Due to possibility of insufficient sterilization.
- Z-2586-2017 · Medtest Holdings, Inc. · posted Jun 14, 2017 · Terminated
Decrease reactivity of R 1 component.
- Z-2528-2017 · Philips Electronics North America Corporation · posted Jun 13, 2017 · Terminated
Potential for early component failure.
- Z-2624-2017 · Smith & Nephew, Inc. · posted Jun 12, 2017 · Terminated
The incorrect Loctite adhesive was used to assemble the impactor heads.
- Z-2625-2017 · Smith & Nephew, Inc. · posted Jun 12, 2017 · Terminated
The incorrect Loctite adhesive was used to assemble the impactor heads.
- Z-2661-2017 · Medacta Usa Inc · posted Jun 12, 2017 · Terminated
One lot of product does not include a screw for attachment to the tibial base plate.
- Z-2172-2017 · Laserex Systems Inc. · posted Jun 6, 2017 · Terminated
It was discovered the unit produced a laser emission without pressing the fire button when the slit-lamp was driven to its lowest position prior to use on any patient.
- da Vinci Xi Surgical System ↗Class IIZ-2630-2017 · Intuitive Surgical, Inc. · posted Jun 6, 2017 · Terminated
da Vinci Xi(R) Surgical System Patient Side Carts (PSC) were shipped to the field with potentially under-torqued brake screws in a section of the system arms.
- Z-2539-2017 · Cook Inc. · posted Jun 5, 2017 · Terminated
Correction to the product Instructions for Use because the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.
- Z-2540-2017 · Cook Inc. · posted Jun 5, 2017 · Terminated
Cook Inc. is initiating a voluntary recall of the Vue Optic Visualization Source and the Flexor Vue Deflecting Endoscopic System because it was determined that the current reprocessing instructions do not provide sufficient detailed information for the…
- Z-2246-2017 · Baxter Healthcare Corporation · posted Jun 2, 2017 · Terminated
Connection issues between the PD transfer set catheter connector and an adapter may lead to leaks and contamination of the fluid pathway which may result in peritonitis.
- Z-2701-2017 · Natus Medical Incorporated · posted Jun 2, 2017 · Terminated
Olympic Pasteurmatic 3000/3500 Systems are recalled because they do not fully meet todays regulatory requirements. Given the age of the products design, component and technology obsolescence, and the time involved in fully complying with the latest design…
- Adenovirus (20/11 & 2/6) ↗Class IIZ-2199-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- Estrogen Receptor (EP1) ↗Class IIZ-2200-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- Estrogen Receptor (SP1) ↗Class IIZ-2201-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- H. pylori (polyclonal} ↗Class IIZ-2202-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- Her2/Neu (c-erbB-2) (CB-1 1) ↗Class IIZ-2203-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- Her2/Neu (EP3) ↗Class IIZ-2204-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- Herpes Simplex Virus I ↗Class IIZ-2205-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- Herpes Simplex Virus II ↗Class IIZ-2206-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- SV40 (MRQ-4) ↗Class IIZ-2207-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- Varicella Zoster Virus ↗Class IIZ-2208-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- CD117, c-kit (YR145) ↗Class IIZ-2209-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- HHV-8 (13810) ↗Class IIZ-2210-2017 · Cell Marque Corporation · posted Jun 1, 2017 · Terminated
Cell Marque has determined the need to issue a product recall after identifying product intended to be sold outside the U.S. was not properly labeled as "For Export Only" as required by section 801 (e )(1) of the FD&C Act.
- Z-2217-2017 · Cardinal Scale Mfg Co · posted Jun 1, 2017 · Terminated
One complaint of a seat on a scale failing unexpectedly which can cause an injury.
- Assist Bed Rail - 6632 ↗Class IIZ-2218-2017 · Invacare Corporation · posted Jun 1, 2017 · Terminated
A typographical error was noted within the user manual for the 6632 Assist Bed Rail which could result in the Assist Bed Rails being installed incorrectly.
- Z-2219-2017 · Exact Sciences Corporation · posted Jun 1, 2017 · Terminated
Exact Sciences is correcting all Sample Mixer v2 Upgrade units because of incidents of Sample Mixer door damage, which has occurred when the Sample Holder opens during a run and the Sample Holder lid makes contact with the Sample Mixer door.
Class III recalls (6)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-2649-2017 · Medtest Holdings, Inc. · posted Jun 29, 2017 · Terminated
Incorrect control range in product labeling
- Z-2650-2017 · Medtest Holdings, Inc. · posted Jun 29, 2017 · Terminated
Linear performance information in product insert does not match that listed in the approved premarket notification
- Z-2631-2017 · Medtest Holdings, Inc. · posted Jun 26, 2017 · Terminated
Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.
- Z-2212-2017 · Acumedia Manufacturers, Inc. · posted Jun 1, 2017 · Terminated
Storage temperature listed on the label was incorrect. Label stated 2-30C. Actual storage temperature is 2-8C.
- Z-2213-2017 · Acumedia Manufacturers, Inc. · posted Jun 1, 2017 · Terminated
Incorrect expiration date was listed on the label. Correct expiration date was November 30, 2016. Product was labeled with expiration date of April 30, 2018.
- Z-2214-2017 · Acumedia Manufacturers, Inc. · posted Jun 1, 2017 · Terminated
Incorrect expiration date on label. Correct expiration date was March 25, 2019. Product was labeled with expiration date of March 28,2019.