Alaris Pump Module model 8100 recallZ-2700-2017 · CareFusion 303, Inc. · initiated Jun 12, 2017

Class II Terminated All recalls posted in June 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential risk that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100.

FDA root cause category: Device Design

What the company is doing

An Urgent Medical Device Recall letter will be sent on 6/12/17 to customers to inform them that BD is issuing this letter to inform you of a potential risk associated with the Alaris Pump module. Firm identified a specific scenario that could cause unintended flow in the older, centered sear door latch design in the Alaris Pump module model 8100. This scenario is reproducible under the following situations: a) the user does not close the roller clamp on the IV administration set before the pump door is opened as recommended, and b) opening the pump door using an atypical technique with the door latch with the centered sear design. The letter instructs customers If their device has a centered-sear that was manufactured between June 2002 to June 2004, clearly mark and sequester (e.g. Biomed department) the Alaris Pump module that exhibited the issue. Notify CareFusion Customer Advocacy at 888-812-3266 or [email protected] to coordinate an onsite remediation or send the devices to the BD Service Depot.

Recall details

Recall number
Z-2700-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 12, 2017
Posted by the FDA
Jun 19, 2017
Recalling firm
CareFusion 303, Inc.
Quantity
35,940 units
Product code
FRN
FDA submissions
K950419
Distribution
US and Canada

Product description

Alaris Pump Module model 8100

Code information (models, lots, serial numbers)

Notification will be distributed to all Alaris Pumps customers with pumps manufactured between June 2002 through June 2004.

Other devices under the same FDA submission

1 device cleared under K950419 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.